Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05414734055659), received between Jul 9, 2018 and Jun 4, 2019 — classified as 2 injuries.

The most frequently reported specific device problem is “Perivalvular Leak” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Perivalvular Leak1

Patient problems

  • Aortic Insufficiency1
  • Aortic Regurgitation1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 4, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFORMATION RECEIVED FROM PATIENT DEVICE TRACKING NOTES THAT ON (B)(6) 2019, A 25MM EPIC VALVE WAS NOT IMPLANTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8666302 ↗
  • InjuryJul 9, 2018Perivalvular Leak; Perivalvular Leak

    ON (B)(6) 2018, A 25MM EPIC SUPRA VALVE WAS IMPLANTED. AFTER REMOVING THE PATIENT FROM BY-PASS, A POST-IMPLANT TEE SHOWED AORTIC VALVE REGURGITATION AND SIGNIFICANT PARAVALVULAR LEAKAGE. THE PATIENT WAS RETURNED TO BY-PASS AND THE EPIC VALVE WAS REMOVED AND REPLACED WITH A MECHANICAL VALVE. THE USER COMMENTS THE COR-KNOT DID NOT FIRE PROPERLY. THE PATIENT IS REPORTED TO BE STABLE.

    Patient problems: Aortic Insufficiency; Aortic Regurgitation

    MAUDE report 7671274 ↗