Fast-Cath™ 406333Recalls & reported problems

St. Jude Medical, Inc. · HEMOSTASIS INTRODUCER · Class 2

All models of Fast-Cath (71)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05414734203876), received between Sep 13, 2021 and Nov 10, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak5
  • Gas/Air Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 10, 2025Fluid/Blood Leak; Fluid/Blood Leak

    DURING AN ATRIAL FIBRILLATION PROCEDURE, A LEAK FROM THE HEMOSTASIS VALVE WAS NOTED. THEREFORE, IT WAS REMOVED AND THE INTRODUCER WAS REPLACED. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23511465 ↗
  • MalfunctionJan 20, 2023Fluid/Blood Leak; Fluid/Blood Leak

    DURING THE PAROXYSMAL SUPRAVENTRICULAR TACHYCARDIA PROCEDURE, THE SHEATH WAS INSERTED INTO THE PATIENT, AND BLOOD LEAKED FROM THE HEMOSTATIC VALVE OF THE SHEATH. THIS OCCURRED BEFORE THE CATHETER OR SEPTAL PUNCTURE NEEDLE WAS INSERTED. THE INTRODUCER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT. THE HOSPITAL DISCARDED THE REPORTED DEVICE. NO ADDITIONAL VENIPUNCTURE WAS PERFORMED FOR SHEATH REPLACEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16218554 ↗
  • MalfunctionFeb 28, 2022Fluid/Blood Leak; Fluid/Blood Leak

    MANUFACTURER REF# 3005334138-2022-00072. DURING A PAROXYSMAL SUPRAVENTRICULAR TACHYCARDIA ABLATION PROCEDURE, BLOOD WAS LEAKING FROM THE HEMOSTASIS VALVES OF THE SHEATH, AFTER INSERTING THE SHEATH INTO A PATIENT PRIOR TO INSERTING CATHETERS. THE SHEATH SET WAS REPLACED BUT BLOOD LEAK WAS NOTED AGAIN. THE SHEATH WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13616364 ↗
  • MalfunctionFeb 28, 2022Fluid/Blood Leak; Fluid/Blood Leak

    RELATED MANUFACTURER REF# 3005334138-2022-00073. DURING A PAROXYSMAL SUPRAVENTRICULAR TACHYCARDIA ABLATION PROCEDURE, A BLOOD WAS NOTED FROM THE HEMOSTASIS VALVES OF THE SHEATH. THE SHEATH SET WAS REPLACED WHICH DID NOT RESOLVE THE ISSUE. THE SHEATH WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13616360 ↗
  • MalfunctionDec 28, 2021Gas/Air Leak; Gas/Air Leak

    DURING A PULMONARY VEIN ISOLATION PROCEDURE, AIR WAS ASPIRATED INTO THE DEVICE WHEN ATTEMPTING TO FLUSH. THE INTRODUCER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13094147 ↗
  • MalfunctionSep 13, 2021Fluid/Blood Leak; Fluid/Blood Leak

    DURING A PAROXYSMAL SUPRAVENTRICULAR TACHYCARDIA (PSVT) ABLATION PROCEDURE, A BLOOD LEAK WAS NOTED FROM THE HEMOSTASIS VALVE OF THE SHEATH. THE SHEATH SET WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12460269 ↗

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →