Swartz™ 407358Recalls & reported problems
St. Jude Medical, Inc. · BRAIDED TRANSSEPTAL GUIDING INTRODUCER · Class 2
All models of Swartz (17)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 05414734205535), received between Feb 14, 2020 and Jan 20, 2026 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Flaked” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Flaked3
- Fluid/Blood Leak2
- Material Separation2
- Difficult to Insert1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 20, 2026Fracture
THIS REPORT IS TO ADVISE OF A FRACTURE FOUND IN THE SHEATH DURING ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24115353 ↗MalfunctionDec 12, 2025Material Separation; Material Separation
DURING AN ATRIAL FIBRILLATION PROCEDURE, AFTER THE CATHETER WAS REMOVED FROM THE PATIENT, IT WAS NOTED THAT THE GRAY TIP OF THE SHEATH WAS MISSING AFTER THE PROCEDURE. IT LOOKED TO BE TORN. THE PROCEDURE WAS COMPLETED WITHOUT REPLACING THE DEVICE AND THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23786582 ↗MalfunctionDec 2, 2025Material Separation; Material Separation
DURING THE CAVO-TRICUSPID ISTHMUS PROCEDURE, THE HUB OF THE SHEATH DISCONNECTED. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23693000 ↗InjuryMay 16, 2024Flaked; Difficult to Insert; Flaked; Difficult to Insert
DURING THE MITRACLIP TAVI/TAVR PROCEDURE, WHILE ADVANCING THE TRANSSEPTAL NEEDLE THROUGH THE SHEATH, RESISTANCE WAS NOTED, REQUIRING AN ADDITIONAL FEMORAL PUNCTURE SITE. THE DEVICE WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19333348 ↗MalfunctionJan 31, 2024Flaked; Flaked
DURING AN ATRIAL FIBRILLATION PROCEDURE, SKIVING OCCURRED. INSERTION DIFFICULTIES WERE NOTED WHILE ADVANCING THE NEEDLE INTO THE SHEATH INTO THE PATIENT. THE SHEATH AND NEEDLE WERE REMOVED AND TESTED ON THE BENCH BY PUSHING AGAINST THE RESISTANCE, THE NEEDLE SHEARED OFF A SECTION OF THE PLASTIC. THE SAME BENCH-TOP TESTING WAS PERFORMED WITH THE SECOND SHEATH AND NEEDLE, AND THE SAME ISSUE WAS NOTED. THE ISSUE WAS EVENTUALLY RESOLVED, AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE PATIENT HEALTH CONSEQUENCES. THERE HAS BEEN NO SUGGESTION THAT ANY PLASTIC WAS DISLODGED IN THE PATIENT'S CIRCULATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18616532 ↗MalfunctionJan 29, 2024Fluid/Blood Leak; Fluid/Blood Leak
DURING A PREMATURE VENTRICULAR CONTRACTION PROCEDURE WITH CARTO, A BLOOD LEAK WAS NOTED FROM THE HEMOSTASIS VALVE. THE INTRODUCER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18596907 ↗MalfunctionSep 2, 2022Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT IT TO ADVISE OF A LEAK IN THE SHEATH DURING ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15348611 ↗MalfunctionFeb 14, 2020Flaked; Flaked
THIS EVENT IS BEING REPORTED BASED ON ANALYSIS REVEALING SKIVING OF THE DILATOR.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9710939 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page