Swartz™ 407358Recalls & reported problems

St. Jude Medical, Inc. · BRAIDED TRANSSEPTAL GUIDING INTRODUCER · Class 2

All models of Swartz (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 05414734205535), received between Feb 14, 2020 and Jan 20, 2026 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Flaked” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Flaked3
  • Fluid/Blood Leak2
  • Material Separation2
  • Difficult to Insert1
  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 20, 2026Fracture

    THIS REPORT IS TO ADVISE OF A FRACTURE FOUND IN THE SHEATH DURING ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24115353 ↗
  • MalfunctionDec 12, 2025Material Separation; Material Separation

    DURING AN ATRIAL FIBRILLATION PROCEDURE, AFTER THE CATHETER WAS REMOVED FROM THE PATIENT, IT WAS NOTED THAT THE GRAY TIP OF THE SHEATH WAS MISSING AFTER THE PROCEDURE. IT LOOKED TO BE TORN. THE PROCEDURE WAS COMPLETED WITHOUT REPLACING THE DEVICE AND THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23786582 ↗
  • MalfunctionDec 2, 2025Material Separation; Material Separation

    DURING THE CAVO-TRICUSPID ISTHMUS PROCEDURE, THE HUB OF THE SHEATH DISCONNECTED. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23693000 ↗
  • InjuryMay 16, 2024Flaked; Difficult to Insert; Flaked; Difficult to Insert

    DURING THE MITRACLIP TAVI/TAVR PROCEDURE, WHILE ADVANCING THE TRANSSEPTAL NEEDLE THROUGH THE SHEATH, RESISTANCE WAS NOTED, REQUIRING AN ADDITIONAL FEMORAL PUNCTURE SITE. THE DEVICE WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19333348 ↗
  • MalfunctionJan 31, 2024Flaked; Flaked

    DURING AN ATRIAL FIBRILLATION PROCEDURE, SKIVING OCCURRED. INSERTION DIFFICULTIES WERE NOTED WHILE ADVANCING THE NEEDLE INTO THE SHEATH INTO THE PATIENT. THE SHEATH AND NEEDLE WERE REMOVED AND TESTED ON THE BENCH BY PUSHING AGAINST THE RESISTANCE, THE NEEDLE SHEARED OFF A SECTION OF THE PLASTIC. THE SAME BENCH-TOP TESTING WAS PERFORMED WITH THE SECOND SHEATH AND NEEDLE, AND THE SAME ISSUE WAS NOTED. THE ISSUE WAS EVENTUALLY RESOLVED, AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE PATIENT HEALTH CONSEQUENCES. THERE HAS BEEN NO SUGGESTION THAT ANY PLASTIC WAS DISLODGED IN THE PATIENT'S CIRCULATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18616532 ↗
  • MalfunctionJan 29, 2024Fluid/Blood Leak; Fluid/Blood Leak

    DURING A PREMATURE VENTRICULAR CONTRACTION PROCEDURE WITH CARTO, A BLOOD LEAK WAS NOTED FROM THE HEMOSTASIS VALVE. THE INTRODUCER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18596907 ↗
  • MalfunctionSep 2, 2022Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT IT TO ADVISE OF A LEAK IN THE SHEATH DURING ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15348611 ↗
  • MalfunctionFeb 14, 2020Flaked; Flaked

    THIS EVENT IS BEING REPORTED BASED ON ANALYSIS REVEALING SKIVING OF THE DILATOR.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9710939 ↗