Swartz™ 407456Recalls & reported problems
St. Jude Medical, Inc. · BRAIDED TRANSSEPTAL GUIDING INTRODUCER · Class 2
All models of Swartz (17)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05414734205801), received between Dec 15, 2020 and Mar 19, 2021 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture3
- Degraded2
- Break1
- Crack1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient2
- Foreign Body Embolism1
- Foreign Body In Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 19, 2021Degraded; Fracture; Degraded; Fracture
DURING A SUPRAVENTRICULAR TACHYCARDIA PROCEDURE, THE FIRST INTRODUCER BROKE AT THE HUB. A SECOND DEVICE WAS USED, AND THE TIP OF THE SHEATH DETACHED AND REMAINED IN THE PATIENT. WHILE INTRODUCING THE SHEATH AND DILATOR OVER THE LONG 032 GUIDEWIRE INTO THE RIGHT FEMORAL VEIN, THE PROXIMAL PORTION OF THE SHEATH FRACTURED AT THE JUNCTURE OF THE SHAFT AND THE VALVE. THE SHEATH WITH DILATOR WERE THEN REMOVED OVER THE J WIRE TO EXCHANGE FOR A NEW ONE. WHILE REMOVING THE SHEATH, THE DISTAL 1/3 OF IT DETACHED AND REMAINED IN THE SUPERIOR VENA CAVA AT THE LEVEL OF THE RIGHT ATRIUM. A SNARE WAS USED TO RETRIEVE THE DETACHED PORTION AND BRING IT DOWN TO THE LEVEL OF THE RIGHT FEMORAL VEIN. WHILE SNARED, THE SHEATH BEGAN TO BREAK INTO MULTIPLE PIECES AND COULD NOT BE REMOVED PERCUTANEOUSLY. VASCULAR SURGERY WAS NECESSARY TO REMOVE THE PIECES WITH A CUT-DOWN TO THE RIGHT FEMORAL VEIN. INSPECTION OF THE REMAINING, INTACT PORTION OF THE SHEATH REVEALED INNUMEROUS MICRO-FRACTURES THAT TRANSVERSED THE ENTIRETY OF THE SHAFT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11527161 ↗InjuryMar 19, 2021Degraded; Material Separation; Degraded; Material Separation
DURING A SUPRAVENTRICULAR TACHYCARDIA PROCEDURE, THE FIRST INTRODUCER BROKE AT THE HUB. A SECOND DEVICE WAS USED, AND THE TIP OF THE SHEATH DETACHED AND REMAINED IN THE PATIENT. WHILE INTRODUCING THE SHEATH AND DILATOR OVER THE LONG 032 GUIDEWIRE INTO THE RIGHT FEMORAL VEIN, THE PROXIMAL PORTION OF THE SHEATH FRACTURED AT THE JUNCTURE OF THE SHAFT AND THE VALVE. THE SHEATH WITH DILATOR WERE THEN REMOVED OVER THE J WIRE TO EXCHANGE FOR A NEW ONE. WHILE REMOVING THE SHEATH, THE DISTAL 1/3 OF IT DETACHED AND REMAINED IN THE SUPERIOR VENA CAVA AT THE LEVEL OF THE RIGHT ATRIUM. A SNARE WAS USED TO RETRIEVE THE DETACHED PORTION AND BRING IT DOWN TO THE LEVEL OF THE RIGHT FEMORAL VEIN. WHILE SNARED, THE SHEATH BEGAN TO BREAK INTO MULTIPLE PIECES AND COULD NOT BE REMOVED PERCUTANEOUSLY. VASCULAR SURGERY WAS NECESSARY TO REMOVE THE PIECES WITH A CUT-DOWN TO THE RIGHT FEMORAL VEIN. INSPECTION OF THE REMAINING, INTACT PORTION OF THE SHEATH REVEALED INNUMEROUS MICRO-FRACTURES THAT TRANSVERSED THE ENTIRETY OF THE SHAFT.
Patient problems: Foreign Body Embolism
MAUDE report 11527144 ↗InjuryMar 16, 2021Break; Crack
PATIENT UNDERGOING PLANNED ABLATION PROCEDURE. DURING SL2 SHEATH ADVANCEMENT INTO RIGHT ATRIUM, THE DISTAL 10CM END OF THE SHEATH WAS FRACTURED. A TOTAL OF THREE PACKAGES OF THIS PRODUCT HAD BEEN OPENED FOR THIS CASE: THE FIRST ONE BROKE AT THE HUB, THE SECOND HAD APPROXIMATELY 10 CM OF THE TIP THAT BROKE OFF IN THE PATIENT (AS NOTED PREVIOUSLY) AND THE THIRD WAS NOTED TO HAVE CRACKING AND WAS NOT USED. THE BROKEN CATHETER TIP AND SHEATH IN THE PATIENT REQUIRED A SURGICAL CUT-DOWN BY VASCULAR SURGERY TO REMOVE ALL THE PIECES. FDA SAFETY REPORT ID# (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 11506589 ↗MalfunctionDec 15, 2020Fracture; Fracture
DURING THE PROCEDURE, THE TIP AND CENTER OF THE INTRODUCERS WERE FRACTURED. THE FRACTURES WERE SEGMENTED ALONG THE INTRODUCER SHAFT, AND THE MATERIAL WAS BEING HELD TOGETHER BY THE BRAIDED MATERIAL. THE INTRODUCERS WERE REPLACED, AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL ACCESS SITE WAS REQUIRED DURING REPLACEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11013872 ↗MalfunctionDec 15, 2020Fracture; Fracture
RELATED MANUFACTURING REF: 3005334138-2020-00630. DURING THE PROCEDURE, THE TIP AND CENTER OF THE INTRODUCERS WERE FRACTURED. THE FRACTURES WERE SEGMENTED ALONG THE INTRODUCER SHAFT AND THE MATERIAL WAS BEING HELD TOGETHER BY THE BRAIDED MATERIAL. THE INTRODUCERS WERE REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL ACCESS SITE WAS REQUIRED DURING REPLACEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11013688 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page