Fast-Cath™ 406700Recalls & reported problems
St. Jude Medical, Inc. · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05414734216937), received between Jun 23, 2020 and Jun 23, 2020 — classified as 1 injury.
The most frequently reported specific device problem is “Material Separation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Material Separation1
Patient problems
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJun 23, 2020Material Separation; Material Separation
DURING A PROCEDURE, A PART OF THE SHEATH DETACHED AND HAD TO BE RETRIEVED. WHEN THE SHEATH WAS ATTEMPTED TO BE REMOVED, THE SHEATH HAD BROKEN AND A PIECE WAS LEFT INSIDE THE FEMORAL VEIN. A 10FR SHEATH AND PTCA BALLOON WERE USED TO RETRIEVE THE PIECE, REQUIRING AN ADDITIONAL ACCESS SITE. IT WAS CONFIRMED WITH FLUOROSCOPY AND ANGIOGRAPHY THAT NO FRAGMENT REMAINED IN THE PATIENT. NO FURTHER INTERVENTION WAS NEEDED. THE PROCEDURE WAS FINISHED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 10183981 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page