Fast-Cath™ 406702Recalls & reported problems
St. Jude Medical, Inc. · Class 2
All models of Fast-Cath (71)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 05414734216951), received between Sep 19, 2019 and Mar 19, 2025 — classified as 5 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Material Separation” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation8
- Fluid/Blood Leak1
- Fracture1
- Material Twisted/Bent1
Patient problems
- Foreign Body Embolism3
- Foreign Body In Patient3
- No Clinical Signs, Symptoms or Conditions3
- Device Embedded In Tissue or Plaque2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 19, 2025Material Separation; Material Separation
AFTER AN ATYPICAL FLUTTER IN THE RIGHT ATRIUM, AFTER THE PHYSICIAN TAKE OUT ALL THE CATHETERS FROM THE PATIENT, HE TRIED TO TAKE OUT THE INTRODUCER BUT NOTED IT WAS DAMAGED. THE INTRODUCER WAS SPLIT IN TWO PIECES. THERE ARE NO PICTURES AVAILABLE. THERE WAS NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21645864 ↗MalfunctionJun 18, 2024Fracture; Fracture
DURING THE PREMATURE VENTRICULAR CONTRACTION PROCEDURE, WHILE PERFORMING THE GROIN PUNCTURE IT WAS DISCOVERED THAT THE SHEATH WAS DAMAGED, IT LOOKED AS IF THE DILATOR HAD SLIT IT OPEN FROM THE INSIDE. THE SHEATH WAS REPLACED, AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19558630 ↗InjuryJun 12, 2024Material Separation; Material Separation
DURING THE PROCEDURE, WHILE CLOSING USING THE FIGURE 8 STITCH AS USUAL, THE SHEATH BROKE AND REQUIRED INTERVENTION TO SUCCESSFULLY RETRIEVE FROM THE PATIENT. THE PATIENT IS STABLE.
Patient problems: Foreign Body In Patient
MAUDE report 19524770 ↗MalfunctionJun 26, 2023Material Separation; Material Separation
DURING AN ATRIAL FIBRILLATION PROCEDURE, STENOSIS WAS OBSERVED WITH THE 1ST FAST-CATH HEMOSTASIS INTRODUCER INSERTION AT THE RIGHT FEMORAL VEIN. BEFORE INSERTING THE 2ND FAST-CATH HEMOSTASIS INTRODUCER, IT WAS FOUND THAT THE TRANSSEPTAL NEEDLE HAD INADVERTENTLY PUNCTURED AND FRACTURED THE FIRST SHEATH. THE BROKEN SHEATH WAS REMOVED WITH A SNARE AND REPLACED WITH A NON-ABBOTT SHEATH TO COMPLETE THE PROCEDURE WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 17206299 ↗InjuryDec 21, 2022Material Separation; Material Separation
DURING THE PROCEDURE, A PORTION OF SHEATH DETACHED IN THE PATIENT. THE PATIENT IS STABLE.
Patient problems: Foreign Body Embolism; Foreign Body In Patient
MAUDE report 16023005 ↗MalfunctionMay 11, 2022Fluid/Blood Leak; Material Twisted/Bent; Fluid/Blood Leak; Material Twisted/Bent
DURING THE PROCEDURE, IT WAS NOTED THAT THERE WAS BLOOD FLOW BACK IN THE SHEATH AND THE FLUSH PORT WAS WARPED. THE PORT WAS CLAMPED WITH A PAIR OF HEMOSTATS TO STOP THE BACK FLOW AND THE CASE WAS COMPLETED WITH NO CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14359571 ↗InjuryApr 13, 2021Material Separation; Material Separation
DURING AN ATRIAL FIBRILLATION ABLATION PROCEDURE UTILIZING A FEMORAL ACCESS SITE, WHEN ATTEMPTING TO REMOVE THE SHEATH RESISTANCE WAS NOTED AND WHEN FULLY REMOVED FROM THE PATIENT, A PART OF THE SHEATH WAS LEFT IN THE PATIENT. A SNARE WAS USED TO REMOVE THE BROKEN OFF PIECE AND THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE PROCEDURE WAS STILL NOTED TO HAVE BEEN COMPLETED. ALL OF THE PIECES WERE ACCOUNTED FOR WITH BOTH FLUOROSCOPY AND ECHOCARDIOGRAM. NO INSERTION DIFFICULTIES OR ANATOMICAL ISSUES WERE NOTICED WHEN INTRODUCING THE INTRODUCER.
Patient problems: Foreign Body Embolism
MAUDE report 11657799 ↗MalfunctionJan 14, 2021Material Separation; Material Separation
DURING A VENTRICULAR TACHYCARDIA ABLATION, WHEN THE SHEATHS WERE PULLED FROM THE RIGHT FEMORAL VEIN WHILE HOLDING PRESSURE, IT WAS NOTED UNDER FLUOROSCOPY THAT THE INTRODUCER HAD SHEARED OFF AND THE LOWER HALF WAS RETAINED IN THE PATIENT. THE SHEATH PORTION WAS RETRIEVED AND REMOVED UNDER FLUOROSCOPY.
Patient problems: Foreign Body Embolism; No Consequences Or Impact To Patient
MAUDE report 11169929 ↗InjurySep 19, 2019Material Separation; Material Separation
MANUFACTURER RELATED REF: 3005334138-2019-00511. DURING THE PROCEDURE, THE SHEATH BROKE INSIDE THE FEMORAL VEIN. A VASCULAR SURGEON WAS ABLE TO REMOVE THE SHEATH. A SECOND SHEATH WAS USED AND THE DEVICE KINKED AND FORMED IN THE DEVICE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9086474 ↗InjurySep 19, 2019Material Separation; Material Separation
MANUFACTURER RELATED REF: 3005334138-2019-00512. DURING THE PROCEDURE, THE SHEATH BROKE INSIDE THE FEMORAL VEIN. A VASCULAR SURGEON WAS ABLE TO REMOVE THE SHEATH. A SECOND SHEATH WAS USED AND THE DEVICE KINKED AND FORMED IN THE DEVICE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9086461 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page