Quattrode™ 3149 (UDI 05414734401593)Recalls & reported problems
St. Jude Medical, Inc. · Lead, 3/4mm, 90 cm · Class 3 · Implant
All models of Quattrode (17)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Quattrode 3149 lead is part of an implanted spinal cord stimulator system used for pain relief. Most of the 50 reports describe leads or the battery pack giving poor or no pain relief, high impedance readings, or a stimulator that stopped communicating, sometimes with tissue erosion or unwanted overstimulation. All 50 reports describe injuries and none report deaths; some patients needed surgery to remove and replace leads or the whole system. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 50
- Deaths reported
- 0
- Injuries reported
- 50
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 05414734401593), received between Oct 23, 2018 and Aug 3, 2026 — classified as 50 injuries.
The most frequently reported specific device problem is “Therapeutic or Diagnostic Output Failure” (17 reports).
14 other recalls cover devices cleared under the same FDA submission (P010032); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Therapeutic or Diagnostic Output Failure17
- Adverse Event Without Identified Device or Use Problem13
- High impedance7
- Impedance Problem6
- Insufficient Device Problem Information3
- Inappropriate/Inadequate Shock/Stimulation2
- Migration2
Patient problems
- Inadequate Pain Relief35
- Erosion5
- Device Overstimulation of Tissue4
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient3
- Pain3
- Burning Sensation2
- Unspecified Infection2
Recent adverse event reports
The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.
InjuryAug 3, 2026Adverse Event Without Identified Device or Use Problem
It was reported that following weight loss, the patient experienced skin erosion at the lead and extension sites. As a result, surgical intervention may be undertaken to address the issue.
Patient problems: Erosion
MAUDE report 50119092 ↗InjuryOct 4, 2024High impedance; High impedance
ADDITIONAL INFORMATION INDICATES SURGICAL INTERVENTION TOOK PLACE ON (B)(6) 2024 WHEREIN ALL THE THREE LEADS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.
Patient problems: Failure of Implant; Inadequate Pain Relief
MAUDE report 20375227 ↗InjuryMar 4, 2024Impedance Problem; Impedance Problem
RELATED MANUFACTURER REFERENCE NUMBER: (B)(4). IT WAS REPORTED THAT THE PATIENT'S LEADS HAD INVALID IMPEDANCES. AS A RESULT, THE PATIENT UNDERWENT SURGICAL INTERVENTION WHEREIN THE LEADS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18833349 ↗InjuryMar 4, 2024Impedance Problem; Impedance Problem
RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2024-07335, 1627487-2024-07336,1627487-2024-07338. IT WAS REPORTED THAT THE PATIENT'S LEADS HAD INVALID IMPEDANCES. AS A RESULT, THE PATIENT UNDERWENT SURGICAL INTERVENTION WHEREIN THE LEADS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18833333 ↗InjuryAug 3, 2023High impedance; High impedance
RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2023-03649. IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY AND BOTH OF THE PATIENT'S LEADS HAD HIGH IMPEDANCES. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN IN THE FUTURE TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 17455278 ↗InjuryAug 3, 2023High impedance; High impedance
RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2023-03650. IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY AND BOTH OF THE PATIENT'S LEADS HAD HIGH IMPEDANCES. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN IN THE FUTURE TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 17455276 ↗InjuryOct 16, 2022Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure
RELATED MANUFACTURER REFERENCE NUMBERS: 1627487-2022-05484; 1627487-2022-05487. IT WAS REPORTED THE PATIENT'S IPG FAILED TO ESTABLISH COMMUNICATION WITH EXTERNAL DEVICES. A MANUFACTURER REPRESENTATIVE CONFIRMED THE ISSUE AND THE IPG WAS DEEMED INOPERABLE. THE PATIENT ALSO EXPERIENCED INEFFECTIVE THERAPY IN THE SUPRAORBITAL AND OCCIPITAL REGIONS BILATERALLY. AS A RESULT, SURGICAL INTERVENTION OCCURRED WHEREIN THE IPG AND LEADS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. THE INVESTIGATION DID NOT DETERMINE WHICH OF THE LEADS ARE RELATED TO INEFFECTIVE THERAPY.
Patient problems: Inadequate Pain Relief
MAUDE report 15613131 ↗InjuryMay 11, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RELATED MANUFACTURER NUMBER:1627487-2022-02645, 1627487-2022-02646, 1627487-2022-02649. IT WAS REPORTED THE PATIENT ENTIRE SYSTEM WAS EXPLANTED DUE TO INEFFECTIVE STIMULATION. DATE OF EXPLANT IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14366671 ↗InjuryAug 26, 2021High impedance; High impedance
ADDITIONAL INFORMATION RECEIVED INDICATES THE LEAD WAS REPLACED TO ADDRESS THE ISSUE. PATIENT HAD EFFECTIVE THERAPY POST OPERATIVELY.
Patient problems: Inadequate Pain Relief
MAUDE report 12374840 ↗InjuryAug 26, 2021High impedance; High impedance
ADDITIONAL INFORMATION RECEIVED INDICATES THE LEAD WAS REPLACED TO ADDRESS THE ISSUE. PATIENT HAD EFFECTIVE THERAPY POST OPERATIVELY.
Patient problems: Inadequate Pain Relief
MAUDE report 12374827 ↗InjuryApr 13, 2021Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure
ADDITIONAL INFORMATION WAS RECEIVED THAT SURGERY OCCURRED IN WHICH THE PATIENT'S SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 11662547 ↗InjuryApr 13, 2021Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure
RELATED MANUFACTURER REFERENCE NUMBERS: 1627487-2021-13162, 1627487-2021-13164, 1627487-2021-13165, 1627487-2021-13166, 1627487-2021-13167, 1627487-2021-13168. IT WAS REPORTED THAT EFFECTIVE THERAPY WAS NEVER ESTABLISHED FOR THE PATIENT. AS A RESULT SURGERY MAY OCCUR TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 11662545 ↗InjuryJan 29, 2021Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure
RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2021-00754, 1627487-2021-00755, 1627487-2021-00757, 1627487-2021-00758, 1627487-2021-00759, 1627487-2021-00760. IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION. SURGICAL INTERVENTION TOOK PLACE WHEREIN THE SYSTEM WAS EXPLANTED.
Patient problems: Inadequate Pain Relief
MAUDE report 11250540 ↗InjuryJan 29, 2021Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure
RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2021-00754, 1627487-2021-00756, 1627487-2021-00757, 1627487-2021-00758, 1627487-2021-00759, 1627487-2021-00760. IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION. SURGICAL INTERVENTION TOOK PLACE WHEREIN THE SYSTEM WAS EXPLANTED.
Patient problems: Inadequate Pain Relief
MAUDE report 11250535 ↗InjuryNov 26, 2020Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure
RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2020-31740, 1627487-2020-31741, 1627487-2020-48257, 1627487-2020-48258, 1627487-2020-48262, 1627487-2020-48263, 1627487-2020-48264, 1627487-2020-48265, 1627487-2020-48266. IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION. SURGICAL INTERVENTION TOOK PLACE WHEREIN BOTH PATIENT¿S SYSTEMS WERE EXPLANTED AND REPLACED WITH ONE IPG, THREE LEADS AND TWO EXTENSIONS. EFFECTIVE THERAPY WAS ESTABLISHED.
Patient problems: Inadequate Pain Relief
MAUDE report 10909214 ↗InjuryNov 26, 2020Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure
RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2020-31740, 1627487-2020-31741, 1627487-2020-48257, 1627487-2020-48258, 1627487-2020-48261, 1627487-2020-48263, 1627487-2020-48264, 1627487-2020-48265, 1627487-2020-48266. IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION. SURGICAL INTERVENTION TOOK PLACE WHEREIN BOTH PATIENT¿S SYSTEMS WERE EXPLANTED AND REPLACED WITH ONE IPG, THREE LEADS AND TWO EXTENSIONS. EFFECTIVE THERAPY WAS ESTABLISHED.
Patient problems: Inadequate Pain Relief
MAUDE report 10909173 ↗InjuryOct 10, 2020Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure
RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2020-32255, 1627487-2020-32256, 1627487-2020-32258, 1627487-2020-32260, 1627487-2020-32261, 1627487-2020-32262. IT WAS REPORTED THE PATIENT EXPERIENCED INADEQUATE STIMULATION WITH THEIR SCS SYSTEM. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED.
Patient problems: Inadequate Pain Relief
MAUDE report 10660058 ↗InjuryOct 10, 2020Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure
RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2020-32255, 1627487-2020-32256, 1627487-2020-32257, 1627487-2020-32260, 1627487-2020-32261, 1627487-2020-32262. IT WAS REPORTED THE PATIENT EXPERIENCED INADEQUATE STIMULATION WITH THEIR SCS SYSTEM. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED.
Patient problems: Inadequate Pain Relief
MAUDE report 10660057 ↗InjuryOct 6, 2020High impedance; High impedance
RELATED MANUFACTURER REFERENCE NUMBERS: 1627487-2020-32027. 1627487-2020-32028. 1627487-2020-32030. 1627487-2020-32031. 1627487-2020-32033. 1627487-2020-32034. IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY ON THE LEFT SIDE. DIAGNOSTIC TESTING REVEALED HIGH IMPEDANCE. REPROGRAMMING DID NOT RESOLVE THE ISSUE. SURGERY MAY OCCUR TO ADDRESS THE ISSUE. WHICH DEVICES ARE RELATED TO THE ISSUE IS UNKNOWN, SO ALL SUSPECTED DEVICES ARE BEING REPORTED.
Patient problems: Inadequate Pain Relief
MAUDE report 10632783 ↗InjuryOct 6, 2020High impedance; High impedance
RELATED MANUFACTURER REFERENCE NUMBERS: 1627487-2020-32028. 1627487-2020-32029. 1627487-2020-32030. 1627487-2020-32031. 1627487-2020-32033. 1627487-2020-32034. IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY ON THE LEFT SIDE. DIAGNOSTIC TESTING REVEALED HIGH IMPEDANCE. REPROGRAMMING DID NOT RESOLVE THE ISSUE. SURGERY MAY OCCUR TO ADDRESS THE ISSUE. WHICH DEVICES ARE RELATED TO THE ISSUE IS UNKNOWN, SO ALL SUSPECTED DEVICES ARE BEING REPORTED.
Patient problems: Inadequate Pain Relief
MAUDE report 10632780 ↗InjurySep 1, 2020Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2020-30596, RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2020-30597, RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2020-30598. IT WAS REPORTED THAT PATIENT EXPERIENCED UNCOMFORTABLE STIMULATION. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN AT UNKNOWN DATE WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED.
Patient problems: Device Overstimulation of Tissue
MAUDE report 10480491 ↗InjurySep 1, 2020Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2020-30596, RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2020-30597, RELATED MANUFACTURER REFERENCE NUMBER :1627487-2020-30599. IT WAS REPORTED THAT PATIENT EXPERIENCED UNCOMFORTABLE STIMULATION. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN AT UNKNOWN DATE WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED.
Patient problems: Device Overstimulation of Tissue
MAUDE report 10480486 ↗InjuryJul 7, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RELATED MANUFACTURER REFERENCE NUMBER 1627487-2020-22528; RELATED MANUFACTURER REFERENCE NUMBER 1627487-2020-22529; RELATED MANUFACTURER REFERENCE NUMBER 1627487-2020-22531. IT WAS REPORTED THE PATIENT STATED HE EXPERIENCED UNCOMFORTABLE STIMULATION. PATIENT STATED PHYSICIAN IMPLANTED LEADS IN A DIFFERENT LOCATION THAN THE LOCATION USED FOR THE TRIAL AND HE FEELS PAIN. PATIENT SAID HE HAD A LEAD REVISION PROCEDURE (DATE NOT PROVIDED), BUT THE ISSUE PERSISTED. PATIENT DESIRES TO HAVE THE SYSTEM REMOVED. AN ABBOTT REPRESENTATIVE CONTACTED THE PATIENT AND INDICATED THE SYSTEM APPEARS TO BE NONFUNCTIONAL AND THAT THE PATIENT¿S PHYSICIAN DOES NOT PLAN TO REMOVE THE SYSTEM AT THIS TIME DUE TO PATIENT¿S HEALTH CONDITIONS. PATIENT PLANS TO CONSULT WITH ANOTHER PHYSICIAN.
Patient problems: Discomfort; Inadequate Pain Relief; Pain
MAUDE report 10242603 ↗InjuryJun 11, 2020Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure
RELATED MANUFACTURER REPORT NUMBER: 1627487-2020-21508, 1627487-2020-21509, 1627487-2020-21512, 1627487-2020-21513, 1627487-2020-21516, 1627487-2020-21520. IT WAS REPORTED THE PATIENT WAS EXPERIENCING INEFFECTIVE THERAPY. IN TURN, REPROGRAMMING WAS UNABLE TO RESOLVE THE ISSUE. AS A RESULT, THE PATIENT UNDERWENT SURGICAL INTERVENTION DURING WHICH THE SYSTEM WAS EXPLANTED AND REPLACED WITH A COMPETITORS DEVICE.
Patient problems: Inadequate Pain Relief
MAUDE report 10143756 ↗InjuryJun 11, 2020Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure
RELATED MANUFACTURER REPORT NUMBERS: 1627487-2020-21508, 1627487-2020-21511, 1627487-2020-21512, 1627487-2020-21513, 1627487-2020-21516, 1627487-2020-21520. IT WAS REPORTED THE PATIENT WAS EXPERIENCING INEFFECTIVE THERAPY. IN TURN, REPROGRAMMING WAS UNABLE TO RESOLVE THE ISSUE. AS A RESULT, THE PATIENT UNDERWENT SURGICAL INTERVENTION DURING WHICH THE SYSTEM WAS EXPLANTED AND REPLACED WITH A COMPETITORS DEVICE.
Patient problems: Inadequate Pain Relief
MAUDE report 10143746 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P010032. One submission often covers many models, so check the product and lot details in each record.
- Z-0458-2026 · Oct 6, 2025 · Open, ClassifiedEterna SCS IPG (Implantable Pulse Generator), Model Number 32400
- Z-2508-2023 · Jul 18, 2023 · Open, ClassifiedProclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
- Z-2506-2023 · Jul 18, 2023 · Open, ClassifiedProclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
- Z-2510-2023 · Jul 18, 2023 · Open, ClassifiedInfinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
- Z-2511-2023 · Jul 18, 2023 · Open, ClassifiedInfinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
- Z-2507-2023 · Jul 18, 2023 · Open, ClassifiedProclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
- Z-2509-2023 · Jul 18, 2023 · Open, ClassifiedProclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
- Z-2505-2023 · Jul 18, 2023 · Open, ClassifiedProclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
- Z-2691-2017 · Jun 2, 2017 · TerminatedProclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
- Z-2692-2017 · Jun 2, 2017 · TerminatedProclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
- Z-1785-2013 · Nov 30, 2012 · TerminatedEon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
- Z-0025-2013 · Sep 4, 2012 · TerminatedThe Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
- Z-1738-2013 · Jul 26, 2012 · TerminatedEon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
- Z-3091-2011 · Jun 2, 2009 · TerminatedSoftware version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im
Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.