Quattrode® 3153Recalls & reported problems

Advanced Neuromodulation Systems, Inc. · Lead, 3/6mm, 30 cm · Class 3 · Implant

All models of Quattrode (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 05414734401616), received between Jul 11, 2018 and May 30, 2025 — classified as 8 injuries.

The most frequently reported specific device problem is “Fracture” (2 reports).

14 other recalls cover devices cleared under the same FDA submission (P010032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • Fracture2
  • Therapeutic or Diagnostic Output Failure1

Patient problems

  • Inadequate Pain Relief5
  • Failure of Implant2
  • Post Operative Wound Infection2
  • Low Oxygen Saturation1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 30, 2025Fracture; Fracture

    IT WAS REPORTED THAT BOTH OF THE PATIENT'S LEADS WERE VISUALLY CONFIRMED TO BE FRACTURED DURING A PROCEDURE FOR NORMAL EOL IPG REPLACEMENT. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED AND 1 LEAD AND 1 IPG WERE REPLACED.

    Patient problems: Failure of Implant; Inadequate Pain Relief

    MAUDE report 22115032 ↗
  • InjuryMay 30, 2025Fracture; Fracture

    IT WAS REPORTED THAT BOTH OF THE PATIENT'S LEADS WERE VISUALLY CONFIRMED TO BE FRACTURED DURING A PROCEDURE FOR NORMAL EOL IPG REPLACEMENT. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED AND 1 LEAD AND 1 IPG WERE REPLACED.

    Patient problems: Failure of Implant; Inadequate Pain Relief

    MAUDE report 22115024 ↗
  • InjuryOct 5, 2021Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2021-17499. IT WAS REPORTED THAT PATIENT WAS EXPERIENCING INEFFECTIVE THERAPY IN THE RIGHT SIDE AS THE QUATTRODE LEADS WERE UNABLE TO PROVIDE ADEQUATE COVERAGE. TO ADDRESS THE ISSUE PATIENT UNDERWENT SURGICAL INTERVENTION WHEREIN THE QUATTRODE LEADS WERE REPOSITIONED ON (B)(6) 2021. REPORTEDLY EFFECTIVE THERAPY WAS RESTORED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 12579935 ↗
  • InjuryApr 14, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE 3 OF 6. REFERENCE MFR. REPORT#: 1627487-2020-03919. REFERENCE MFR. REPORT#: 1627487-2020-03920. REFERENCE MFR. REPORT#: 1627487-2020-03922. REFERENCE MFR. REPORT#: 1627487-2020-03923. REFERENCE MFR. REPORT#: 1627487-2020-03924. IT WAS REPORTED THE PATIENT'S SCS SYSTEM WAS EXPLANTED DUE TO INFECTION AT THE PATIENT'S IPG AND LEAD SITES. THE INFECTION RESOLVED OVER TIME VIA ANTIBIOTIC INTRAVENOUS THERAPY.

    Patient problems: Post Operative Wound Infection

    MAUDE report 9959990 ↗
  • InjuryApr 14, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE 2 OF 6. REFERENCE MFR. REPORT#: 1627487-2020-03919. REFERENCE MFR. REPORT#: 1627487-2020-03921. REFERENCE MFR. REPORT#: 1627487-2020-03922. REFERENCE MFR. REPORT#: 1627487-2020-03923. REFERENCE MFR. REPORT#: 1627487-2020-03924. IT WAS REPORTED THE PATIENT'S SCS SYSTEM WAS EXPLANTED DUE TO INFECTION AT THE PATIENT'S IPG AND LEAD SITES. THE INFECTION RESOLVED OVER TIME VIA ANTIBIOTIC INTRAVENOUS THERAPY.

    Patient problems: Post Operative Wound Infection

    MAUDE report 9959988 ↗
  • InjuryMay 31, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE 2 OF 2. REFERENCE MFR. REPORT: 1627487-2019-06426. IT WAS REPORTED PATIENT EXPERIENCED INEFFECTIVE COVERAGE OF THERAPY FOR MORE THAN A YEAR APPROXIMATELY (B)(6) 2018. MRI RESULTS SHOWED A PROGRESSION OF THE DISEASE. SURGICAL INTERVENTION MAY TAKE PLACE AT A LATER DATE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 8660474 ↗
  • InjuryMay 31, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE 1 OF 2. REFERENCE MFR. REPORT: 1627487-2019-06429. IT WAS REPORTED PATIENT EXPERIENCED INEFFECTIVE COVERAGE OF THERAPY FOR MORE THAN A YEAR APPROXIMATELY (B)(6) 2018. MRI RESULTS SHOWED A PROGRESSION OF THE DISEASE. SURGICAL INTERVENTION MAY TAKE PLACE AT A LATER DATE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 8660467 ↗
  • InjuryJul 11, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE 2 OF 3. REFERENCE MFR REPORT: 1627487-2018-06209, REFERENCE MFR REPORT: 1627487-2018-06211. IT WAS REPORTED THE PATIENT WAS EXPERIENCING INEFFECTIVE STIMULATION (REFERENCE MFR REPORTS: 1627487-2018-06255, 1627487-2018-06256). AS A RESULT, THE PATIENT UNDERWENT SURGICAL INTERVENTION TO HAVE THE DEVICE EXPLANTED; HOWEVER, THE PATIENT DESATURATED PRIOR TO INCISION AND THE PROCEDURE WAS ABORTED.

    Patient problems: Low Oxygen Saturation

    MAUDE report 7679535 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P010032. One submission often covers many models, so check the product and lot details in each record.

  • Z-0458-2026 · Oct 6, 2025 · Open, Classified
    Eterna SCS IPG (Implantable Pulse Generator), Model Number 32400
  • Z-2507-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2509-2023 · Jul 18, 2023 · Open, Classified
    Proclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
  • Z-2511-2023 · Jul 18, 2023 · Open, Classified
    Infinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2510-2023 · Jul 18, 2023 · Open, Classified
    Infinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2506-2023 · Jul 18, 2023 · Open, Classified
    Proclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
  • Z-2508-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2505-2023 · Jul 18, 2023 · Open, Classified
    Proclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
  • Z-2691-2017 · Jun 2, 2017 · Terminated
    Proclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-2692-2017 · Jun 2, 2017 · Terminated
    Proclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-1785-2013 · Nov 30, 2012 · Terminated
    Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
  • Z-0025-2013 · Sep 4, 2012 · Terminated
    The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
  • Z-1738-2013 · Jul 26, 2012 · Terminated
    Eon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
  • Z-3091-2011 · Jun 2, 2009 · Terminated
    Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im

Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices

All STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page