Octrode™ 3189 (UDI 05414734401715)Recalls & reported problems
St. Jude Medical, Inc. · Lead Kit, 90cm Length · Class 3 · Implant
All models of Octrode (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This St. Jude Medical Octrode lead kit, a spinal cord stimulator lead for pain relief, has 2,236 reports, including 8 deaths and 2,222 injuries. Most describe inadequate pain relief. Many also describe the leads migrating out of place, and some describe repeat surgery to reposition or replace them. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 2,238
- Deaths reported
- 8
- Injuries reported
- 2,225
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2,238 adverse event reports that list this exact device (DI 05414734401715), received between Jun 11, 2018 and Sep 29, 2026 — classified as 8 deaths, 2,225 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Migration” (629 reports).
14 other recalls cover devices cleared under the same FDA submission (P010032); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Migration629
- Adverse Event Without Identified Device or Use Problem590
- High impedance381
- Therapeutic or Diagnostic Output Failure345
- Fracture128
- Impedance Problem53
- Low impedance51
- Inappropriate/Inadequate Shock/Stimulation42
Patient problems
- Inadequate Pain Relief1,627
- No Clinical Signs, Symptoms or Conditions111
- Failure of Implant108
- Undesired Nerve Stimulation77
- No Consequences Or Impact To Patient75
- Erosion63
- Unspecified Infection56
- Post Operative Wound Infection44
Recent adverse event reports
The 25 most recent of 2,238 reports, in the reporter's own words as published by the FDA.
InjurySep 29, 2026Low impedance
It was reported that the patient had low impedances on both leads resulting in ineffective stimulation. Surgical intervention was undertaken on (b)(6) 2026, where the system was explanted and replaced.
Patient problems: Inadequate Pain Relief
MAUDE report 50935447 ↗InjurySep 29, 2026Low impedance
It was reported that the patient had low impedances on both leads resulting in ineffective stimulation. Surgical intervention was undertaken on (b)(6)2026, where the system was explanted and replaced.
Patient problems: Inadequate Pain Relief
MAUDE report 50935442 ↗InjurySep 25, 2026Inappropriate/Inadequate Shock/Stimulation
It was reported that the patient experienced ineffective therapy on the right side of both the thoracic and cervical systems. Additionally, the patient experienced uncomfortable stimulation with the bilateral cervical leads. Reprogramming has been unable to resolve the issue. As a result, surgical intervention was undertaken wherein the bilateral cervical leads were replaced to address the issue. Surgical intervention may be undertaken for the thoracic right lead in the future.
Patient problems: Undesired Nerve Stimulation
MAUDE report 50899054 ↗InjurySep 25, 2026Therapeutic or Diagnostic Output Failure
It was reported that the patient experienced ineffective therapy on the right side of both the thoracic and cervical systems. Additionally, the patient experienced uncomfortable stimulation with the bilateral cervical leads. Reprogramming has been unable to resolve the issue. As a result, surgical intervention was undertaken wherein the bilateral cervical leads were replaced to address the issue. Surgical intervention may be undertaken for the thoracic right lead in the future.
Patient problems: Inadequate Pain Relief
MAUDE report 50899051 ↗InjurySep 23, 2026Therapeutic or Diagnostic Output Failure
It was reported that the patient experienced ineffective therapy. Reprogramming has been unable to resolve the issue. As a result, surgical intervention was undertaken wherein system was explanted to address the issue.
Patient problems: Inadequate Pain Relief
MAUDE report 50872160 ↗InjurySep 18, 2026Material Fragmentation
It was reported that during the lead revision procedure on (b)(6) 2026 (Related Manufacturer Reference Number: 1627487-2026-10156) the lead was cut and a portion of the lead remains implanted and connected to the IPG. As such, surgical intervention may take place at a later date to explant the remaining portion of the lead.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50829590 ↗InjurySep 18, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient experienced inflammation (swelling/redness) at the lead site. As such, surgical intervention took place (b)(6) 2026 wherein the doctor cut and removed the portion of the lead that was under the affected area. The remaining portion of the lead remains implanted and connected to the IPG(Related Manufacturer Reference Number: 1627487-2026-10157).
Patient problems: Inflammation; Swelling/ Edema
MAUDE report 50829586 ↗InjurySep 18, 2026High impedance
It was reported that the patient experienced ineffective therapy. Impedance check revealed high impedances on all contacts of the lead. As such, surgical intervention took place wherein the lead was explanted and replaced to address the issue. Reportedly, therapy was restored post op.
Patient problems: Inadequate Pain Relief
MAUDE report 50828897 ↗InjurySep 14, 2026Therapeutic or Diagnostic Output Failure
It was reported that the patient is experiencing inadequate coverage/therapy. Additionally, patient is experiencing pain at the IPG site following weight loss. As such, surgical intervention may take place at later date to address the issue.
Patient problems: Inadequate Pain Relief
MAUDE report 50756969 ↗InjurySep 14, 2026Therapeutic or Diagnostic Output Failure
It was reported that the patient is experiencing inadequate coverage/therapy. Additionally, patient is experiencing pain at the IPG site following weight loss. As such, surgical intervention may take place at later date to address the issue.
Patient problems: Inadequate Pain Relief
MAUDE report 50756957 ↗InjurySep 3, 2026High impedance
It was reported that the patient experienced ineffective therapy. Diagnostics revealed high impedances on both leads. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the leads were explanted and replaced to address the issue.
Patient problems: Inadequate Pain Relief
MAUDE report 50587101 ↗InjuryAug 24, 2026Impedance Problem
It was reported that the patient had the Abbott IPG and one lead replaced at an unknown date with a competitor product. Recently, the system, including the one remaining Abbott lead was found to have high impedances. As a result, surgical intervention was undertaken on (b)(6) 2026 where the system was removed and replaced to address the issue.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50421047 ↗InjuryAug 24, 2026Insufficient Device Problem Information
It was reported that the patient had the Abbott IPG and one lead replaced at an unknown date with a competitor product. Recently, the system, including the one remaining Abbott lead was found to have high impedances. As a result, surgical intervention was undertaken on (b)(6) 2026 where the system was removed and replaced to address the issue.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50421043 ↗InjuryAug 23, 2026Therapeutic or Diagnostic Output Failure
It was reported that following recent weight loss, the patient experienced pain at the IPG site. Additionally, the patient experienced ineffective therapy. As a result, surgical intervention was undertaken wherein the system was explanted to address the issue. It is unknown which lead caused ineffective therapy.
Patient problems: Inadequate Pain Relief
MAUDE report 50396288 ↗InjuryAug 11, 2026Migration
It was reported that the patient experienced ineffective therapy following unknown trauma. X-ray imaging revealed migration of both leads. As a result, surgical intervention may be undertaken in the future.
Patient problems: Inadequate Pain Relief
MAUDE report 50211089 ↗InjuryJul 31, 2026Migration
It was reported that the patient experienced ineffective therapy. X-ray imaging revealed migration of both leads. As a result, surgical intervention may be undertaken in the future.
Patient problems: Inadequate Pain Relief
MAUDE report 50104965 ↗InjuryJul 30, 2026Migration
It was reported that the patient was not receiving effective therapy due to lead migration. The issue was confirmed via X-Rays. Surgical intervention was undertaken on (b)(6) 2026 wherein the lead was repositioned to address the issue. Therapy was restored post-surgery.
Patient problems: Inadequate Pain Relief
MAUDE report 50090462 ↗InjuryJul 28, 2026Migration
It was reported patient's leads had migrated causing patient ineffective therapy. Surgical intervention was undertaken wherein the leads were explanted and replaced to address the issue. Therapy was restored post-op.
Patient problems: Inadequate Pain Relief
MAUDE report 50055344 ↗InjuryJul 28, 2026Migration
It was reported patient's leads had migrated causing patient ineffective therapy. Surgical intervention was undertaken wherein the leads were explanted and replaced to address the issue. Therapy was restored post-op.
Patient problems: Inadequate Pain Relief
MAUDE report 50055309 ↗InjuryJul 15, 2026Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY OF THEIR CERVICAL SYSTEM. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE. IT IS UNKNOWN WHICH LEAD IS RESPONSIBLE FOR INEFFECTIVE STIM.
Patient problems: Inadequate Pain Relief
MAUDE report 25834169 ↗InjuryJun 29, 2026Low impedance
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY/COVERAGE. IMPEDANCE CHECK REVEALED LOW IMPEDANCE ON ONE THE LEADS. REPROGRAMMING WAS UNABLE TO RESOLVE THE ISSUE. AS SUCH, SURGICAL INTERVENTION MAY TAKE PLACE AT A LATER DATE TO ADDRESS THE ISSUE. INVESTIGATION WAS UNABLE TO DETERMINE WHICH OF THE LEAD HAD LOW IMPEDANCE.
Patient problems: Inadequate Pain Relief
MAUDE report 25657160 ↗InjuryJun 28, 2026High impedance
IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. FURTHER INVESTIGATION CONFIRMED THAT ONE OF THE PATIENT¿S LEADS EXHIBITED HIGH IMPEDANCE WHILE THE OTHER SHOWED LOW IMPEDANCE, PREVENTING THE PATIENT FROM UTILIZING MRI MODE. AS A RESULT, THE PHYSICIAN REPLACED THE SYSTEM TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 25648537 ↗InjuryJun 25, 2026Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AT THE IPG SITE, INEFFECTIVE THERAPY, AND SHOCKING SENSATION. SURGICAL INTERVENTION WAS UNDERTAKEN IN OCT 2023 WHEREIN THE IPG WAS REPOSITIONED; HOWEVER, THE ISSUES PERSISTED. AS A RESULT, THE PATIENT UNDERWENT ANOTHER SURGICAL INTERVENTION WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED ON AN UNKNOWN DATE.
Patient problems: Inadequate Pain Relief
MAUDE report 25627390 ↗InjuryJun 19, 2026Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. AS SUCH, SUCH SURGICAL INTERVENTION TOOK PLACE IN 2024 WHEREIN THE SYSTEM WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. EXACT DATE OF EXPLANT IS UNKNOWN. INVESTIGATION WAS UNABLE TO DETERMINE WHICH OF THE LEADS ATTRIBUTED TO THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 25557172 ↗InjuryJun 16, 2026Device Misassembled During Manufacturing /Shipping
IT WAS REPORTED THAT DURING A TRIAL WHEREIN PATIENT WAS IMPLANTED WITH 3189 LEADS, DURING THE PROCEDURE IT WAS DISCOVERED THE LEADS ONLY HAD 4 CONTACTS. DEVICE PACKAGES WERE RECEIVED CONTAINING THE INCORRECT PRODUCT. PATIENT DID NOT EXPERIENCE ANY ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25522651 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P010032. One submission often covers many models, so check the product and lot details in each record.
- Z-0458-2026 · Oct 6, 2025 · Open, ClassifiedEterna SCS IPG (Implantable Pulse Generator), Model Number 32400
- Z-2511-2023 · Jul 18, 2023 · Open, ClassifiedInfinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
- Z-2510-2023 · Jul 18, 2023 · Open, ClassifiedInfinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
- Z-2505-2023 · Jul 18, 2023 · Open, ClassifiedProclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
- Z-2506-2023 · Jul 18, 2023 · Open, ClassifiedProclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
- Z-2509-2023 · Jul 18, 2023 · Open, ClassifiedProclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
- Z-2508-2023 · Jul 18, 2023 · Open, ClassifiedProclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
- Z-2507-2023 · Jul 18, 2023 · Open, ClassifiedProclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
- Z-2692-2017 · Jun 2, 2017 · TerminatedProclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
- Z-2691-2017 · Jun 2, 2017 · TerminatedProclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
- Z-1785-2013 · Nov 30, 2012 · TerminatedEon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
- Z-0025-2013 · Sep 4, 2012 · TerminatedThe Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
- Z-1738-2013 · Jul 26, 2012 · TerminatedEon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
- Z-3091-2011 · Jun 2, 2009 · TerminatedSoftware version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im
Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page