Lamitrode Tripole™ 3219 (UDI 05414734401821)Recalls & reported problems
Advanced Neuromodulation Systems, Inc. · 16 Lead, 60 cm · Class 3 · Implant
Quick summary
This Lamitrode Tripole spinal cord stimulator lead appears in 268 adverse event reports, including 1 death and 266 injuries. The reports mostly describe patients getting too little pain relief, with testing often showing high readings on the lead, lead movement or breaks, and several patients needing surgery to replace or remove the lead. Some also describe unwanted nerve stimulation and tingling. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 270
- Deaths reported
- 1
- Injuries reported
- 268
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 270 adverse event reports that list this exact device (DI 05414734401821), received between Jun 26, 2018 and Sep 29, 2026 — classified as 1 death, 268 injuries, 1 malfunction.
The most frequently reported specific device problem is “Therapeutic or Diagnostic Output Failure” (74 reports).
14 other recalls cover devices cleared under the same FDA submission (P010032); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Therapeutic or Diagnostic Output Failure74
- High impedance54
- Adverse Event Without Identified Device or Use Problem52
- Migration30
- Fracture12
- Inappropriate/Inadequate Shock/Stimulation12
- Insufficient Device Problem Information10
- Use of Device Problem9
Patient problems
- Inadequate Pain Relief172
- No Clinical Signs, Symptoms or Conditions27
- Undesired Nerve Stimulation16
- Failure of Implant15
- Implant Pain9
- No Consequences Or Impact To Patient9
- Hematoma8
- Post Operative Wound Infection6
Recent adverse event reports
The 25 most recent of 270 reports, in the reporter's own words as published by the FDA.
InjurySep 29, 2026Inappropriate/Inadequate Shock/Stimulation
It was reported the patient experienced shocking. As a result the system was explanted to address the issue.
Patient problems: Undesired Nerve Stimulation
MAUDE report 50932099 ↗InjurySep 22, 2026High impedance
It was reported that during the Normal End of Life IPG replacement on (b)(6) 2026 impedances were found on the existing lead. As a result, the lead was also replaced to address the issue.
Patient problems: Inadequate Pain Relief
MAUDE report 50858832 ↗InjuryJul 31, 2026High impedance
Additional information received indicates that patient was unable to set the system into MRI mode due to the high impedance.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50100411 ↗InjuryJul 28, 2026Therapeutic or Diagnostic Output Failure
It was reported that the patient experienced ineffective therapy. As a result, surgical intervention may be undertaken to address the issue.
Patient problems: Inadequate Pain Relief
MAUDE report 50063823 ↗InjuryJul 27, 2026Therapeutic or Diagnostic Output Failure
It was reported that following recent weight loss, the patient experienced pain and an itching sensation at the IPG site. Additionally, the patient experienced ineffective therapy. As a result, surgical intervention may be undertaken to address the issue.
Patient problems: Inadequate Pain Relief
MAUDE report 50047706 ↗InjuryJul 27, 2026High impedance
It was reported that the patient experienced ineffective stimulation due to high impedances. As a result, surgical intervention was undertaken on (b)(6)2026 wherein the lead was explanted and replaced to address the issue.
Patient problems: Inadequate Pain Relief
MAUDE report 50024496 ↗InjuryJun 11, 2026Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON AN UNKNOWN DATE WHEREIN THE SYSTEM WAS EXPLANTED.
Patient problems: Inadequate Pain Relief
MAUDE report 25474485 ↗InjuryJun 5, 2026Fracture
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. LEAD FRACTURE WAS CONFIRMED VIA IMAGING. AS SUCH, SURGICAL INTERVENTION TOOK PLACE WHEREIN THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. EFFECTIVE THERAPY WAS CONFIRMED POST OP.
Patient problems: Inadequate Pain Relief
MAUDE report 25412926 ↗InjuryMay 8, 2026Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. REPROGRAMMING HAS BEEN UNABLE TO RESOLVE THE ISSUE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 25124591 ↗InjuryApr 15, 2026Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED PATIENT WAS FEELING TINGLING IN THEIR STOMACH THAT WAS MAKING THEM SICK WHEN THERAPY WAS ON. PROGRAMMING WAS NOT ABLE TO RESOLVE THE ISSUE AND THERAPY WAS TURNED OFF. NEXT COURSE OF ACTION IS UNDETERMINED AT THIS TIME.
Patient problems: Undesired Nerve Stimulation
MAUDE report 24891883 ↗InjuryApr 6, 2026Patient-Device Incompatibility; No Apparent Adverse Event
ADDITIONAL INFORMATION INDICATES THE LEADS WERE REPLACED ELECTIVELY FOR MRI CAPABILITY. THERE IS NO DEVICE MALFUNCTION; THEREFORE, THIS DEVICE IS NO LONGER CONSIDERED REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24791936 ↗InjuryMar 25, 2026Material Twisted/Bent
IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AT THE LEAD SIDE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEAD WAS TUNNELED DEEPER. DURING THE PROCEDURE IT WAS NOTED THAT THE LEAD WAS COILED UP. EFFECTIVE THERAPY WAS RESTORED.
Patient problems: Implant Pain
MAUDE report 24688694 ↗InjuryMar 24, 2026High impedance
IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION DUE TO HIGH IMPEDANCES. SURGICAL INTERVENTION TOOK PLACE WHEREIN THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. THERAPY WAS RESTORED.
Patient problems: Inadequate Pain Relief
MAUDE report 24683207 ↗InjuryMar 3, 2026High impedance
IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY, PATIENTS LEAD HAD HIGH IMPEDANCES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 24500919 ↗InjuryMar 2, 2026Migration
IT WAS REPORTED THAT FOLLOWING A CAR ACCIDENT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. X-RAY IMAGING REVEALED MIGRATION OF THE LEAD. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE LEAD WAS REPLACED TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 24489730 ↗InjuryMar 2, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT¿S IPG AND LEAD WERE EXPLANTED. THE DATE OF EXPLANT AND REASON FOR EXPLANT ARE UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 24483935 ↗InjuryFeb 20, 2026Use of Device Problem
IT WAS REPORTED THE PATIENT EXPERIENCED COMPLICATIONS AT THE IPG POCKET SITE. THE IPG WAS EXPLANTED AND THE PADDLE LEAD WAS CUT AND PARTIAL REMAINED IMPLANTED. DATE OF SURGERY AND COMPLICATIONS ARE UNKNOWN. ON (B)(6) 2026 THE PHYSICIAN REMOVED THE CUT LEAD AND REPLACED THE SYSTEM. EFFECTIVE THERAPY WAS ESTABLISHED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 24400186 ↗InjuryFeb 16, 2026Use of Device Problem
IT WAS REPORTED THAT DURING A SYSTEM EXPLANT PROCEDURE ON (B)(6) 2016 [RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2016-04418], THE LEAD WAS CUT AND A PORTION OF THE LEAD REMAINS IMPLANTED. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 24348617 ↗InjuryJan 13, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AT THE ANCHOR SITE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE ANCHORS WERE EXPLANTED TO ADDRESS THE ISSUE. THERAPY WAS RESTORED POST-OPERATIVELY.
Patient problems: Implant Pain
MAUDE report 24059631 ↗InjuryJan 7, 2026High impedance
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. DIAGNOSTICS REVEALED HIGH IMPEDANCES. REPROGRAMMING HAS BEEN UNABLE TO RESOLVE THE ISSUE. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 23995476 ↗InjuryDec 29, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT DURING A LEAD REVISION PROCEDURE ON (B)(6) 2025, PATIENTS LEAD WAS CUT, AND THE REMAINING PORTION OF THE LEAD REMAINS IMPLANTED. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN ON A LATER DATE TO REMOVE REMAINING PORTION OF THE LEAD.
Patient problems: Foreign Body In Patient
MAUDE report 23908683 ↗InjuryDec 24, 2025Use of Device Problem; Use of Device Problem
ADDITIONAL INFORMATION WAS RECEIVED INDICATING THAT SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026, WHEREIN THE REMAINING PORTION OF THE LEAD WAS EXPLANTED TO ADDRESS THE ISSUE.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 23890746 ↗InjuryDec 18, 2025High impedance; High impedance
IT WAS REPORTED PATIENT'S LEAD HAS INVALID IMPEDANCES. SURGICAL INTERVENTION MAY BE PENDING TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23839370 ↗InjuryDec 17, 2025High impedance; High impedance
IT WAS REPORTED THAT THE PATIENT'S LEAD EXHIBITED HIGH IMPEDANCES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. EFFECTIVE THERAPY WAS RESTORED POST OPERATIVELY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23832687 ↗InjuryDec 15, 2025High impedance; High impedance
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. TROUBLESHOOTING REVEALED HIGH IMPEDANCES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. EFFECTIVE THERAPY WAS RESTORED POST OPERATIVELY.
Patient problems: Inadequate Pain Relief
MAUDE report 23811376 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P010032. One submission often covers many models, so check the product and lot details in each record.
- Z-0458-2026 · Oct 6, 2025 · Open, ClassifiedEterna SCS IPG (Implantable Pulse Generator), Model Number 32400
- Z-2510-2023 · Jul 18, 2023 · Open, ClassifiedInfinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
- Z-2505-2023 · Jul 18, 2023 · Open, ClassifiedProclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
- Z-2507-2023 · Jul 18, 2023 · Open, ClassifiedProclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
- Z-2509-2023 · Jul 18, 2023 · Open, ClassifiedProclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
- Z-2506-2023 · Jul 18, 2023 · Open, ClassifiedProclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
- Z-2508-2023 · Jul 18, 2023 · Open, ClassifiedProclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
- Z-2511-2023 · Jul 18, 2023 · Open, ClassifiedInfinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
- Z-2692-2017 · Jun 2, 2017 · TerminatedProclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
- Z-2691-2017 · Jun 2, 2017 · TerminatedProclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
- Z-1785-2013 · Nov 30, 2012 · TerminatedEon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
- Z-0025-2013 · Sep 4, 2012 · TerminatedThe Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
- Z-1738-2013 · Jul 26, 2012 · TerminatedEon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
- Z-3091-2011 · Jun 2, 2009 · TerminatedSoftware version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im
Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.