Exclaim™ 3225 (UDI 05414734401883)Recalls & reported problems

St. Jude Medical, Inc. · Lead, 90 cm · Class 3 · Implant

Quick summary

This St. Jude Medical Exclaim spinal cord stimulator lead is an implanted lead used for pain relief. Reports most often describe ineffective therapy, high electrical resistance in the lead, and leads migrating, with some patients needing surgery to reposition or replace them. The database holds 243 reports, including 1 death and 237 injuries, with 5 malfunctions. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
246
Deaths reported
1
Injuries reported
240
Recalls naming it
0

Summary

The FDA's MAUDE database contains 246 adverse event reports that list this exact device (DI 05414734401883), received between Jun 12, 2018 and Sep 30, 2026 — classified as 1 death, 240 injuries, 5 malfunctions.

The most frequently reported specific device problem is “High impedance” (77 reports).

14 other recalls cover devices cleared under the same FDA submission (K063080, P010032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 0 reportsJan 2024: 3 reportsFeb 2024: 3 reportsMar 2024: 1 reportsApr 2024: 6 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 3 reportsAug 2024: 3 reportsSep 2024: 4 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 5 reportsJul 2025: 3 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 4 reportsFeb 2026: 4 reportsMar 2026: 1 reportsApr 2026: 3 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High impedance77
  • Therapeutic or Diagnostic Output Failure48
  • Adverse Event Without Identified Device or Use Problem47
  • Migration42
  • Fracture18
  • Impedance Problem9
  • Failure to Deliver Energy5
  • Inappropriate/Inadequate Shock/Stimulation5

Patient problems

  • Inadequate Pain Relief195
  • Failure of Implant13
  • Post Operative Wound Infection9
  • No Clinical Signs, Symptoms or Conditions8
  • Undesired Nerve Stimulation5
  • Pain4
  • Unspecified Infection4
  • Numbness3

Recent adverse event reports

The 25 most recent of 246 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Therapeutic or Diagnostic Output Failure

    It was reported that the patient reported that the lead placed as trial to perm lead required an extended hospital stay during the trail and now does not have effective therapy as the lead has migrated and system is not working at this time. Patient reports no communication, but it is not known if the charging was maintained. As a result, surgical intervention may be taken at a later date.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50948561 ↗
  • InjurySep 25, 2026Inappropriate/Inadequate Shock/Stimulation

    It was reported that the patient experienced shocking sensations. Evaluation of the system indicated high impedances on multiple contacts. As a result, surgical intervention may be undertaken to address the issue. It is unknown which lead is related to this issue.

    Patient problems: Burning Sensation

    MAUDE report 50901218 ↗
  • InjurySep 18, 2026Therapeutic or Diagnostic Output Failure

    It was reported that the patient experienced ineffective therapy. Reprogramming has been unable to resolve the issue. As a result, surgical intervention was undertaken wherein system was explanted to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50813435 ↗
  • InjuryAug 5, 2026Therapeutic or Diagnostic Output Failure

    It was reported that the patient experienced ineffective therapy, shocking sensations, numbness in the extremities, vasovagal syncope, cardiac event and atrial fibrillation. As a result, surgical intervention was undertaken on an unknown date wherein the entire system was explanted to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50147495 ↗
  • InjuryAug 3, 2026Therapeutic or Diagnostic Output Failure

    It was reported that the patient experienced ineffective therapy. As a result, surgical intervention was undertaken on an unknown date wherein the entire system was explanted to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50122451 ↗
  • InjuryJun 16, 2026High impedance

    IT WAS REPORTED THAT THE PATIENT'S CERVICAL LEAD EXHIBITED HIGH IMPEDANCES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN IN (B)(6) 2025 WHEREIN THE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE. AFTER EXPLANT, IT WAS REPORTED THE PATIENT EXPERIENCED AN INFECTION AT LEAD INCISION SITE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 25527839 ↗
  • InjuryJun 3, 2026High impedance

    IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY, PATIENTS LEAD HAS HIGH IMPEDANCES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 25384249 ↗
  • InjuryApr 16, 2026Migration

    IT WAS REPORTED THAT THE PATIENT EXPERIENCE INEFFECTIVE THERAPY. PATIENT¿S LEAD HAD MIGRATED. AS SUCH, SURGICAL INTERVENTION TOOK PLACE ON (B)(6) 2026 WHEREIN THE LEAD WAS REPOSITIONED TO ADDRESS THE ISSUE. THERAPY WAS RESTORED POST OP.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24905592 ↗
  • InjuryApr 12, 2026High impedance

    IT WAS REPORTED THAT THE PATIENT REPORTED INEFFECTIVE STIMULATION. DIAGNOSTIC REPORTS INDICATED THE LEAD HAD HIGH IMPEDANCES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE LEAD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24853014 ↗
  • InjuryApr 3, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24779205 ↗
  • InjuryMar 4, 2026High impedance; No Apparent Adverse Event

    IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY, PATIENTS LEAD HAS HIGH IMPEDANCES. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24514267 ↗
  • InjuryFeb 25, 2026Malposition of Device

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION DUE TO MALPOSITION OF THE DEVICE. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24445287 ↗
  • InjuryFeb 19, 2026High impedance

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY DUE TO HIGH IMPEDANCES ON BOTH LEADS. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24393509 ↗
  • InjuryFeb 19, 2026High impedance

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY DUE TO HIGH IMPEDANCES ON BOTH LEADS. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24393503 ↗
  • InjuryFeb 13, 2026High impedance

    ADDITIONAL INFORMATION RECEIVED INDICATES SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEAD WAS EXPLANTED AND REPLACED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24339843 ↗
  • InjuryJan 27, 2026High impedance

    IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION DUE TO HIGH IMPEDANCES. SURGICAL INTERVENTION TOOK PLACE WHEREIN THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24196368 ↗
  • InjuryJan 26, 2026Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED UNCOMFORTABLE STIMULATION. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.

    Patient problems: Undesired Nerve Stimulation

    MAUDE report 24185172 ↗
  • InjuryJan 20, 2026Use of Device Problem

    IT WAS REPORTED THAT DURING A PERM IMPLANT PROCEDURE ON (B)(6) 2025 THE PATIENT¿S LEAD WAS CUT IN ACCIDENT AND THE CASE WAS ABANDONED. AS SUCH, SURGICAL INTERVENTION TOOK PLACE ON (B)(6) 2025 WHEREIN THE SEVERED LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24123745 ↗
  • InjuryJan 12, 2026Fracture

    IT WAS REPORTED THAT THE PATIENT¿S IPG HAD REACHED END OF LIFE. IT IS UNKNOWN IF THIS OCCURRED PREMATURELY. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEAD WAS CONFIRMED FRACTURED. AS A RESULT, THE IPG AND LEAD WERE EXPLANTED AND REPLACED DURING THE PROCEDURE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24047140 ↗
  • InjuryDec 28, 2025High impedance; High impedance

    IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY. DIAGNOSTICS REVEALED HIGH IMPEDANCES ON ONE LEAD. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE LEADS AND ANCHORS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 23904038 ↗
  • InjuryDec 17, 2025Impedance Problem; Impedance Problem

    IT WAS REPORTED THAT THE PATIENT'S LEADS HAD IMPEDANCE ISSUES. SURGICAL INTERVENTION WAS UNDERTAKEN, AND THE LEADS WERE EXPLANTED AND REPLACED. THE IPG WAS REPLACED FOR AN UNKNOWN REASON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23835235 ↗
  • InjuryNov 17, 2025Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED SHOCKING SENSATIONS AND INEFFECTIVE THERAPY DUE TO HIGH IMPEDANCES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE LEAD WAS REPLACED TO ADDRESS THE ISSUE. THERAPY WAS RESTORED POST-OPERATIVELY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 23576649 ↗
  • InjuryNov 7, 2025High impedance; High impedance

    ADDITIONAL INFORMATION INDICATES SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025, WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 23504018 ↗
  • InjuryOct 24, 2025Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION AND THE IPG WAS UNABLE TO COMMUNICATE WITH THE EXTERNAL DEVICES. SURGICAL INTERVENTION WAS UNDERTAKEN AND THE SYSTEM WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 23377988 ↗
  • InjurySep 12, 2025Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON AN UNKNOWN DATE WHEREIN THE LEAD WAS LEFT IMPLANTED, THE IPG WAS EXPLANTED, AND THE SYSTEM WAS REPLACED WITH A COMPETITOR'S SYSTEM TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 23034883 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K063080, P010032. One submission often covers many models, so check the product and lot details in each record.

  • Z-0458-2026 · Oct 6, 2025 · Open, Classified
    Eterna SCS IPG (Implantable Pulse Generator), Model Number 32400
  • Z-2506-2023 · Jul 18, 2023 · Open, Classified
    Proclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
  • Z-2509-2023 · Jul 18, 2023 · Open, Classified
    Proclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
  • Z-2507-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2510-2023 · Jul 18, 2023 · Open, Classified
    Infinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2511-2023 · Jul 18, 2023 · Open, Classified
    Infinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2508-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2505-2023 · Jul 18, 2023 · Open, Classified
    Proclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
  • Z-2692-2017 · Jun 2, 2017 · Terminated
    Proclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-2691-2017 · Jun 2, 2017 · Terminated
    Proclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-1785-2013 · Nov 30, 2012 · Terminated
    Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
  • Z-0025-2013 · Sep 4, 2012 · Terminated
    The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
  • Z-1738-2013 · Jul 26, 2012 · Terminated
    Eon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
  • Z-3091-2011 · Jun 2, 2009 · Terminated
    Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im

Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices

All STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.