Lamitrode® 3288Recalls & reported problems

Advanced Neuromodulation Systems, Inc. · 88 Lead Kit, 60cm Length · Class 3 · Implant

Quick summary

This Lamitrode 3288 is an implanted spinal cord stimulator lead kit used for pain relief. The 58 reports mostly describe leads that stop working well, with high impedance readings and lost stimulation, so patients got little pain relief; several had surgery to remove or replace the lead and generator. There are 57 injuries and no deaths reported, along with one malfunction. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
58
Deaths reported
0
Injuries reported
57
Recalls naming it
0

Summary

The FDA's MAUDE database contains 58 adverse event reports that list this exact device (DI 05414734402248), received between Sep 26, 2018 and Jan 12, 2026 — classified as 57 injuries, 1 malfunction.

The most frequently reported specific device problem is “Therapeutic or Diagnostic Output Failure” (17 reports).

14 other recalls cover devices cleared under the same FDA submission (P010032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Therapeutic or Diagnostic Output Failure17
  • High impedance12
  • Adverse Event Without Identified Device or Use Problem10
  • Disconnection4
  • Insufficient Device Problem Information4
  • Impedance Problem3
  • Migration3
  • Fracture2

Patient problems

  • Inadequate Pain Relief40
  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient3
  • Hematoma2
  • Post Operative Wound Infection2
  • Unspecified Infection2
  • Bacterial Infection1
  • Cerebrospinal Fluid Leakage1

Recent adverse event reports

The 25 most recent of 58 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 12, 2026High impedance

    IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY, PATIENTS LEAD HAD HIGH IMPEDANCES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. PATIENTS IPG WAS ALSO EXPLANTED AND REPLACED FOR AN UNKNOWN REASON.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24048398 ↗
  • InjuryDec 5, 2025High impedance; High impedance

    IT WAS REPORTED PATIENT WAS UNABLE TO SET THE IPG INTO MRI MODE DUE TO HIGH IMPEDANCE ON ONE OF THE LEADS, IT IS UNKNOWN WHICH LEAD ATTRIBUTED TO THE ISSUE. SURGICAL INTERVENTION MAY BE PENDING TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23730328 ↗
  • InjuryJul 24, 2025High impedance; High impedance

    IT WAS REPORTED THAT FOLLOWING JIU JITSU PATIENT EXPERIENCED INEFFECTIVE THERAPY, PATIENTS LEAD HAS HIGH IMPEDANCES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN PATIENTS' SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 22594290 ↗
  • InjuryJul 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING IMPLANT PROCEDURE ON (B)(6) 2025, THE PATIENT SUSTAINED A DURAL TEAR. AS A RESULT, THE TEAR WAS SUTURED DURING THE PROCEDURE TO ADDRESS THE ISSUE. THE PATIENT EXPERIENCED A HEADACHE POST OP. PATIENTS IS STABLE AND SYMPTOMS HAVE RESOLVED.

    Patient problems: Cerebrospinal Fluid Leakage; Headache

    MAUDE report 22487175 ↗
  • InjuryDec 30, 2024High impedance; Impedance Problem; High impedance; Impedance Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION DUE TO HIGH IMPEDANCES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2024 THAT IDENTIFIED NO CONNECTION ISSUE. NO INTERVENTION WAS UNDERTAKEN DURING THE PROCEDURE WITH THE DEVICES AND ALL DEVICES REMAIN IMPLANTED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 21044643 ↗
  • InjuryDec 5, 2024Impedance Problem; Impedance Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. IMPEDANCE CHECK REVEALED HIGH AND LOW IMPEDANCES. AS SUCH, SURGICAL INTERVENTION TOOK PLACE WHEREIN THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. REPORTEDLY, EFFECTIVE THERAPY WAS CONFIRMED POST OP.

    Patient problems: Failure of Implant; Inadequate Pain Relief

    MAUDE report 20856742 ↗
  • InjuryJul 22, 2024Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INEFFECTIVE THERAPY AND DISCOMFORT AT THE IPG SITE. AS A RESULT, THE PATIENT UNDERWENT SURGICAL INTERVENTION DURING WHICH THE IPG AND LEAD WERE EXPLANTED AND REPLACED. EFFECTIVE THERAPY WAS ESTABLISHED POST-OPERATIVELY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 19796928 ↗
  • InjuryMar 29, 2024Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION WITH THE SCS SYSTEM. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2024 WHEREIN THE SCS IPG WAS EXPLANTED AND A DRG SYSTEM WAS IMPLANTED. THE SCS LEAD REMAINS IMPLANTED BUT INACTIVE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 19004315 ↗
  • InjuryFeb 6, 2024High impedance; High impedance

    RELATED MANUFACTURER REFERENCE NUMBER 3006705815-2024-00879. IT WAS REPORTED THE PATIENT IS NOT RECEIVING EFFECTIVE THERAPY DUE TO HIGH IMPEDANCES. AS SUCH, SURGICAL INTERVENTION WAS TAKEN AND THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 18651527 ↗
  • InjuryNov 1, 2023High impedance; High impedance

    IT WAS REPORTED THAT THE PATIENT'S LEAD HAD HIGH IMPEDANCES. SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2023 WHEREIN THE LAMITRODE LEAD WAS EXPLANTED AND REPLACED WITH A PENTA LEAD TO ADDRESS THE ISSUE. THERAPY WAS RESTORED POST PROCEDURE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 18049473 ↗
  • InjuryJul 11, 2023Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    ADDITIONAL INFORMATION RECEIVED IDENTIFIED THAT PATIENT UNDERWENT SURGICAL INTERVENTION ON (B)(6) 2023 WHEREIN, THE LEAD WAS EXPLANTED AND REPLACED, THE IPG WAS ALSO RELOCATED. EFFECTIVE THERAPY WAS RESTORED POST OPERATIVELY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 17302587 ↗
  • InjuryMay 31, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED INDICATES THAT THE DOCTOR HAS INDICATED THAT THE INFECTION HAS CLEARED AND IV ANTIBIOTICS HAVE BEEN STOPPED.

    Patient problems: Post Operative Wound Infection

    MAUDE report 17035008 ↗
  • InjuryApr 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT 72 HOURS FOLLOWING A PERMANENT IMPLANT PROCEDURE, THE PATIENT EXPERIENCED AN EXTRADURAL HEMATOMA AFTER STRAINING TO EVACUATE THEIR BOWELS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2023 WHERE IN THE ENTIRE SYSTEM WAS EXPLANTED. CURRENTLY, THE PATIENT HAS COMPLETE LOSS OF LOWER LIMB MOTOR FUNCTION AND SIGNIFICANT LOSS OF SENSORY FUNCTION.

    Patient problems: Hematoma; Paralysis

    MAUDE report 16710657 ↗
  • InjuryApr 3, 2023Migration; Migration

    RELATED MANUFACTURER REFERENCE NUMBER 3006705815-2023-01812. IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY DUE TO LEAD MIGRATION. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEAD WAS EXPLANTED AND REPLACED. THE IPG FAILED TO ESTABLISH COMMUNICATION WITH EXTERNAL DEVICES AFTER THE PROCEDURE AND WAS EXPLANTED AND REPLACED. EFFECTIVE THERAPY WAS RESTORED POST OPERATIVELY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 16665414 ↗
  • InjuryFeb 9, 2023Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THE PATIENT WAS EXPERIENCING SHOCKING SENSATION. AS SUCH, SURGICAL INTERVENTION WAS UNDERTAKEN WHERE THE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Undesired Nerve Stimulation

    MAUDE report 16333943 ↗
  • InjuryDec 18, 2022Disconnection; Migration; Disconnection; Migration

    RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2022-06898. IT WAS REPORTED THAT X-RAYS SHOWED DISCONNECTION BETWEEN THE LEAD AND IPG. AS SUCH, THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. A MESSAGE WAS RECEIVED STATING THE IS AN OPEN CIRCUIT/HIGH IMPEDANCES. HOWEVER, IMPEDANCE CHECK REVELATED THAT THE IMPEDANCE IS IN NORMAL RANGE. SURGICAL INTERVENTION MAY TAKE PLACE AT A LATER DATE TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 16001459 ↗
  • InjuryDec 7, 2022Material Fragmentation; Use of Device Problem; Material Fragmentation; Use of Device Problem

    RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2022-06689. IT WAS REPORTED THAT DURING AN IPG EXPLANT PROCEDURE THE PHYSICIAN CUT THE LEADS. THE CUT PORTION OF THE LEADS REMAIN IMPLANTED IN THE PATIENT. SURGICAL INTERVENTION MAY BE PENDING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15934106 ↗
  • InjuryNov 21, 2022High impedance; High impedance

    IT WAS REPORTED THE PATIENTS LEAD DISPLAYED HIGH IMPEDANCE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 15833316 ↗
  • InjuryOct 4, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RELATED MANUFACTURER REFERENCE NUMBER:1627487-2022-05457. IT WAS REPORTED THAT PATIENT EXPERIENCED AN INFECTION POSSIBLY 6 WEEKS AFTER IMPLANT. REPORTEDLY, THE ENTIRE SYSTEM WAS EXPLANTED ON AN UNKNOWN DATE AND THE PATIENT WAS HOSPITALIZED SEVERAL WEEKS TO TREAT THE INFECTION. THE INFECTION HAS BEEN RESOLVED.

    Patient problems: Post Operative Wound Infection

    MAUDE report 15540280 ↗
  • InjuryAug 1, 2022Fracture; Fracture

    IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. LEAD DIAGNOSTICS INDICATED HIGH IMPEDANCES ON ALL CONTACTS. REPROGRAMMING DID NOT RESOLVE THE ISSUE. X-RAY CONFIRMED LEAD FRACTURE AND THE LEADS ARE TANGLED IN THE IPG POCKET SITE. IN TURN, SURGICAL INTERVENTION MAY OCCUR TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 15137095 ↗
  • InjuryJun 17, 2022Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION. IN TURN, THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 14734493 ↗
  • InjuryJun 3, 2022Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    RELATED MANUFACTURER REPORT NUMBER: 1627487-2022-03095. IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INEFFECTIVE THERAPY. SURGICAL INTERVENTION MAY OCCUR TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 14595243 ↗
  • InjuryJun 3, 2022Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    RELATED MANUFACTURER REPORT NUMBER: 1627487-2022-03096. IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INEFFECTIVE THERAPY. SURGICAL INTERVENTION MAY OCCUR TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 14595236 ↗
  • InjuryMay 25, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PATIENT UNDERWENT SURGICAL PROCEDURE ON (B)(6) 2022 WHEREIN THE CERVICAL WOUND WAS WASHED OUT DUE TO POSTOPERATIVE INFECTION OR HEMATOMA.

    Patient problems: Hematoma; Unspecified Infection

    MAUDE report 14495809 ↗
  • InjuryNov 14, 2021Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. SURGERY OCCURRED TO EXPLANT AND REPLACE THE LEAD TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 12808429 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P010032. One submission often covers many models, so check the product and lot details in each record.

  • Z-0458-2026 · Oct 6, 2025 · Open, Classified
    Eterna SCS IPG (Implantable Pulse Generator), Model Number 32400
  • Z-2506-2023 · Jul 18, 2023 · Open, Classified
    Proclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
  • Z-2507-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2511-2023 · Jul 18, 2023 · Open, Classified
    Infinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2505-2023 · Jul 18, 2023 · Open, Classified
    Proclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
  • Z-2510-2023 · Jul 18, 2023 · Open, Classified
    Infinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2509-2023 · Jul 18, 2023 · Open, Classified
    Proclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
  • Z-2508-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2692-2017 · Jun 2, 2017 · Terminated
    Proclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-2691-2017 · Jun 2, 2017 · Terminated
    Proclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-1785-2013 · Nov 30, 2012 · Terminated
    Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
  • Z-0025-2013 · Sep 4, 2012 · Terminated
    The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
  • Z-1738-2013 · Jul 26, 2012 · Terminated
    Eon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
  • Z-3091-2011 · Jun 2, 2009 · Terminated
    Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im

Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices

All STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.