St. Jude Medical, Inc. · IPG, 16-Channel Rechargeable · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This St. Jude Medical Protégé spinal cord stimulator has 1,229 adverse event reports, including 1 death and 1,223 injuries. Most reports mention inadequate pain relief, and others describe the implanted pulse generator losing its wireless connection to external devices, with some leading to surgery to remove or replace it. No recalls name this device. Reports in the last 12 months (20) are fewer than in the 12 months before (45).
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,226
- Deaths reported
- 1
- Injuries reported
- 1,220
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,226 adverse event reports that list this exact device (DI 05414734405690), received between Jun 11, 2018 and Jul 31, 2026 — classified as 1 death, 1,220 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (377 reports).
14 other recalls cover devices cleared under the same FDA submission (P010032); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method377
- Wireless Communication Problem326
- Adverse Event Without Identified Device or Use Problem157
- Use of Device Problem111
- Therapeutic or Diagnostic Output Failure102
- Insufficient Device Problem Information75
- Premature Discharge of Battery44
- No Device Output38
Patient problems
- Inadequate Pain Relief974
- No Consequences Or Impact To Patient62
- No Clinical Signs, Symptoms or Conditions42
- Failure of Implant40
- Pain35
- Insufficient Information32
- Discomfort31
- Implant Pain27
Recent adverse event reports
The 25 most recent of 1,226 reports, in the reporter's own words as published by the FDA.
InjuryJul 31, 2026Improper or Incorrect Procedure or Method; Wireless Communication Problem
It was reported that the patient's IPG became unable to communicate with external devices. As a result, surgical intervention may be undertaken in the future.
Patient problems: Inadequate Pain Relief
MAUDE report 50105498 ↗InjuryJun 11, 2026Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT THE PATIENT DID NOT CHARGE THEIR IPG AS RECOMMENDED AND THE IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE IPG WAS EXPLANTED TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 25470801 ↗InjuryMay 28, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT HAD THEIR SYSTEM EXPLANTED ON AN UNKNOWN DATE FOR AN UNKNOWN REASON.
Patient problems: Insufficient Information
MAUDE report 25322004 ↗InjuryFeb 27, 2026Wireless Communication Problem
IT WAS REPORTED PATIENT'S THORACIC LEAD HAS HIGH IMPEDANCES. PATIENT WAS UNABLE TO CHARGE THE IPG SINCE 2024. TO ADDRESS THE ISSUES, SURGICAL INTERVENTION MAY BE PENDING.
Patient problems: Inadequate Pain Relief
MAUDE report 24474871 ↗InjuryFeb 16, 2026Wireless Communication Problem
IT WAS REPORTED THE PATIENT'S IPG WAS INOPERABLE. IT IS UNKNOWN IF THE PATIENT STOPPED CHARGING THE DEVICE. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE IPG WAS EXPLANTED AND REPLACED. EFFECTIVE THERAPY WAS ESTABLISHED.
Patient problems: Inadequate Pain Relief
MAUDE report 24347451 ↗InjuryJan 27, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT HAD THEIR IPG EXPLANTED ON AN UNKNOWN DATE TO HAVE PAIN PUMP IMPLANTED.
Patient problems: Insufficient Information
MAUDE report 24193687 ↗InjuryJan 9, 2026Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT THE PATIENT DID NOT CHARGE THEIR IPG AS RECOMMENDED AND THE IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE IPG WAS EXPLANTED AND REPLACED ON (B)(6) 2025 TO ADDRESS THE ISSUE. DURING IPG REPLACEMENT THE PATIENTS LEAD HAD A HIGH IMPEDANCE. AS A RESULT, THE LEAD WAS ALSO REPLACED ON (B)(6) 2025 TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 24027712 ↗InjuryJan 2, 2026Wireless Communication Problem
IT WAS REPORTED THAT THE PATIENT DID NOT CHARGE THEIR IPG AS RECOMMENDED AND THE IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 23950426 ↗InjuryJan 2, 2026Wireless Communication Problem
IT WAS REPORTED THAT THE PATIENT'S IPG IS INOPERABLE. SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 23947863 ↗InjuryDec 29, 2025Improper or Incorrect Procedure or Method; Wireless Communication Problem; Improper or Incorrect Procedure or Method; Wireless Communication Problem
ADDITIONAL INFORMATION INDICATES THAT THE PATIENT STOPPED CHARGING DUE TO AN UNKNOWN REASON.
Patient problems: Inadequate Pain Relief
MAUDE report 23917043 ↗InjuryDec 17, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
ADDITIONAL INFORMATION INDICATES SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 23832232 ↗InjuryDec 9, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT UNDERWENT SURGICAL INTERVENTION ON AN UNKNOWN DATE WHEREIN THE SYSTEM WAS EXPLANTED FOR AN UNKNOWN REASON.
Patient problems: Insufficient Information
MAUDE report 23761599 ↗InjuryDec 4, 2025Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED PATIENT UNDERWENT SURGICAL INTERVENTION ON (B)(6) 2025 WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. THERAPY WAS RESTORED POST-OP.
Patient problems: Inadequate Pain Relief
MAUDE report 23719576 ↗InjuryDec 3, 2025No Apparent Adverse Event; Insufficient Device Problem Information; No Apparent Adverse Event; Insufficient Device Problem Information
ADDITIONAL INFORMATION INDICATES THERE IS NO ALLEGATION OF DEVICE DEFICIENCY. REPLACEMENT OF THE DEVICE IS ELECTIVE TO HAVE AN IMAGING PROCEDURE PERFORMED AS THE PATIENT IS NOT IMPLANTED WITH MRI CONDITIONAL SYSTEM REQUIRED FOR PATIENT¿S IMAGING NEEDS.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23712257 ↗InjuryDec 3, 2025Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THE PATIENT¿S IPG HAD AN INCREASED RECHARGE BURDEN. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE IPG WAS REPLACED ON (B)(6) 2025 TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23707293 ↗InjuryNov 24, 2025Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THE PATIENT'S IPG WAS INOPERABLE. SURGICAL INTERVENTION WAS UNDERTAKEN DURING WHICH THE IPG WAS EXPLANTED AND REPLACED. EFFECTIVE THERAPY WAS CONFIRMED.
Patient problems: Inadequate Pain Relief
MAUDE report 23631695 ↗InjuryNov 18, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT THE PATIENT DID NOT CHARGE THEIR IPG AS RECOMMENDED AND THE IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. AS A RESULT, SURGICAL INTERVENTION MAY BE PENDING TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 23583086 ↗InjuryOct 30, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT PATIENT STOPPED USING THE STIMULATOR FOR AN UNKNOWN REASON. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.
Patient problems: Insufficient Information
MAUDE report 23426033 ↗InjuryOct 24, 2025No Apparent Adverse Event; Insufficient Device Problem Information; No Apparent Adverse Event; Insufficient Device Problem Information
DEVICE IS NO LONGER REPORTABLE AS THERE IS NO ALLEGATION AGAINST THE DEVICE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23377588 ↗InjuryOct 15, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT THE PATIENT DID NOT CHARGE THEIR IPG AS RECOMMENDED AND THE IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 23303657 ↗InjurySep 4, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT THE PATIENT DID NOT CHARGE THEIR IPG AS RECOMMENDED AND THE IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 22974662 ↗InjurySep 3, 2025Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE PATIENT'S IPG WAS INOPERABLE DUE TO NOT HOLDING A CHARGE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE IPG WAS EXPLANTED AND REPLACED.
Patient problems: Failure of Implant; Inadequate Pain Relief
MAUDE report 22959921 ↗InjurySep 3, 2025Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE PATIENT'S IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.
Patient problems: Failure of Implant; Inadequate Pain Relief
MAUDE report 22954771 ↗InjurySep 2, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED PATIENT'S IPGS BECAME INOPERABLE DUE TO PATIENT STOPPED CHARGING. SURGICAL INTERVENTION MAY BE PENDING TO ADDRESS THE ISSUES.
Patient problems: Inadequate Pain Relief
MAUDE report 22935545 ↗InjurySep 2, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THE PATIENT DID NOT RECHARGE THE IPG AS RECOMMENDED. IN TURN, THE IPG BECAME INOPERABLE. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 22932764 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P010032. One submission often covers many models, so check the product and lot details in each record.
- Z-0458-2026 · Oct 6, 2025 · Open, ClassifiedEterna SCS IPG (Implantable Pulse Generator), Model Number 32400
- Z-2509-2023 · Jul 18, 2023 · Open, ClassifiedProclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
- Z-2510-2023 · Jul 18, 2023 · Open, ClassifiedInfinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
- Z-2505-2023 · Jul 18, 2023 · Open, ClassifiedProclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
- Z-2507-2023 · Jul 18, 2023 · Open, ClassifiedProclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
- Z-2511-2023 · Jul 18, 2023 · Open, ClassifiedInfinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
- Z-2506-2023 · Jul 18, 2023 · Open, ClassifiedProclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
- Z-2508-2023 · Jul 18, 2023 · Open, ClassifiedProclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
- Z-2691-2017 · Jun 2, 2017 · TerminatedProclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
- Z-2692-2017 · Jun 2, 2017 · TerminatedProclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
- Z-1785-2013 · Nov 30, 2012 · TerminatedEon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
- Z-0025-2013 · Sep 4, 2012 · TerminatedThe Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
- Z-1738-2013 · Jul 26, 2012 · TerminatedEon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
- Z-3091-2011 · Jun 2, 2009 · TerminatedSoftware version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im
Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.