Octrode™ 3183 (UDI 05414734406123)Recalls & reported problems

St. Jude Medical, Inc. · Lead Kit, 30cm Length · Class 3 · Implant

All models of Octrode (10)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
12
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 05414734406123), received between Sep 11, 2018 and Jan 21, 2025 — classified as 12 injuries.

The most frequently reported specific device problem is “Migration” (5 reports).

14 other recalls cover devices cleared under the same FDA submission (P010032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration5
  • Adverse Event Without Identified Device or Use Problem4
  • High impedance3
  • Therapeutic or Diagnostic Output Failure2

Patient problems

  • Inadequate Pain Relief9
  • No Clinical Signs, Symptoms or Conditions2
  • Failure of Implant1
  • Pain1
  • Skin Erosion1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 21, 2025High impedance; High impedance

    IT WAS REPORTED PATIENT HAD SEVERAL HIGH IMPEDANCES ON THE LEAD. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. THERAPY WAS RESTORED POST-OP.

    Patient problems: Failure of Implant; Inadequate Pain Relief

    MAUDE report 21198874 ↗
  • InjuryJun 3, 2024Migration; Migration

    IT WAS REPORTED THAT PATIENTS LEADS HAD MIGRATED. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2024 WHEREIN LEADS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19455640 ↗
  • InjuryJun 3, 2024Migration; Migration

    IT WAS REPORTED THAT PATIENTS LEADS HAD MIGRATED. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2024 WHEREIN LEADS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19455639 ↗
  • InjuryMar 7, 2024Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    RELATED MANUFACTURER REPORT NUMBER: (B)(4). IT WAS REPORTED THAT THE PATIENT UNDERWENT A DRG TRIAL ON 16FEB2024 DUE TO INEFFECTIVE STIMULATION THEY WERE EXPERIENCING WITH BOTH SCS SYSTEMS. THE INVESTIGATION DID NOT DETERMINE WHICH 3183 LEAD ATTRIBUTED TO THE ISSUE. FURTHER INTERVENTION MAY BE PENDING.

    Patient problems: Inadequate Pain Relief

    MAUDE report 18861175 ↗
  • InjuryJan 23, 2024High impedance; Migration; High impedance; Migration

    RELATED MANUFACTURER REPORT NUMBER: 1627487-2024-00381. IT WAS REPORTED THAT THE PATIENT HAD HIGH IMPEDANCES ON BOTH IMPLANTED LEADS AS WELL AS BOTH IMPLANTED LEADS HAD MIGRATED. X-RAYS TAKEN INTRAOPERATIVELY CONFIRMED BOTH LEADS HAD MIGRATED. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2024 WHEREIN BOTH PATIENT'S LEADS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 18561352 ↗
  • InjuryJan 23, 2024High impedance; Migration; High impedance; Migration

    RELATED MANUFACTURER REPORT NUMBER: (B)(4) IT WAS REPORTED THAT THE PATIENT HAD HIGH IMPEDANCES ON BOTH IMPLANTED LEADS AS WELL AS BOTH IMPLANTED LEADS HAD MIGRATED. X-RAYS TAKEN INTRAOPERATIVELY CONFIRMED BOTH LEADS HAD MIGRATED. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2024 WHEREIN BOTH PATIENT'S LEADS WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 18561348 ↗
  • InjuryJan 26, 2023Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT SURGICAL INTERVENTION WAS UNDERTAKEN IN WHICH THE PATIENT'S LEAD WAS EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 16246006 ↗
  • InjuryDec 16, 2020Migration; Migration

    REFERENCE MANUFACTURER NUMBER: 1627487-2020-48884. IT WAS REPORTED THAT THE PATIENT SUFFERED FROM SEVERAL FALLS THAT COINCIDED WITH A LOSS OF THERAPY. X-RAYS WERE TAKEN WHICH CONFIRMED SLIGHT LEAD MIGRATION OF THE LEFT SCS PERCUTANEOUS LEAD. IN TURN, THE PATIENT UNDERWENT SURGICAL INTERVENTION WHEREIN BOTH SCS LEADS WERE EXPLANTED AND REPLACED WITH A PADDLE LEAD TO ADDRESS THE ISSUE. SINCE IT IS NOT KNOWN WHICH DEVICE CONTRIBUTED TO THE ISSUE, ALL POSSIBLE DEVICES ARE BEING REPORTED ON.

    Patient problems: Inadequate Pain Relief

    MAUDE report 11019998 ↗
  • InjuryMar 12, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE 2 OF 3: REFERENCE MFR REPORT # 1627487-2019-03297, REFERENCE MFR REPORT # 3006705815-2019-00929. IT IS REPORTED THAT PATIENT WAS NOT RECEIVING ADEQUATE THERAPY FROM THE SCS SYSTEM. IN TURN, THE SYSTEM WAS EXPLANTED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 8412820 ↗
  • InjuryFeb 21, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE 5 OF 6. REFERENCE MFR REPORT # 1627487-2019-02487, REFERENCE MFR REPORT # 1627487-2019-02488, REFERENCE MFR REPORT # 1627487-2019-02489, REFERENCE MFR REPORT # 1627487-2019-02490, REFERENCE MFR REPORT # 1627487-2019-02492. IT WAS REPORTED THAT PATIENT HAD BOTH OF THEIR PNS SYSTEMS EXPLANTED. PER PATIENT, THE SYSTEMS WERE NO LONGER WORKING AND WANTED TO HAVE THEM EXPLANTED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 8358359 ↗
  • InjuryNov 5, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    FURTHER FOLLOW-UP INDICATES THE PATIENT HAD SURGICAL INTERVENTION ON (B)(6) 2018, DURING WHICH THE PATIENT'S LEAD WAS REPOSITIONED TO RESOLVE THIS ISSUE.

    Patient problems: Pain; Skin Erosion

    MAUDE report 8039929 ↗
  • InjurySep 11, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE 1 OF 2: REFERENCE MFR. REPORT#1627487-2018-08220. IT WAS REPORTED THAT THE PATIENT NEVER EXPERIENCED EFFECTIVE STIMULATION FROM THE SCS SYSTEM. MULTIPLE REPROGRAMMING'S WERE ATTEMPTED TO NO AVAIL. AS SUCH, THE SCS SYSTEM WAS EXPLANTED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 7867267 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P010032. One submission often covers many models, so check the product and lot details in each record.

  • Z-0458-2026 · Oct 6, 2025 · Open, Classified
    Eterna SCS IPG (Implantable Pulse Generator), Model Number 32400
  • Z-2507-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2506-2023 · Jul 18, 2023 · Open, Classified
    Proclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
  • Z-2505-2023 · Jul 18, 2023 · Open, Classified
    Proclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
  • Z-2508-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2510-2023 · Jul 18, 2023 · Open, Classified
    Infinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2511-2023 · Jul 18, 2023 · Open, Classified
    Infinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2509-2023 · Jul 18, 2023 · Open, Classified
    Proclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
  • Z-2692-2017 · Jun 2, 2017 · Terminated
    Proclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-2691-2017 · Jun 2, 2017 · Terminated
    Proclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-1785-2013 · Nov 30, 2012 · Terminated
    Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
  • Z-0025-2013 · Sep 4, 2012 · Terminated
    The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
  • Z-1738-2013 · Jul 26, 2012 · Terminated
    Eon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
  • Z-3091-2011 · Jun 2, 2009 · Terminated
    Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im

Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices

All STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices →