Skip to content

Lamitrode™ 3244Recalls & reported problems

St. Jude Medical, Inc. · 44 Lead Kit, 60cm Length · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This St. Jude Medical Lamitrode 3244 is a lead kit used with an implanted spinal cord stimulator for pain relief. The 56 reports mostly describe leads that stopped delivering effective therapy, with high impedance readings and sometimes broken or disconnected wiring, leading patients to lose pain relief;55 reports involved injuries and no deaths were reported, and some leads had to be removed or replaced in surgery. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
56
Deaths reported
0
Injuries reported
55
Recalls naming it
0

Summary

The FDA's MAUDE database contains 56 adverse event reports that list this exact device (DI 05414734406215), received between Jun 29, 2018 and Nov 7, 2024 — classified as 55 injuries, 1 malfunction.

The most frequently reported specific device problem is “High impedance” (26 reports).

14 other recalls cover devices cleared under the same FDA submission (P010032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High impedance26
  • Adverse Event Without Identified Device or Use Problem13
  • Therapeutic or Diagnostic Output Failure10
  • Disconnection4
  • Insufficient Device Problem Information3
  • Fracture2
  • Failure to Advance1
  • Failure to Deliver Energy1

Patient problems

  • Inadequate Pain Relief47
  • Fall3
  • No Clinical Signs, Symptoms or Conditions3
  • Failure of Implant2
  • Insufficient Information2
  • No Consequences Or Impact To Patient2
  • No Information2
  • Erosion1

Recent adverse event reports

The 25 most recent of 56 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 7, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT PATIENTS LEAD WAS EXPLANTED ON AN UNKNOWN DATE FOR AN UNKNOWN REASON.

    Patient problems: Insufficient Information

    MAUDE report 20636009 ↗
  • InjuryAug 14, 2024High impedance; High impedance

    ADDITIONAL INFORMATION WAS OBTAINED, REVEALING THAT THE LEADS WERE REPLACED VIA SURGICAL INTERVENTION ON (B)(6) 2024. IT WAS ALSO CONFIRMED THE PATIENT DID NOT HAVE EFFECTIVE THERAPY PRIOR TO SURGERY. THERAPY WAS RESTORED POST OP.

    Patient problems: Failure of Implant; Inadequate Pain Relief; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19980639 ↗
  • InjuryJul 17, 2024Use of Device Problem; Failure to Advance; Use of Device Problem; Failure to Advance

    IT WAS REPORTED THAT THE DURING AN IPG REPLACEMENT SURGERY, THE PHYSICIAN NICKED THE INSULATION OF THE LEAD WHEN OPENING THE BATTERY POCKET. THE WIRE WAS SUPERFICIAL ON TOP OF THE BATTERY. THE SURGICAL PROCEDURE WAS ABORTED, AND THE PHYSICIAN CLOSED THE INCISION. THE PATIENT HAD NO THERAPY AT THIS TIME. SURGICAL INTERVENTION MAY TAKE PLACE AT A FUTURE DATE TO ADDRESS THE ISSUE.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19765078 ↗
  • InjuryFeb 29, 2024High impedance; High impedance

    IT WAS REPORTED THAT PATIENT'S LEAD HAD HIGH IMPEDANCES ON ALL CONTACTS. AS A RESULT, THE PATIENT UNDERWENT SURGICAL INTERVENTION DURING WHICH THE LEAD WAS EXPLANTED AND REPLACED. EFFECTIVE THERAPY WAS ESTABLISHED POST-OPERATIVELY.

    Patient problems: Failure of Implant; Inadequate Pain Relief

    MAUDE report 18809590 ↗
  • InjuryAug 14, 2023High impedance; High impedance

    RELATED MANUFACTURER REFERENCE NUMBER 1627487-2023-03818. IT WAS REPORTED THE PATIENT IS NOT RECEIVING EFFECTIVE THERAPY DUE TO HIGH IMPEDANCES. AS SUCH, SURGICAL INTERVENTION MAY BE PENDING TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 17540173 ↗
  • InjuryAug 14, 2023High impedance; High impedance

    RELATED MANUFACTURER REFERENCE NUMBER 1627487-2023-03817. IT WAS REPORTED THE PATIENT IS NOT RECEIVING EFFECTIVE THERAPY DUE TO HIGH IMPEDANCES. AS SUCH, SURGICAL INTERVENTION MAY BE PENDING TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 17540170 ↗
  • InjuryJun 21, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    RELATED MANUFACTURER REPORT NUMBER: 1627487-2023-02839, 1627487-2023-02954, 1627487-2023-02956. IT WAS REPORTED THAT THE LEAD WAS EXPLANTED ON AN UNKNOWN DATE FOR AN UNKNOWN REASON. NO OTHER INFORMATION IS KNOWN AT THIS TIME.

    Patient problems: Insufficient Information

    MAUDE report 17174972 ↗
  • InjuryMay 31, 2023Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT WAS NOT GETTING RELIEF FROM THEIR SYSTEM. AS A RESULT, PATIENT HAD SYSTEM EXPLANTED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 17035396 ↗
  • InjuryMay 26, 2023High impedance; High impedance

    ADDITIONAL INFORMATION WAS RECEIVED INDICATING THAT THE PATIENT UNDERWENT SURGICAL INTERVENTION ON (B)(6) 2023 WHERE IN THE PATIENT'S SCS SYSTEM WAS FULLY EXPLANTED AND REPLACED. THERAPY WAS RESTORED POST OPERATIVELY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 17015154 ↗
  • InjuryDec 9, 2022Impedance Problem; Impedance Problem

    RELATED MANUFACTURING REPORT NUMBER: 1627487-2022-06750. IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY DUE TO INVALID IMPEDANCES ON THE SCS LEAD AND THE PATIENT LOST EFFECTIVE COVERAGE DUE TO THE L5 DRG LEAD BEING MIGRATED. AS A RESULT, THE PATIENT UNDERWENT SURGICAL INTERVENTION ON (B)(6) 2022 TO HAVE THEIR BOTH LEADS REPLACED. PATIENT NOW HAS EFFECTIVE THERAPY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 15954688 ↗
  • InjuryNov 17, 2022Failure to Deliver Energy; Failure to Deliver Energy

    IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION. SURGICAL INTERVENTION MAY TAKE PLACE TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 15811819 ↗
  • InjuryMay 5, 2022High impedance; High impedance

    IT WAS REPORTED THE PATIENTS LEAD DISPLAYED HIGH IMPEDANCE. SURGICAL INTERVENTION MAY BE PENDING TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 14290346 ↗
  • InjuryMar 2, 2022High impedance; High impedance

    RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2022-01201. IT WAS REPORTED THE PATIENTS SYSTEM DISPLAYED HIGH IMPEDANCES ON ALL CONTACTS. SURGICAL INTERVENTION MAY BE PENDING TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 13655961 ↗
  • InjuryMar 2, 2022High impedance; High impedance

    RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2022-01202. IT WAS REPORTED THE PATIENTS SYSTEM DISPLAYED HIGH IMPEDANCES ON ALL CONTACTS. SURGICAL INTERVENTION MAY BE PENDING TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 13655958 ↗
  • InjuryNov 18, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED INDICATES SURGERY TOOK PLACE ON (B)(6) 2022 WHEREIN THE SECOND LEAD WAS EXPLANTED DUE TO THE LEAD ERODING THROUGH THE SKIN. THE SYSTEM WAS REPLACED. THE IPG WAS ELECTIVELY REPLACED.

    Patient problems: Erosion

    MAUDE report 12840390 ↗
  • InjuryNov 16, 2021Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2021-18398. IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE PADDLE LEAD AND PERCUTANEOUS LEAD WERE EXPLANTED AND REPLACED WITH A PENTA PADDLE LEAD. DURING THE SURGERY, THE PHYSICIAN ELECTIVELY REPLACED THE IPG AS WELL. POST-OPERATIVELY, STIMULATION THERAPY WAS RESTORED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 12825130 ↗
  • InjurySep 12, 2021Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE STIM. AS A RESULT, THE LEAD WAS EXPLANTED AND REPLACED RESOLVING THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 12457993 ↗
  • InjuryJun 1, 2021Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THE PATIENT WAS EXPERIENCING INEFFECTIVE THERAPY. AS A RESULT, THE PATIENT UNDERWENT SURGICAL INTERVENTION DURING WHICH THE LEAD WAS EXPLANTED AND REPLACED. EFFECTIVE THERAPY WAS ESTABLISHED POST-OPERATIVELY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 11917365 ↗
  • InjuryFeb 9, 2021High impedance; Therapeutic or Diagnostic Output Failure; High impedance; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION DUE TO HIGH IMPEDANCES. SURGICAL INTERVENTION TOOK PLACE WHEREIN THE LEAD WAS EXPLANTED AND REPLACED. EFFECTIVE THERAPY WAS ESTABLISHED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 11300618 ↗
  • InjuryFeb 3, 2021Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT HAD INEFFECTIVE THERAPY THAT COULD NOT BE RESOLVED BY SYSTEM REPROGRAMMING. IN TURN, THE PATIENT MAY UNDERGO SURGICAL INTERVENTION TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 11276117 ↗
  • InjuryJan 6, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RELATED MFR REPORTS: 1627487-2021-00098 AND 1627487-2021-00100. IT WAS REPORTED THE PATIENT'S SCS SYSTEM WAS REMOVED AND REPLACED WITH A COMPETITOR'S SYSTEM AT AN UNKNOWN DATE AND FOR UNSPECIFIED REASONS. NO FURTHER INFORMATION WAS AVAILABLE AT THIS TIME.

    Patient problems: No Information

    MAUDE report 11128439 ↗
  • InjuryJan 6, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RELATED MFR REPORTS: 1627487-2021-00099 AND 1627487-2021-00100. IT WAS REPORTED THE PATIENT'S SCS SYSTEM WAS REMOVED AND REPLACED WITH A COMPETITOR'S SYSTEM AT AN UNKNOWN DATE AND FOR UNSPECIFIED REASONS. NO FURTHER INFORMATION WAS AVAILABLE AT THIS TIME.

    Patient problems: No Information

    MAUDE report 11128433 ↗
  • InjuryDec 6, 2020Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    ADDITIONAL INFORMATION WAS RECEIVED THAT SURGERY OCCURRED ON (B)(6) 2020 TO IMPLANT AN ADDITIONAL LEAD TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 10956058 ↗
  • InjuryNov 18, 2020High impedance; High impedance

    IT WAS REPORTED THAT THE PATIENT UNDERWENT SURGICAL INTERVENTION WHEREIN THE SCS SYSTEM WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 10864322 ↗
  • InjuryNov 16, 2020High impedance; High impedance

    IT WAS REPORTED THE PATIENT'S SCS SYSTEM WAS EXHIBITING HIGH IMPEDANCE ON MULTIPLE LEAD CONTACTS. AS A RESULT, THE PATIENT IS NOT RECEIVING STIMULATION THERAPY COVERAGE OF THE PAIN AREA. SURGICAL INTERVENTION MAY BE TAKEN TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 10847417 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P010032. One submission often covers many models, so check the product and lot details in each record.

  • Z-0458-2026 · Oct 6, 2025 · Open, Classified
    Eterna SCS IPG (Implantable Pulse Generator), Model Number 32400
  • Z-2510-2023 · Jul 18, 2023 · Open, Classified
    Infinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2509-2023 · Jul 18, 2023 · Open, Classified
    Proclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
  • Z-2505-2023 · Jul 18, 2023 · Open, Classified
    Proclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
  • Z-2508-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2506-2023 · Jul 18, 2023 · Open, Classified
    Proclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
  • Z-2511-2023 · Jul 18, 2023 · Open, Classified
    Infinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2507-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2691-2017 · Jun 2, 2017 · Terminated
    Proclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-2692-2017 · Jun 2, 2017 · Terminated
    Proclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-1785-2013 · Nov 30, 2012 · Terminated
    Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
  • Z-0025-2013 · Sep 4, 2012 · Terminated
    The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
  • Z-1738-2013 · Jul 26, 2012 · Terminated
    Eon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
  • Z-3091-2011 · Jun 2, 2009 · Terminated
    Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im

Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices

All STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page