Ellipse™ CD2275-36QRecalls & reported problems
St. Jude Medical, Inc. · Tiered-therapy cardioverter/defibrillator VVEV VVIR · Class 3 · Implant
All models of Ellipse (12)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05414734505918), received between Sep 6, 2018 and Jun 19, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Connection Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Connection Problem1
- Failure to Convert Rhythm1
- Inappropriate/Inadequate Shock/Stimulation1
- Incorrect Interpretation of Signal1
- Migration1
Patient problems
- Discomfort1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJun 19, 2023Failure to Convert Rhythm; Incorrect Interpretation of Signal; Inappropriate/Inadequate Shock/Stimulation; Failure to Convert Rhythm; Incorrect Interpretation of Signal; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE ANTI-TACHYCARDIA PACING AND HIGH VOLTAGE (HV) SHOCK DUE TO A SUPRAVENTRICULAR TACHYCARDIA (SVT) RHYTHM. THE HV SHOCK INDUCED VENTRICULAR TACHYCARDIA (VT) WHICH WAS UNABLE TO BE TERMINATED BY THE DEVICE WITH AN ADDITIONAL HV SHOCK. THE DEVICE WAS REPROGRAMMED TO RESOLVE THE EVENT. THE PATIENT WAS STABLE.
Patient problems: Discomfort
MAUDE report 17152826 ↗InjurySep 6, 2018Connection Problem; Migration; Connection Problem; Migration
DURING A FOLLOW-UP IN CLINIC, THERE WAS DECREASED SENSING AND AN INCREASED CAPTURE THRESHOLD NOTED ON THE RIGHT VENTRICULAR (RV) LEAD. DIAGNOSTIC IMAGING WAS PERFORMED AND REVEALED THAT THE DEVICE HAD MIGRATED IN THE POCKET AND CAUSED BOTH THE RV AND RIGHT ATRIAL (RA) LEADS TO DISLODGE. A SURGERY WAS PERFORMED TO ADDRESS THE DISLODGEMENT AND THE SENSING VALUE OF THE RA LEAD DECREASED. BOTH LEADS WERE THEN EXPLANTED AND REPLACED. WHEN THE NEW LEADS WERE BEING RECONNECTED TO THE OLD DEVICE, THE RV LEAD COULD NOT BE PLACED IN THE DEVICE HEADER, SO THE DEVICE WAS ALSO REPLACED. THE PATIENT WAS STABLE FOLLOWING THE PROCEDURE AND SUFFERED NO ADVERSE CONSEQUENCES DUE TO THE EVENT. RELATED MANUFACTURER REPORT NUMBERS: 2938836-2018-10423, 2938836-2018-10424.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7850709 ↗
Other Pulse generator, permanent, implantable devices
- Assurity™ PM2240St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall84 deaths4,076 reports
- Fortify Assura™ CD2357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable63 deaths3,571 reports
- Advisa DR MRI™ SureScan® A2DR01Medtronic, Inc. · Pulse generator, permanent, implantable51 deaths2,298 reports
- Fortify Assura™ CD1357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable35 deaths1,933 reports
- Ellipse™ CD2411-36QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable42 deaths1,498 reports
- Fortify Assura™ CD2357-40CSt. Jude Medical, Inc. · Pulse generator, permanent, implantable25 deaths1,378 reports
- Accent® DR RFSt. Jude Medical, Inc. · Pulse generator, permanent, implantable73 deaths1,370 reports
- Edora 8 DR-T 407145BIOTRONIK SE & Co. KG · Pulse generator, permanent, implantable10 deaths982 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page