Ellipse™ CD2275-36QRecalls & reported problems

St. Jude Medical, Inc. · Tiered-therapy cardioverter/defibrillator VVEV VVIR · Class 3 · Implant

All models of Ellipse (12)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05414734505918), received between Sep 6, 2018 and Jun 19, 2023 — classified as 2 injuries.

The most frequently reported specific device problem is “Connection Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Connection Problem1
  • Failure to Convert Rhythm1
  • Inappropriate/Inadequate Shock/Stimulation1
  • Incorrect Interpretation of Signal1
  • Migration1

Patient problems

  • Discomfort1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 19, 2023Failure to Convert Rhythm; Incorrect Interpretation of Signal; Inappropriate/Inadequate Shock/Stimulation; Failure to Convert Rhythm; Incorrect Interpretation of Signal; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE ANTI-TACHYCARDIA PACING AND HIGH VOLTAGE (HV) SHOCK DUE TO A SUPRAVENTRICULAR TACHYCARDIA (SVT) RHYTHM. THE HV SHOCK INDUCED VENTRICULAR TACHYCARDIA (VT) WHICH WAS UNABLE TO BE TERMINATED BY THE DEVICE WITH AN ADDITIONAL HV SHOCK. THE DEVICE WAS REPROGRAMMED TO RESOLVE THE EVENT. THE PATIENT WAS STABLE.

    Patient problems: Discomfort

    MAUDE report 17152826 ↗
  • InjurySep 6, 2018Connection Problem; Migration; Connection Problem; Migration

    DURING A FOLLOW-UP IN CLINIC, THERE WAS DECREASED SENSING AND AN INCREASED CAPTURE THRESHOLD NOTED ON THE RIGHT VENTRICULAR (RV) LEAD. DIAGNOSTIC IMAGING WAS PERFORMED AND REVEALED THAT THE DEVICE HAD MIGRATED IN THE POCKET AND CAUSED BOTH THE RV AND RIGHT ATRIAL (RA) LEADS TO DISLODGE. A SURGERY WAS PERFORMED TO ADDRESS THE DISLODGEMENT AND THE SENSING VALUE OF THE RA LEAD DECREASED. BOTH LEADS WERE THEN EXPLANTED AND REPLACED. WHEN THE NEW LEADS WERE BEING RECONNECTED TO THE OLD DEVICE, THE RV LEAD COULD NOT BE PLACED IN THE DEVICE HEADER, SO THE DEVICE WAS ALSO REPLACED. THE PATIENT WAS STABLE FOLLOWING THE PROCEDURE AND SUFFERED NO ADVERSE CONSEQUENCES DUE TO THE EVENT. RELATED MANUFACTURER REPORT NUMBERS: 2938836-2018-10423, 2938836-2018-10424.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7850709 ↗

Other Pulse generator, permanent, implantable devices

All Pulse generator, permanent, implantable devices →