Ellipse™ DRRecalls & reported problems
St. Jude Medical, Inc. · Tiered-therapy cardioverter/defibrillator VVEV VVIR · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Ellipse DR is a St. Jude Medical implantable cardioverter-defibrillator. The 79 adverse event reports mention oversensing, inappropriate shocks or reset, lead capture problems, charging failures, and battery longevity issues, including 1 death and 29 injuries. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 79
- Deaths reported
- 1
- Injuries reported
- 29
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 79 adverse event reports that list this exact device (DI 05414734505987), received between Jun 13, 2018 and Mar 1, 2024 — classified as 1 death, 29 injuries, 49 malfunctions.
The most frequently reported specific device problem is “Over-Sensing” (29 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Over-Sensing29
- Inappropriate/Inadequate Shock/Stimulation10
- Inappropriate or Unexpected Reset9
- Failure to Charge6
- Adverse Event Without Identified Device or Use Problem5
- Defective Device5
- Incorrect Interpretation of Signal5
- Delayed Charge Time3
Patient problems
- No Known Impact Or Consequence To Patient41
- No Clinical Signs, Symptoms or Conditions15
- Therapeutic Effects, Unexpected10
- Insufficient Information5
- Arrhythmia2
- No Consequences Or Impact To Patient2
- Syncope2
- Unspecified Infection2
Recent adverse event reports
The 25 most recent of 79 reports, in the reporter's own words as published by the FDA.
InjuryMar 1, 2024Signal Artifact/Noise; Failure to Capture; Signal Artifact/Noise; Failure to Capture
IT WAS REPORTED, THAT THE PATIENT PRESENTED IN CLINIC FOR FOLLOW UP. UPON INTERROGATION, IT WAS NOTED, THAT THE RIGHT VENTRICULAR (RV) LEAD WAS EXHIBITING FAILURE TO CAPTURE. THE RV LEAD WAS CAPPED AND REPLACED ON (B)(6) 2024. THE PATIENT WAS IN STABLE CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18822821 ↗MalfunctionSep 28, 2023Defective Device; Defective Device
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 17831143 ↗MalfunctionNov 4, 2022Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT A PATIENT PRESENTED TO CLINIC FOR FOLLOW UP. DEVICE INTERROGATION REVEALED THE THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD REACHED ERI. IT WAS NOTED, THAT A PREVIOUS BATTERY LONGEVITY HAD BEEN OVERESTIMATED. THE ICD WAS EXPLANTED AND REPLACED DUE TO ERI. THE PATIENT CONDITION WAS STABLE BEFORE, DURING, AND AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15739793 ↗MalfunctionAug 25, 2022Defective Device; Defective Device
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 15296039 ↗MalfunctionJul 28, 2022Over-Sensing; Over-Sensing
IT WAS REPORTED THAT A PATIENT PRESENTED IN-CLINIC. UPON INTERROGATION, THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR WAS FOUND TO HAVE MYOPOTENTIALS OVER-SENSING, WHICH RESULTED IN NON-SUSTAINED V OVERSENSING EPISODES. CORRECTIVE PROGRAMMING CHANGES WERE MADE. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15122932 ↗MalfunctionApr 22, 2022Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR EXHIBITED MYOPOTENTIAL OVERSENSING. AFTER CONSIDERATION, IT WAS NOTED THAT IT WAS DUE TO A COUGHING SPELL AND IT ONLY HAPPENED WHEN THE PATIENT HAD COVID. NO INTERVENTION WAS PERFORMED. THE PATIENT WAS IN STABLE CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14190087 ↗MalfunctionJan 25, 2022Defective Device; Defective Device
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 13349134 ↗MalfunctionNov 17, 2021Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE PATIENT PRESENTED REMOTELY VIA MERLIN.NET. IT WAS NOTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS POST-PACED T-WAVE OVER-SENSING (PPTWOS). THE PATIENT WAS ASYMPTOMATIC. THE ICD WAS REPROGRAMMED AND THE PATIENT WAS IN STABLE CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12829823 ↗MalfunctionNov 12, 2021Defective Device; Defective Device
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 12805661 ↗InjuryOct 28, 2021Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset
NEW INFORMATION RECEIVED NOTED THAT THE PATIENT'S IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR (ICD) WAS EXPLANTED AND REPLACED. THE PATIENT WAS ASYMPTOMATIC BEFORE, DURING AND AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12716185 ↗InjuryOct 5, 2021Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset
IT WAS REPORTED THAT THE PATIENT PRESENTED FOR A GENERATOR CHANGE OUT PROCEDURE. UPON REVIEW, THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR WAS FOUND TO BE IN THE BACK-UP MODE. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR WAS EXPLANTED AND REPLACED. PATIENT WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12577930 ↗InjuryJun 21, 2021Over-Sensing; Defective Device; Over-Sensing; Defective Device
IT WAS REPORTED THAT IMPLANTABLE CARDIOVERTER DEFIBRILLATOR EXHIBITED FAR FIELD OVERSENSING AND NOISE THAT WAS OVERSENSED BY THE DEVICE. BOTH SENSING ANOMALIES WERE UNABLE TO BE REPRODUCED IN CLINIC. THE PHYSICIAN ALLEGED THE SOURCE OF THE NOISE WAS AN ISSUE INSIDE THE HEADER. THE DEVICE WAS EXPLANTED AND REPLACED. THE PATIENT WAS IN STABLE CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12037465 ↗MalfunctionJun 1, 2021Over-Sensing; Over-Sensing
NEW INFORMATION NOTES THAT THE PROGRAMMING CHANGES WERE MADE ON THE DEVICE. NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11917925 ↗MalfunctionMay 21, 2021Failure to Interrogate; Failure to Interrogate
IT WAS REPORTED THAT THE ASYMPTOMATIC PATIENT PRESENTED FOR A FOLLOW UP IN CLINIC. IT WAS NOTED THAT THERE WAS NO RADIOFREQUENCY (RF) TELEMETRY ON THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR. A FIRMWARE UPGRADE WAS COMPLETED SUCCESSFULLY AND TELEMETRY WAS RESTORED. THE PATIENT WAS STABLE BEFORE, DURING AND AFTER THE UPGRADE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11863923 ↗MalfunctionMay 12, 2021Over-Sensing; Pacing Problem; Over-Sensing; Pacing Problem
NEW INFORMATION RECEIVED NOTES THAT CORRECTIVE PROGRAMMING CHANGES WERE PERFORMED ON (B)(6) 2021.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11813222 ↗MalfunctionApr 27, 2021Over-Sensing; Over-Sensing
IT WAS REPORTED THAT A PATIENT PRESENTED REMOTELY ON (B)(6) 2021 WITH EPISODES OF POST-PACED T-WAVE OVERSENSING ON THEIR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR. PROGRAMMING CHANGES WERE MADE (B)(6) 2021. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11730235 ↗MalfunctionApr 21, 2021Failure to Disconnect; Failure to Disconnect
IT WAS REPORTED THAT THE PATIENT PRESENTED IN HOSPITAL FOR A SCHEDULED GENERATOR CHANGE. DURING THE PROCEDURE, THE PHYSICIAN WAS UNABLE TO REMOVE THE RIGHT VENTRICULAR (RV) LEAD PIN FOR THE RV PORT AFTER SUCCESSFULLY ENGAGING AND LOOSENING THE SET SCREW. THE PHYSICIAN PROCEEDED TO REMOVE THE SILICONE SEPTUM AND SET SCREW FROM THE DEVICE HEADER USING MOSQUITO FORCEPS BUT REMAINED UNABLE TO REMOVE THE RV LEAD PIN FROM THE RV PORT. THE PHYSICIAN ATTEMPTED TO APPLY LEVERAGE TO THE LEAD PIN THROUGH THE SET SCREW CHANNEL USING VARIOUS INSTRUMENTS (TIP SCALPEL, TIP MOSQUITO FORCEPS, LONG END OF AN "L" SHAPED WRENCH), WITHOUT SUCCESS. ORTHPEDIC BONE NIPPERS WERE SUBSEQUENTLY USED TO CUT THROUGH THE HEADER TO FREE THE LEAD, WITH SUCCESS. THE RV LEAD WAS TESTED FOLLOWING IT'S REMOVAL, WITH PACING SYSTEM ANALYZER MEASUREMENTS STABLE WHEN COMPARED TO TESTING PERFORMED IN HOLDING PRIOR TO CASE START. THE CASE PROCEEDED AND CONCLUDED WITHOUT FURTHER COMPLICATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11700955 ↗InjuryFeb 3, 2021Failure to Charge; Failure to Charge
IT WAS REPORTED THE PATIENT PRESENTED IN CLINIC. THEIR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR EXHIBITED CHARGE TIME LIMIT. THE PHYSICIAN EXPLANTED AND REPLACED THE DEVICE. THE PATIENT WAS IN STABLE CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11276953 ↗MalfunctionDec 1, 2020Failure to Disconnect; Failure to Disconnect
RELATED MANUFACTURER REFERENCE NUMBER: 2017865-2020-19435. IT WAS REPORTED THAT THE PATIENT PRESENTED IN CLINIC FOR A GENERATOR CHANGE PROCEDURE. DURING THE PROCEDURE, THE RIGHT VENTRICULAR LEAD WAS STUCK IN THE HEADER OF THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR DESPITE THE SET SCREW BEING LOOSENED. THE LEAD WAS REMOVED USING A BONE CUTTER. IT WAS ALSO NOTED THAT THE RIGHT ATRIAL LEAD HAD EXTERNALIZED CONDUCTORS DUE TO PROCEDURAL DAMAGE. THE RIGHT ATRIAL LEAD WAS CAPPED AND NOT REPLACED DURING PROCEDURE ON (B)(6) 2020 DUE TO THE PATIENT EXPERIENCING CHRONIC ATRIAL FIBRILLATION. THE PATIENT WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10928783 ↗InjuryNov 29, 2020Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR EXHIBITED FAR-FIELD P-WAVE OVERSENSING RESULTING IN INAPPROPRIATE SHOCK. PROGRAMMING CHANGES WERE MADE. THE PATIENT WAS STABLE.
Patient problems: Therapeutic Effects, Unexpected
MAUDE report 10916652 ↗InjuryNov 10, 2020Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR EXHIBITED INAPPROPRIATE ANTI-TACHYCARDIA PACING DUE TO P WAVES FALLING INTO BLANKING AND CAUSING THE DEVICE TO IDENTIFY VENTRICLE RATE BRANCH MORE THAN THE ATRIAL RATE BRANCH. NO INTERVENTION WAS PERFORMED. NO PATIENT CONSEQUENCES.
Patient problems: Therapeutic Effects, Unexpected
MAUDE report 10820498 ↗MalfunctionOct 8, 2020Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE ICD EXHIBITED POST SENSED T WAVE OVERSENSING. PROGRAMMING CHANGES WERE SUGGESTED. NO PATIENT SYMPTOMS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10653556 ↗InjurySep 9, 2020Incorrect Interpretation of Signal; Inappropriate/Inadequate Shock/Stimulation; Incorrect Interpretation of Signal; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THE PATIENT PRESENTED TO THE EMERGENCY ROOM HAVING RECEIVED SHOCK THERAPY. UPON INTERROGATION, IT WAS OBSERVED THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR WAS INCORRECTLY INTERPRETING TACHYCARDIA EPISODES AND INAPPROPRIATELY DELIVERING SHOCK. PROGRAMMING CHANGES WERE COMPLETED TO ADDRESS THIS ISSUE. THE PATIENT WAS STABLE.
Patient problems: Therapeutic Effects, Unexpected
MAUDE report 10508417 ↗InjuryAug 25, 2020Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY ROOM. UPON REVIEW, IT WAS REVEALED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR PROVIDED AN INAPPROPRIATE SHOCK. PROGRAMMING CHANGES WERE MADE. THE PATIENT WAS STABLE.
Patient problems: Therapeutic Effects, Unexpected
MAUDE report 10446863 ↗MalfunctionAug 22, 2020Failure to Charge; Failure to Charge
IT WAS REPORTED THAT A PATIENT PRESENTED WITH AN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR THAT REACHED CAPACITOR CHARGE TIME. NO FURTHER INTERVENTION WAS MADE AT THE TIME. PATIENT WOULD BE FURTHER MONITORED. PATIENT WAS STABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10440534 ↗
Other Pulse generator, permanent, implantable devices
- Assurity™ PM2240St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall84 deaths4,076 reports
- Fortify Assura™ CD2357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable63 deaths3,571 reports
- Advisa DR MRI™ SureScan® A2DR01Medtronic, Inc. · Pulse generator, permanent, implantable51 deaths2,298 reports
- Fortify Assura™ CD1357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable35 deaths1,933 reports
- Ellipse™ CD2411-36QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable42 deaths1,498 reports
- Fortify Assura™ CD2357-40CSt. Jude Medical, Inc. · Pulse generator, permanent, implantable25 deaths1,378 reports
- Accent® DR RFSt. Jude Medical, Inc. · Pulse generator, permanent, implantable73 deaths1,370 reports
- Edora 8 DR-T 407145BIOTRONIK SE & Co. KG · Pulse generator, permanent, implantable10 deaths982 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page