Merlin@home™ EX1151Recalls & reported problems
St. Jude Medical, Inc. · Wireless USB Adapter for use with Merlin@home Transmitter · Class 3
All models of Merlin@home (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 05414734507011), received between Dec 27, 2019 and Nov 27, 2024 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Thermal Decomposition of Device” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Thermal Decomposition of Device3
- Device Sensing Problem1
- Inappropriate or Unexpected Reset1
- Inappropriate/Inadequate Shock/Stimulation1
- Incorrect Interpretation of Signal1
- Melted1
- Over-Sensing1
- Pacing Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 27, 2024Thermal Decomposition of Device; Thermal Decomposition of Device
IT WAS REPORTED THAT BURN MARKS WERE OBSERVED ON THE TRANSMITTER'S USB CELLULAR ADAPTER. THE TRANSMITTER WITH ADAPTER WAS EXCHANGED. THERE WERE NO REPORTED PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20802693 ↗InjuryOct 31, 2023Incorrect Interpretation of Signal; Inappropriate/Inadequate Shock/Stimulation; Incorrect Interpretation of Signal; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE DEVICE INAPPROPRIATELY DELIVERED ANTITACHYCARDIA PACING FOR SUPRAVENTRICULAR TACHYCARDIA. ABBOTT TECHNICAL SUPPORT RECOMMENDED REPROGRAMMING TO RESOLVE THE EVENT, HOWEVER, NO INTERVENTION HAS BEEN PERFORMED AT THIS TIME TO RESOLVE THE EVENT. THE PATIENT WAS IN STABLE CONDITION AND WILL CONTINUE TO BE MONITORED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18043183 ↗MalfunctionAug 17, 2022Over-Sensing; Device Sensing Problem; Over-Sensing; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED REMOTELY VIA MERLIN.NET. UPON REVIEW OF TRANSMISSION, IT WAS FOUND THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED R-WAVE AMPLITUDE VARIATION AND MYOPOTENTIAL OVER-SENSING. PROGRAMMING CHANGES WERE MADE. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15245521 ↗MalfunctionApr 1, 2022Thermal Decomposition of Device; Smoking; Thermal Decomposition of Device; Smoking
IT WAS REPORTED THAT THE GREEN STRIPES ON THE TRANSMITTER'S ALTERNATING CURRENT (AC) POWER ADAPTER LOOKED BURNT AND VISIBLE SMOKE WAS OBSERVED. THE AC POWER ADAPTER WAS REPLACED. THERE WERE NO REPORTED PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13981339 ↗MalfunctionOct 5, 2021Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset
PRIOR TO AN DEVICE REPLACEMENT DUE TO NORMAL ELECTIVE REPLACEMENT INDICATOR (ERI), THE DEVICE WAS INTERROGATED AND WAS OBSERVED TO BE IN BACKUP MODE. THE EVENT WAS RESOLVED BY EXPLANTING AND REPLACING THE DEVICE DUE TO NORMAL ERI. THE PATIENT WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12577571 ↗MalfunctionSep 9, 2021Pacing Problem; Pacing Problem
DURING A CLINIC FOLLOW-UP, THE LOW VOLTAGE OUTPUT PACING ON THE DEVICE WAS LOWER THAN THE PROGRAMMING MINIMUM VOLTAGE. THE DEVICE WAS REPROGRAMMING AS AN INTERIM RESOLUTION AS THE DEVICE IS ANTICIPATED TO BE EXPLANTED DUE TO NORMAL ELECTIVE REPLACEMENT INDICATOR. THE PATIENT WAS STABLE AND WILL CONTINUE TO BE MONITORED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12447194 ↗MalfunctionMar 2, 2021Thermal Decomposition of Device; Thermal Decomposition of Device
IT WAS REPORTED THAT A BURN MARK WAS OBSERVED AROUND THE USB CORD OF CELL ADAPTER. THE DEVICE WOULD BE REPLACED. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11400404 ↗MalfunctionDec 27, 2019Melted; Melted
RELATED MANUFACTURER REFERENCE NUMBER: 2017865-2019-18569. IT WAS REPORTED THAT DURING A SCHEDULED DATA TRANSMISSION, THE MERLIN TRANSMITTER USB PORT AND USB CELLULAR ADAPTER WERE SUSPECTED TO HAVE BEEN MELTED DUE TO HEAD CAUSED BY LONGTIME USE. THE DEVICES WERE REPLACED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9527813 ↗
Other Pulse generator, permanent, implantable devices
- Assurity™ PM2240St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall84 deaths4,076 reports
- Fortify Assura™ CD2357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable63 deaths3,571 reports
- Advisa DR MRI™ SureScan® A2DR01Medtronic, Inc. · Pulse generator, permanent, implantable51 deaths2,298 reports
- Fortify Assura™ CD1357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable35 deaths1,933 reports
- Ellipse™ CD2411-36QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable42 deaths1,498 reports
- Fortify Assura™ CD2357-40CSt. Jude Medical, Inc. · Pulse generator, permanent, implantable25 deaths1,378 reports
- Accent® DR RFSt. Jude Medical, Inc. · Pulse generator, permanent, implantable73 deaths1,370 reports
- Edora 8 DR-T 407145BIOTRONIK SE & Co. KG · Pulse generator, permanent, implantable10 deaths982 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page