Merlin@home™ EX1151Recalls & reported problems

St. Jude Medical, Inc. · Wireless USB Adapter for use with Merlin@home Transmitter · Class 3

All models of Merlin@home (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 05414734507011), received between Dec 27, 2019 and Nov 27, 2024 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Thermal Decomposition of Device” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Thermal Decomposition of Device3
  • Device Sensing Problem1
  • Inappropriate or Unexpected Reset1
  • Inappropriate/Inadequate Shock/Stimulation1
  • Incorrect Interpretation of Signal1
  • Melted1
  • Over-Sensing1
  • Pacing Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 27, 2024Thermal Decomposition of Device; Thermal Decomposition of Device

    IT WAS REPORTED THAT BURN MARKS WERE OBSERVED ON THE TRANSMITTER'S USB CELLULAR ADAPTER. THE TRANSMITTER WITH ADAPTER WAS EXCHANGED. THERE WERE NO REPORTED PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20802693 ↗
  • InjuryOct 31, 2023Incorrect Interpretation of Signal; Inappropriate/Inadequate Shock/Stimulation; Incorrect Interpretation of Signal; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT THE DEVICE INAPPROPRIATELY DELIVERED ANTITACHYCARDIA PACING FOR SUPRAVENTRICULAR TACHYCARDIA. ABBOTT TECHNICAL SUPPORT RECOMMENDED REPROGRAMMING TO RESOLVE THE EVENT, HOWEVER, NO INTERVENTION HAS BEEN PERFORMED AT THIS TIME TO RESOLVE THE EVENT. THE PATIENT WAS IN STABLE CONDITION AND WILL CONTINUE TO BE MONITORED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18043183 ↗
  • MalfunctionAug 17, 2022Over-Sensing; Device Sensing Problem; Over-Sensing; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED REMOTELY VIA MERLIN.NET. UPON REVIEW OF TRANSMISSION, IT WAS FOUND THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED R-WAVE AMPLITUDE VARIATION AND MYOPOTENTIAL OVER-SENSING. PROGRAMMING CHANGES WERE MADE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15245521 ↗
  • MalfunctionApr 1, 2022Thermal Decomposition of Device; Smoking; Thermal Decomposition of Device; Smoking

    IT WAS REPORTED THAT THE GREEN STRIPES ON THE TRANSMITTER'S ALTERNATING CURRENT (AC) POWER ADAPTER LOOKED BURNT AND VISIBLE SMOKE WAS OBSERVED. THE AC POWER ADAPTER WAS REPLACED. THERE WERE NO REPORTED PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13981339 ↗
  • MalfunctionOct 5, 2021Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    PRIOR TO AN DEVICE REPLACEMENT DUE TO NORMAL ELECTIVE REPLACEMENT INDICATOR (ERI), THE DEVICE WAS INTERROGATED AND WAS OBSERVED TO BE IN BACKUP MODE. THE EVENT WAS RESOLVED BY EXPLANTING AND REPLACING THE DEVICE DUE TO NORMAL ERI. THE PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12577571 ↗
  • MalfunctionSep 9, 2021Pacing Problem; Pacing Problem

    DURING A CLINIC FOLLOW-UP, THE LOW VOLTAGE OUTPUT PACING ON THE DEVICE WAS LOWER THAN THE PROGRAMMING MINIMUM VOLTAGE. THE DEVICE WAS REPROGRAMMING AS AN INTERIM RESOLUTION AS THE DEVICE IS ANTICIPATED TO BE EXPLANTED DUE TO NORMAL ELECTIVE REPLACEMENT INDICATOR. THE PATIENT WAS STABLE AND WILL CONTINUE TO BE MONITORED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12447194 ↗
  • MalfunctionMar 2, 2021Thermal Decomposition of Device; Thermal Decomposition of Device

    IT WAS REPORTED THAT A BURN MARK WAS OBSERVED AROUND THE USB CORD OF CELL ADAPTER. THE DEVICE WOULD BE REPLACED. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11400404 ↗
  • MalfunctionDec 27, 2019Melted; Melted

    RELATED MANUFACTURER REFERENCE NUMBER: 2017865-2019-18569. IT WAS REPORTED THAT DURING A SCHEDULED DATA TRANSMISSION, THE MERLIN TRANSMITTER USB PORT AND USB CELLULAR ADAPTER WERE SUSPECTED TO HAVE BEEN MELTED DUE TO HEAD CAUSED BY LONGTIME USE. THE DEVICES WERE REPLACED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9527813 ↗

Other Pulse generator, permanent, implantable devices

All Pulse generator, permanent, implantable devices →