QUADRA ALLURE MP™ PM3160Recalls & reported problems

St. Jude Medical, Inc. · Cardiac Resynchronization Therapy Device, Pulse Generator DDDRV · Class 3 · Implant

All models of QUADRA ALLURE MP (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05414734509084), received between Aug 23, 2022 and Jan 24, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Defective Device” (1 report).

5 other recalls cover devices cleared under the same FDA submission (P030035); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Defective Device1
  • Loose or Intermittent Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Unspecified Infection1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 24, 2024Loose or Intermittent Connection; Loose or Intermittent Connection

    RELATED MANUFACTURER REFERENCE NUMBER: 2017865-2024-01749. IT WAS REPORTED THAT THE PATIENT PRESENTED FOR A GENERATOR CHANGE PROCEDURE DUE TO NORMAL BATTERY DEPLETION. DURING THE PROCEDURE, IT WAS NOTED THAT THE CHRONIC LEFT VENTRICULAR (LV) LEAD EXHIBITED AN ELEVATED CAPTURE THRESHOLD. THE PHYSICIAN INTENDED TO REPLACE THE LV LEAD. DURING THE REPLACEMENT, IT WAS FOUND THAT THE GUIDEWIRE WAS FAILING TO BE REMOVED FROM THE NEW LV LEAD. AN ADDITIONAL ATTEMPT TO REMOVE THE GUIDEWIRE DAMAGED THE NEW LV LEAD. THUS, NO NEW LV LEAD WAS IMPLANTED. PROGRAMMING CHANGES WERE MADE ON THE CHRONIC LV LEAD TO ADDRESS HIGH CAPTURE THRESHOLD. WHEN THE PHYSICIAN ATTEMPTED TO CONNECT THE CHRONIC LV LEAD TO THE NEW PACEMAKER, IT WAS FOUND THAT THE PACEMAKER SETSCREW EXHIBITED LOOSE CONNECTIONS. THE PACEMAKER WAS NOT USED. THE PHYSICIAN CONTINUED WITH SECOND NEW PACEMAKER AND COMPLETED THE PROCEDURE. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18568151 ↗
  • InjuryJul 19, 2023Defective Device; Defective Device

    RELATED MANUFACTURER REFERENCE NUMBER: 2017865-2023-36102 IT WAS REPORTED THAT THE PATIENT PRESENTED FOR A FOLLOW-UP IN CLINIC. UPON INTERROGATION, IT WAS NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED FAILURE TO CAPTURE, FAILURE TO SENSE AND LOW PACING IMPEDANCE WHICH MAY BE DUE TO HEADER ANOMALY ON THE PACEMAKER. THE PACEMAKER AND THE RV LEAD WERE EXPLANTED AND REPLACED. THE PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17355920 ↗
  • InjuryAug 23, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RELATED MANUFACTURER REPORT NUMBER: 2017865-2022-21222; RELATED MANUFACTURER REPORT NUMBER: 2017865-2022-21223; RELATED MANUFACTURER REPORT NUMBER: 2017865-2022-21225. IT WAS REPORTED THAT THE PATIENT PRESENTED FOR FOLLOW-UP WITH DISCHARGE AT THE IMPLANT SITE OF THE PULSE GENERATOR. THE GENERATOR, RIGHT ATRIAL, RIGHT VENTRICULAR, AND LEFT VENTRICULAR LEADS WERE EXPLANTED DUE TO POCKET INFECTION. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: Unspecified Infection

    MAUDE report 15277994 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030035. One submission often covers many models, so check the product and lot details in each record.

  • Z-1497-2022 · Jun 17, 2022 · Open, Classified
    Merlin.net model MN5000 Software
  • Z-1496-2022 · Jun 17, 2022 · Open, Classified
    Merlin 2 PCS MER37000 programmer Model MER3400 software
  • Z-1495-2022 · Jun 17, 2022 · Open, Classified
    Merlin PCS 3650 programmer Model 3330 software
  • Z-0430-2012 · Sep 23, 2011 · Terminated
    Anthem CRT-P models PM3110 and PM3210 St. Jude Medical pacemakers monitor and regulate a patient's heart rate with pacing therapy.
  • Z-0429-2012 · Sep 23, 2011 · Terminated
    Implantable Pacemaker Accent DR models PM2110 and PM2210 St. Jude Medical pacemakers monitor and regulate a patient's heart rate with pacing therapy.

Other Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P) devices

All Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P) devices →