DR 400 5520/100Recalls & reported problems
Agfa · DR 400 SYSTEM · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05414904176252), received between Jan 3, 2025 and Jul 30, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 30, 2025Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
A CUSTOMER IN THE US REPORTED THE COLLIMATOR FELL OFF A DR 400 SYSTEM WHILE THE CUSTOMER WAS ROTATING THE COLLIMATOR. THE LOCAL AGFA FIELD SERVICE ENGINEER RESPONDED AND CONFIRMED A PORTION OF THE RESIN FLANGE WAS BROKEN. A NEW COLLIMATOR AND METAL FLANGE WAS INSTALLED. PER THE CUSTOMER REQUEST A SYSTEM PREVENTIVE MAINTENANCE (PM) WAS ALSO PERFORMED. THE SYSTEM IS NOW WORKING AS INTENDED AND THERE HAVE BEEN NO ADDITONAL EVENTS REPORTED. THERE WAS NO REPORTED HARM TO USER OR PATIENTS AND THERE ARE NO FUTURE ACTIONS RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22670830 ↗MalfunctionJan 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
A CUSTOMER IN GERMANY REPORTED THE COLLIMATOR FELL OFF A DR 400 SYSTEM. THE CUSTOMER WAS TURNING THE TUBE FROM THE TABLE TO THE WALLSTAND AND THE PLASTIC FLANGE BROKE AND THE COLLIMATOR FELL FROM THE SYSTEM. THE LOCAL AGFA FIELD SERVICE ENGINEER WAS AT THE CUSTOMER SITE COMPLETING AN UNRELATED SERVICE AND WAS ABLE TO REVIEW THE BROKEN FLANGE. THE LOCAL SERVICE ENGINEER REPLACED THE BROKEN FLANGE AND REATTACHED THE SYSTEM COLIMATOR. THERE WERE NO REPORTED INJURIES TO USER OR PATIENTS. THE SYSTEM IS NOW WORKING AS INTENDED AND THERE ARE NO FUTURE ACTIONS RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21069400 ↗
Other Solid state x-ray imager (flat panel/digital imager) devices
- SEDECAL SM-40HF-B-D-CSedecal · Solid State X-Ray Imager (Flat Panel/Digital Imager)24 recalls1 report
- SEDECAL SM-40HF-B-D-VIRSedecal · Solid State X-Ray Imager (Flat Panel/Digital Imager)12 recalls1 report
- Micro-X RoverMicro-X Limited · Solid State X-Ray Imager (Flat Panel/Digital Imager)7 recalls0 reports
- DigitalDiagnost C90 712034Philips Medical Systems DMC GmbH · Solid state x-ray imager (flat panel/digital imager)3 recalls90 reports
- DigitalDiagnost C90 712035Philips Medical Systems DMC GmbH · Solid state x-ray imager (flat panel/digital imager)4 recalls17 reports
- DigitalDiagnost 712031Philips Medical Systems DMC GmbH · Solid state x-ray imager (flat panel/digital imager)2 recalls65 reports
- NA GM85Samsung Electronics Co., Ltd. · Solid State X-Ray Imager (Flat Panel/Digital Imager)3 recalls4 reports
- Micro-X RoverMicro-X Limited · Solid State X-Ray Imager (Flat Panel/Digital Imager)3 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.