DR 400 5520/100Recalls & reported problems

Agfa · DR 400 SYSTEM · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05414904176252), received between Jan 3, 2025 and Jul 30, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 30, 2025Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component

    A CUSTOMER IN THE US REPORTED THE COLLIMATOR FELL OFF A DR 400 SYSTEM WHILE THE CUSTOMER WAS ROTATING THE COLLIMATOR. THE LOCAL AGFA FIELD SERVICE ENGINEER RESPONDED AND CONFIRMED A PORTION OF THE RESIN FLANGE WAS BROKEN. A NEW COLLIMATOR AND METAL FLANGE WAS INSTALLED. PER THE CUSTOMER REQUEST A SYSTEM PREVENTIVE MAINTENANCE (PM) WAS ALSO PERFORMED. THE SYSTEM IS NOW WORKING AS INTENDED AND THERE HAVE BEEN NO ADDITONAL EVENTS REPORTED. THERE WAS NO REPORTED HARM TO USER OR PATIENTS AND THERE ARE NO FUTURE ACTIONS RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22670830 ↗
  • MalfunctionJan 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    A CUSTOMER IN GERMANY REPORTED THE COLLIMATOR FELL OFF A DR 400 SYSTEM. THE CUSTOMER WAS TURNING THE TUBE FROM THE TABLE TO THE WALLSTAND AND THE PLASTIC FLANGE BROKE AND THE COLLIMATOR FELL FROM THE SYSTEM. THE LOCAL AGFA FIELD SERVICE ENGINEER WAS AT THE CUSTOMER SITE COMPLETING AN UNRELATED SERVICE AND WAS ABLE TO REVIEW THE BROKEN FLANGE. THE LOCAL SERVICE ENGINEER REPLACED THE BROKEN FLANGE AND REATTACHED THE SYSTEM COLIMATOR. THERE WERE NO REPORTED INJURIES TO USER OR PATIENTS. THE SYSTEM IS NOW WORKING AS INTENDED AND THERE ARE NO FUTURE ACTIONS RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21069400 ↗

Other Solid state x-ray imager (flat panel/digital imager) devices