Lamitrode Tripole™ 3219 (UDI 05415067017628)Recalls & reported problems

St. Jude Medical, Inc. · 16 Lead, 60 cm · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
22
Deaths reported
0
Injuries reported
22
Recalls naming it
0

Summary

The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 05415067017628), received between Dec 20, 2018 and Mar 11, 2026 — classified as 22 injuries.

The most frequently reported specific device problem is “High impedance” (8 reports).

14 other recalls cover devices cleared under the same FDA submission (P010032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High impedance8
  • Therapeutic or Diagnostic Output Failure4
  • Adverse Event Without Identified Device or Use Problem3
  • Impedance Problem2
  • Insufficient Device Problem Information2
  • Low impedance1
  • Material Fragmentation1
  • Material Split, Cut or Torn1

Patient problems

  • Inadequate Pain Relief12
  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information2
  • Therapeutic Effects, Unexpected1
  • Undesired Nerve Stimulation1
  • Unspecified Infection1

Recent adverse event reports

The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 11, 2026High impedance

    ADDITIONAL INFORMATION INDICATES SURGICAL INTERVENTION WAS UNDERTAKEN ON 01 APRIL 2026 WHEREIN LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24570159 ↗
  • InjuryFeb 19, 2026High impedance

    IT WAS REPORTED THE PATIENT'S IPG WAS INOPERABLE AND UNABLE TO ENTER MRI MODE DUE TO HIGH IMPEDANCES. IN TURN, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEAD AND IPG WERE EXPLANTED AND REPLACED. EFFECTIVE THERAPY WAS ESTABLISHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24387359 ↗
  • InjuryDec 12, 2025Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT'S LEAD WAS EXPLANTED AN REPLACED FOR AN UNKNOWN REASON.

    Patient problems: Inadequate Pain Relief

    MAUDE report 23792707 ↗
  • InjuryFeb 7, 2025High impedance; High impedance

    DIAGNOSTICS REVEALED HIGH IMPEDANCE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21333915 ↗
  • InjuryJan 9, 2024High impedance; Impedance Problem; High impedance; Impedance Problem

    ADDITIONAL INFORMATION WAS RECEIVED STATING PRIOR TO THE SURGERY THE PATIENT HAD HIGH IMPEDANCES. ALSO, SURGICAL INTERVENTION TOOK PLACE WHERE THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. EFFECTIVE STIMULATION WAS RESTORED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 18475943 ↗
  • InjuryOct 6, 2023High impedance; High impedance

    IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING IMPEDANCES ISSUES. SURGICAL INTERVENTION MAY TAKE PLACE AT A LATER DATE TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 17885361 ↗
  • InjurySep 25, 2023High impedance; High impedance

    ADDITIONAL INFORMATION RECEIVED INDICATES THAT SURGICAL INTERVENTION TOOK PLACE WHERE THE PATIENTS LEAD WAS EXPLANTED AND REPLACED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 17807812 ↗
  • InjuryJul 5, 2023Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT PATIENT HAS BEEN GETTING NO THERAPY RELIEF. SURGICAL INTERVENTION MAY TAKE PLACE AT A LATER DATE TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 17254941 ↗
  • InjuryMay 30, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE PATIENT'S LEADS WERE EXPLANTED AFTER BEING CUT DURING A PRIOR SURGERY. THE LEAD WAS REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17022253 ↗
  • InjuryMay 8, 2023High impedance; High impedance

    IT WAS REPORTED THE PATIENT EXPERIENCED UNDESIRED NERVE STIMULATION WITH THEIR SCS SYSTEM. IN TURN, REPROGRAMMING WAS UNABLE TO RESOLVE THE ISSUE. AS A RESULT, SURGICAL INTERVENTION MAY BE PENDING TO ADDRESS THE ISSUE.

    Patient problems: Undesired Nerve Stimulation

    MAUDE report 16885102 ↗
  • InjuryFeb 27, 2023Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON 8 MAR 2023 WHEREIN NEW LEAD WAS IMPLANTED TO ADDRESS THE ISSUE. EFFECTIVE THERAPY WAS RESTORED POST OPERATIVELY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 16449173 ↗
  • InjuryJan 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED AN INFECTION WAS PRESENT AT THE LEAD INSERTION SITE. PATIENT WAS PRESCRIBED ORAL ANTIBIOTICS. PATIENT IS TO FOLLOW-UP WITH PHYSICIAN AS THE NEXT COURSE OF ACTION.

    Patient problems: Unspecified Infection

    MAUDE report 16108974 ↗
  • InjuryJan 3, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2023-00047. IT WAS REPORTED THAT PATIENT UNDERWENT SURGICAL INTERVENTION AT UNKNOWN DATE FOR UNKNOWN REASON WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED.

    Patient problems: Insufficient Information

    MAUDE report 16092002 ↗
  • InjuryDec 23, 2022Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE PATIENT¿S LEAD WAS ACCIDENTALLY CUT DURING A PROCEDURE. IN TURN SURGICAL INTERVENTION MAY OCCUR TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 16042520 ↗
  • InjuryDec 7, 2022Use of Device Problem; Impedance Problem; Use of Device Problem; Impedance Problem

    IT WAS REPORTED THAT THE PATIENT LEAD WAS ABANDONED AND LEFT IMPLANTED DURING A SYSTEM REVISION ON (B)(6) 2022. NO INTERVENTION IS PLANNED TO RETRIEVE THE LEAD.

    Patient problems: Inadequate Pain Relief

    MAUDE report 15937244 ↗
  • InjuryNov 17, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    RELATED MANUFACTURER REPORT NUMBER: 1627487-2022-06261. IT WAS REPORTED THE PATIENT DID NOT RECHARGE THE IPG AS RECOMMENDED. AS SUCH, THE IPG BECAME INOPERABLE. IN TURN, THE PATIENT UNDERWENT SURGICAL INTERVENTION WHEREIN THE PATIENT'S IPG WAS EXPLANTED AND REPLACED. IT WAS REPORTED THE PATIENT'S LEAD WAS EXPLANTED AND REPLACED FOR AN UNKNOWN REASON.

    Patient problems: Insufficient Information

    MAUDE report 15819737 ↗
  • InjuryOct 19, 2022Low impedance; Low impedance

    RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2022-04989. IT WAS REPORTED THE PATIENT HAD EFFECTIVE THERAPY, BUT LOW IMPEDANCES ON CONTACTS 8-15. AS A RESULT, THE LEAD WAS REPLACED SO THE PATIENT COULD UTILIZE MRI MODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15635937 ↗
  • InjuryJul 11, 2022High impedance; High impedance

    IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY. SYSTEM DIAGNOSTICS RECORDED HIGH IMPEDANCE ON ALL LEAD CONTACTS. AS A RESULT, SURGERY INTERVENTION WAS UNDERTAKEN WHEREIN THE LEAD WAS EXPLANTED AND REPLACED. EFFECTIVE THERAPY WAS RESTORED POST OPERATIVELY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 14977444 ↗
  • InjuryMar 29, 2022Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    RELATED MANUFACTURER REFERENCE NUMBER: 3006705815-2022-01608. IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION WITH THE IMPLANTED LEAD AND THE IPG HAD REACHED END OF LIFE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE IPG AND LEAD WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 13940938 ↗
  • InjuryNov 12, 2020Migration; Migration

    IT WAS REPORTED THAT PATIENT EXPERIENCED A FALL WHICH RESULTED TO LEAD MIGRATION. AS A RESULT THE SURGICAL INTERVENTION WAS UNDERTAKEN WHERE IN THE LEAD WAS EXPLANTED AND REPLACED ADDRESSING THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 10827135 ↗
  • InjuryJan 14, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE 2 OF 2: REFERENCE MFR. REPORT: 3006705815-2019-00210. IT WAS REPORTED THE PATIENT EXPERIENCED INADEQUATE RELIEF. AS A RESULT, THE SYSTEM WAS EXPLANTED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 8246486 ↗
  • InjuryDec 20, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE 1 OF 2: REFERENCE MFR. REPORT# 3006705815-2018-03418. IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING STIMULATION IN THE WRONG LOCATION. IN TURN, THE PATIENTS SYSTEM WAS EXPLANTED AND REPLACED. EFFECTIVE STIMULATION WAS CONFIRMED POST-OP.

    Patient problems: Therapeutic Effects, Unexpected

    MAUDE report 8185731 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P010032. One submission often covers many models, so check the product and lot details in each record.

  • Z-0458-2026 · Oct 6, 2025 · Open, Classified
    Eterna SCS IPG (Implantable Pulse Generator), Model Number 32400
  • Z-2511-2023 · Jul 18, 2023 · Open, Classified
    Infinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2507-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2510-2023 · Jul 18, 2023 · Open, Classified
    Infinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2508-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2505-2023 · Jul 18, 2023 · Open, Classified
    Proclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
  • Z-2506-2023 · Jul 18, 2023 · Open, Classified
    Proclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
  • Z-2509-2023 · Jul 18, 2023 · Open, Classified
    Proclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
  • Z-2691-2017 · Jun 2, 2017 · Terminated
    Proclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-2692-2017 · Jun 2, 2017 · Terminated
    Proclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-1785-2013 · Nov 30, 2012 · Terminated
    Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
  • Z-0025-2013 · Sep 4, 2012 · Terminated
    The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
  • Z-1738-2013 · Jul 26, 2012 · Terminated
    Eon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
  • Z-3091-2011 · Jun 2, 2009 · Terminated
    Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im

Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices

All STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices →