Exclaim™ 3225 (UDI 05415067017758)Recalls & reported problems

St. Jude Medical, Inc. · Lead, 90 cm · Class 3 · Implant

Quick summary

This St. Jude Medical Exclaim 3225 is an implanted spinal cord stimulator lead, a thin wire placed under the skin to send mild pulses to the spine for pain relief. Most of the 32 reports describe pain relief fading because the lead stopped working, moved out of position, or broke, with several patients needing surgery to remove and replace the lead or the whole system. All 32 reports involve injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
32
Deaths reported
0
Injuries reported
32
Recalls naming it
0

Summary

The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 05415067017758), received between Jul 17, 2018 and Apr 29, 2026 — classified as 32 injuries.

The most frequently reported specific device problem is “High impedance” (14 reports).

14 other recalls cover devices cleared under the same FDA submission (K063080, P010032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High impedance14
  • Adverse Event Without Identified Device or Use Problem8
  • Migration7
  • Fracture6
  • Break1
  • Failure to Deliver Energy1
  • Therapeutic or Diagnostic Output Failure1

Patient problems

  • Inadequate Pain Relief29
  • Failure of Implant2
  • No Clinical Signs, Symptoms or Conditions2
  • Inflammation1
  • Unspecified Infection1

Recent adverse event reports

The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 29, 2026High impedance

    IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE THERAPY. DIAGNOSTICS SHOWED HIGH IMPEDANCES. AS A RESULT, THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 25033478 ↗
  • InjurySep 2, 2025High impedance; High impedance

    ADDITIONAL INFORMATION INDICATES SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE ENTIRE SYSTEM WAS REPLACED TO ADDRESS THE ISSUE. THERAPY WAS RESTORED POST-OPERATIVELY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 22948021 ↗
  • InjuryJul 26, 2025Migration; Migration

    IT WAS REPORTED THAT THE PATIENT'S CERVICAL PADDLE LEAD HAD MIGRATED. AS A RESULT, SURGICAL INTERVENTION TOOK PLACE ON (B)(6) 2025, WHEREIN THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22621773 ↗
  • InjuryFeb 1, 2024High impedance; High impedance

    IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE THERAPY/COVERAGE DUE TO HIGH IMPEDANCE ON THE LEAD. AS SUCH, SURGICAL INTERVENTION TOOK PLACE ON (B)(6) 2024 WHEREIN THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. REPORTEDLY, THERAPY WAS CONFIRMED POST OPERATIVELY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 18622595 ↗
  • InjurySep 11, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING AN IPG REPLACEMENT, THE PHYSICIAN NOTICED THE LEAD WAS FRACTURED. AS A RESULT, THE PHYSICIAN IS REFERRING THE PATIENT TO A SURGEON FOR A LEAD REPLACEMENT TO ADDRESS THE ISSUE.

    Patient problems: Failure of Implant; Inadequate Pain Relief

    MAUDE report 17725323 ↗
  • InjuryJun 15, 2023Migration; Migration

    ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE PATIENT UNDERWENT SURGICAL INTERVENTION ON (B)(6) 2023 WHEREIN THE LEAD WAS EXPLANTED AND REPLACED RESTORING THERAPY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 17133581 ↗
  • InjuryMay 25, 2023Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT'S CERVICAL SCS LEAD HAD FRACTURED. SURGICAL INTERVENTION MAY TAKE PLACE AT A LATER DATE TO ADDRESS THE ISSUE.

    Patient problems: Failure of Implant; Inadequate Pain Relief

    MAUDE report 17005326 ↗
  • InjuryJan 3, 2023Fracture; Fracture

    IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY DUE TO LEAD FRACTURE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN AT UNKNOWN DATE WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED. THE INVESTIGATION WAS UNABLE TO DETERMINE THE LEAD THAT WAS FRACTURED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 16086683 ↗
  • InjuryDec 29, 2022Migration; Migration

    RELATED MANUFACTURER REPORT NUMBER: 1627487-2023-02147, 1627487-2023-02148. ADDITIONAL INFORMATION RECEIVED REPORTS THAT THE PATIENT UNDERWENT SURGICAL INTERVENTION ON (B)(6) 2023 IN WHICH THEIR PADDLE LEAD WAS EXPLANTED AND REPLACED. DURING THE PROCEDURE THE PATIENT'S QUATTRODE LEADS WERE ALSO REMOVED DUE TO SCARRING AROUND THE LEADS.

    Patient problems: Inadequate Pain Relief

    MAUDE report 16070333 ↗
  • InjuryDec 29, 2022High impedance; High impedance

    IT WAS REPORTED THE PATIENT IS NOT RECEIVING EFFECTIVE THERAPY DUE TO HIGH IMPEDANCES. SURGICAL INTERVENTION TOOK PLACE IN WHERE THE LEAD WAS REPLACED. RELATED MANUFACTURER REFERENCE NUMBER: (B)(4).

    Patient problems: Inadequate Pain Relief

    MAUDE report 16068455 ↗
  • InjuryApr 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RELATED MANUFACTURER REFERENCE NUMBER: 3006705815-2022-13600. IT WAS REPORTED THE PATIENT EXPERIENCED AN INFECTION AT THE IPG AND LEAD SITE. SURGICAL INTERVENTION TOOK PLACE WHEREIN THE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Unspecified Infection

    MAUDE report 14043594 ↗
  • InjuryDec 10, 2021High impedance; High impedance

    ADDITIONAL INFORMATION RECEIVED INDICATES THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION. SURGICAL INTERVENTION TOOK PLACE WHEREIN THE LEAD WAS EXPLANTED AND REPLACED. EFFECTIVE THERAPY WAS ESTABLISHED.

    Patient problems: Inadequate Pain Relief; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12976023 ↗
  • InjuryApr 22, 2021High impedance; Migration; High impedance; Migration

    IT WAS REPORTED THAT THE PATIENT HAD A CAR ACCIDENT, THE LEADS MIGRATED, AND HIGH IMPEDANCES WERE MEASURED AT THE LEAD CONTACTS. X-RAY IMAGING CONFIRMED MIGRATION OCCURRED, AND THE BOTTOM OF THE LEAD MOVED OUT OF THE EPIDURAL SPACE. AS A RESULT, SURGERY OCCURRED IN WHICH THE LEAD WAS EXPLANTED AND REPLACED, WHICH RESOLVED THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 11704867 ↗
  • InjuryJan 19, 2021Migration; Migration

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY DUE TO THE LEAD MIGRATING OUT OF THE EPIDURAL SPACE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2021 WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 11196794 ↗
  • InjuryDec 23, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REFERENCE MANUFACTURER NUMBER: 3006705815-2020-33390. IT WAS REPORTED THAT THE PATIENT'S BODY REJECTED THEIR SCS SYSTEM LIKELY DUE TO AN ALLERGIC RESPONSE. IN TURN, THE PATIENT UNDERWENT SURGICAL INTERVENTION WHEREIN THE DEVICES WERE EXPLANTED ON AN UNKNOWN DATE TO ADDRESS THE ISSUE. NOTE: SINCE IT IS NOT KNOWN WHICH DEVICE CONTRIBUTED TO THE ISSUE, ALL POSSIBLE DEVICES ARE BEING REPORTED ON.

    Patient problems: Inflammation

    MAUDE report 11065821 ↗
  • InjuryNov 11, 2020Break; Therapeutic or Diagnostic Output Failure; Break; Therapeutic or Diagnostic Output Failure

    FOLLOW UP INFORMATION RECEIVED THAT THE PATIENT HAD THE LEAD EXPLANTED AND REPLACED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 10826126 ↗
  • InjuryNov 11, 2020Migration; Migration

    IT WAS REPORTED THAT THE PATIENT'S CERVICAL PADDLE LEAD HAD MIGRATED. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEAD WAS EXPLANTED AND REPLACED. POST-OPERATIVELY, STIMULATION THERAPY WAS RESTORED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 10822803 ↗
  • InjuryNov 8, 2020Fracture; High impedance; Migration; Fracture; High impedance; Migration

    ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE PATIENT LEAD WAS EXPLANTED AND REPLACED ON (B)(6) 2021. THE PHYSICIAN NOTED THAT THE LEAD WAS FRACTURED. EFFECTIVE THERAPY WAS ESTABLISHED POST-OP.

    Patient problems: Inadequate Pain Relief

    MAUDE report 10807021 ↗
  • InjuryOct 13, 2020High impedance; High impedance

    IT WAS REPORTED THAT THE PATIENT'S SYSTEM WAS AUTO REDUCING DUE TO HIGH IMPEDANCES. IN TURN, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEAD WAS EXPLANTED AND REPLACED. POSTOPERATIVELY, THE PATIENT HAS EFFECTIVE THERAPY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 10673277 ↗
  • InjurySep 22, 2020Failure to Deliver Energy; High impedance; Failure to Deliver Energy; High impedance

    ADDITIONAL INFORMATION INDICATES THE LEAD WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 10568249 ↗
  • InjuryAug 17, 2020Fracture; High impedance; Fracture; High impedance

    MANUFACTURER REPORT REFERENCE NUMBER: 1627487-2020-24094, 1627487-2020-24096. IT WAS REPORTED THAT THE PATIENT EXPERIENCED HIGH IMPEDANCE. AS A RESULT, THE PATIENT UNDERWENT SURGICAL INTERVENTION IN WHICH THE LEADS WERE EXPLANTED AND REPLACED WITH ONE PENTA LEAD. DURING THE PROCEDURE IT WAS NOTED THAT THE EXCLAIM LEADS WERE VISIBLY FRACTURED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 10416237 ↗
  • InjuryAug 4, 2020High impedance; High impedance

    IT WAS REPORTED THE PATIENT WAS NOT RECEIVING EFFECTIVE STIMULATION AND DIAGNOSTICS INDICATED HIGH IMPEDANCES. IN TURN, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEAD WAS EXPLANTED. THIS ADDRESSED THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 10364254 ↗
  • InjuryJul 28, 2020Fracture; Fracture

    IT WAS REPORTED THE PATIENT¿S SYSTEM WAS EXPLANTED AND REPLACED DUE TO A LEAD FRACTURE. EFFECTIVE THERAPY WAS RESTORED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 10334034 ↗
  • InjuryJun 10, 2020High impedance; High impedance

    IT WAS REPORTED THE PATIENT'S SCS LEAD WAS EXHIBITING HIGH IMPEDANCE ON MULTIPLE CONTACTS. REPORTEDLY, THE PATIENT STATED THEY HAD A SEIZURE APPROXIMATELY THREE MONTHS PRIOR AND FELL, THE SCS SYSTEM HAS NOT WORKED SINCE. CONSEQUENTLY, SURGICAL INTERVENTION WAS TAKEN AND THE PATIENT'S SCS LEAD WAS SUCCESSFULLY REPLACED WITH A NEW ONE AND THE REPORTED ISSUE RESOLVED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 10136578 ↗
  • InjuryAug 2, 2019High impedance; High impedance

    RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2019-08751. IT WAS REPORTED THE PATIENT EXPERIENCED INEFFECTIVE STIMULATION DUE TO HIGH IMPEDANCES. X-RAYS WERE INCONCLUSIVE. AS A RESULT, THE SYSTEM WAS EXPLANTED AND REPLACED. ISSUE RESOLVED

    Patient problems: Inadequate Pain Relief

    MAUDE report 8856276 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K063080, P010032. One submission often covers many models, so check the product and lot details in each record.

  • Z-0458-2026 · Oct 6, 2025 · Open, Classified
    Eterna SCS IPG (Implantable Pulse Generator), Model Number 32400
  • Z-2505-2023 · Jul 18, 2023 · Open, Classified
    Proclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
  • Z-2511-2023 · Jul 18, 2023 · Open, Classified
    Infinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2508-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2507-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2509-2023 · Jul 18, 2023 · Open, Classified
    Proclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
  • Z-2506-2023 · Jul 18, 2023 · Open, Classified
    Proclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
  • Z-2510-2023 · Jul 18, 2023 · Open, Classified
    Infinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2692-2017 · Jun 2, 2017 · Terminated
    Proclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-2691-2017 · Jun 2, 2017 · Terminated
    Proclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-1785-2013 · Nov 30, 2012 · Terminated
    Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
  • Z-0025-2013 · Sep 4, 2012 · Terminated
    The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
  • Z-1738-2013 · Jul 26, 2012 · Terminated
    Eon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
  • Z-3091-2011 · Jun 2, 2009 · Terminated
    Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im

Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices

All STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices →