Protégé™ 3789 (UDI 05415067020833)Recalls & reported problems

St. Jude Medical, Inc. · IPG, 16-Channel Rechargeable · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This St. Jude Protégé spinal cord stimulator is an implanted device that sends electrical pulses to relieve pain. The 446 reports often describe the implanted battery unit losing its ability to communicate with external devices, in some cases after patients did not recharge it, and many describe inadequate pain relief. There are 446 injuries and no deaths, and no recalls name this device. Reports dropped from 27 in the year before to 10 in the most recent year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
446
Deaths reported
0
Injuries reported
446
Recalls naming it
0

Summary

The FDA's MAUDE database contains 446 adverse event reports that list this exact device (DI 05415067020833), received between Jun 11, 2018 and Jun 25, 2026 — classified as 446 injuries.

The most frequently reported specific device problem is “Wireless Communication Problem” (136 reports).

14 other recalls cover devices cleared under the same FDA submission (P010032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 8 reportsJan 2024: 4 reportsFeb 2024: 4 reportsMar 2024: 4 reportsApr 2024: 4 reportsMay 2024: 2 reportsJun 2024: 4 reportsJul 2024: 4 reportsAug 2024: 2 reportsSep 2024: 5 reportsOct 2024: 2 reportsNov 2024: 4 reportsDec 2024: 3 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 2 reportsApr 2025: 4 reportsMay 2025: 4 reportsJun 2025: 4 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 3 reportsMay 2026: 0 reportsJun 2026: 3 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Wireless Communication Problem136
  • Improper or Incorrect Procedure or Method125
  • Adverse Event Without Identified Device or Use Problem53
  • Use of Device Problem50
  • Therapeutic or Diagnostic Output Failure22
  • Insufficient Device Problem Information17
  • Premature Discharge of Battery17
  • No Device Output16

Patient problems

  • Inadequate Pain Relief369
  • Failure of Implant25
  • No Clinical Signs, Symptoms or Conditions16
  • Implant Pain15
  • No Consequences Or Impact To Patient12
  • Insufficient Information9
  • Unspecified Infection8
  • Pain7

Recent adverse event reports

The 25 most recent of 446 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 25, 2026Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT THE PATIENT DID NOT CHARGE THEIR IPG AS RECOMMENDED AND THE IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 25627005 ↗
  • InjuryJun 10, 2026Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT THE PATIENT DID NOT CHARGE THEIR IPG AS RECOMMENDED AND THE IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHERE THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 25459388 ↗
  • InjuryJun 8, 2026Wireless Communication Problem

    IT WAS REPORTED THAT THE PATIENT'S IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. TROUBLESHOOTING HAS BEEN UNABLE TO RESOLVE THE ISSUE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TOA DDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 25437218 ↗
  • InjuryApr 27, 2026Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT THE PATIENT DID NOT CHARGE THEIR IPG AS RECOMMENDED AND THE IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. EFFECTIVE THERAPY WAS RESTORED POST OPERATIVELY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24999075 ↗
  • InjuryApr 7, 2026No Apparent Adverse Event

    ADDITIONAL INFORMATION INDICATES THE IPG MEETS OR EXCEEDS 75% OF ITS EXPECTED LONGEVITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24808163 ↗
  • InjuryApr 6, 2026Wireless Communication Problem

    IT WAS REPORTED THE PATIENT¿S IPG WAS UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. A MANUFACTURER REPRESENTATIVE CONFIRMED THE ISSUE AND THE IPG DEEMED INOPERABLE. SURGICAL INTERVENTION MAY BE TAKEN AT A LATER DATE TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24794159 ↗
  • InjuryMar 30, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AT THE IPG SITE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Implant Pain

    MAUDE report 24714589 ↗
  • InjuryJan 14, 2026Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT THE PATIENT DID NOT CHARGE THEIR IPG AS RECOMMENDED AND THE IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24071490 ↗
  • InjuryDec 2, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT THE PATIENT'S IPG WAS INOPERABLE. SURGICAL INTERVENTION WAS UNDERTAKEN AND THE IPG WAS EXPLANTED AND REPLACED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 23700333 ↗
  • InjuryNov 3, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED PATIENT'S SYSTEM WAS EXPLANTED AT AN UNKNOWN TIME FOR AN UNKNOWN REASON.

    Patient problems: Insufficient Information

    MAUDE report 23457818 ↗
  • InjurySep 2, 2025Premature Discharge of Battery; Premature Discharge of Battery

    IT WAS REPORTED THE PATIENT EXPERIENCED INCREASED IPG RECHARGE BURDEN. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Failure of Implant

    MAUDE report 22947507 ↗
  • InjuryJul 31, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT THE PATIENT DID NOT CHARGE THEIR IPG AS RECOMMENDED AND THE IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. SURGICAL INTERVENTION MAY BE PENDING TO ADDRESS THIS ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 22682891 ↗
  • InjuryJul 22, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT THE PATIENT'S IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. TROUBLESHOOTING HAS BEEN UNABLE TO RESOLVE THE ISSUE. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN IN THE FUTURE.

    Patient problems: Failure of Implant; Inadequate Pain Relief

    MAUDE report 22567976 ↗
  • InjuryJun 19, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT THE PATIENT DID NOT CHARGE THEIR IPG AS RECOMMENDED AND THE IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 22289772 ↗
  • InjuryJun 5, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT THE PATIENT'S IPG WAS UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES AND WAS DEEMED INOPERABLE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE IPG WAS EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant; Inadequate Pain Relief

    MAUDE report 22155451 ↗
  • InjuryJun 4, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT THE PATIENT'S IPG WAS UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE IPG WAS EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant; Inadequate Pain Relief

    MAUDE report 22146910 ↗
  • InjuryJun 2, 2025Premature Discharge of Battery; Wireless Communication Problem; Premature Discharge of Battery; Wireless Communication Problem

    IT WAS REPORTED THE PATIENT¿S IPG HAD AN INCREASED RECHARGE BURDEN. THE IPG WAS SUBSEQUENTLY EXPLANTED AND REPLACED.

    Patient problems: Inadequate Pain Relief; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22122710 ↗
  • InjuryMay 16, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THE SCS IPG WAS NOT FUNCTIONING AND WOULD NOT HOLD A CHARGE. AS SUCH, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE IPG WAS REPLACED AND THERAPY RESTORED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 22037565 ↗
  • InjuryMay 9, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT THE PATIENT¿S IPG HAD REACHED END OF LIFE. IT IS UNKNOWN IF THIS OCCURRED PREMATURELY. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 21997514 ↗
  • InjuryMay 1, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    SURGICAL INTERVENTION MAY BE UNDERTAKEN IN THE FUTURE AND HAS NOT TAKEN PLACE YET.

    Patient problems: Inadequate Pain Relief

    MAUDE report 21947072 ↗
  • InjuryMay 1, 2025Wireless Communication Problem; Wireless Communication Problem

    ADDITIONAL INFORMATION RECEIVED INDICATES THAT SURGICAL INTERVENTION TOOK PLACE WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. REPORTEDLY, THERAPY WAS CONFIRMED POST OPERATIVELY.

    Patient problems: Failure of Implant; Inadequate Pain Relief

    MAUDE report 21944749 ↗
  • InjuryApr 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION INDICATES SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. ADDITIONAL INFORMATION INDICATES, PROGRAMMING WAS DONE AROUND THE IMPEDANCES, AND EFFECTIVE THERAPY HAS BEEN RESTORED. THERE IS NO DEVICE MALFUNCTION; THEREFORE, THE LEADS ARE NO LONGER CONSIDERED REPORTABLE.

    Patient problems: Implant Pain

    MAUDE report 21923463 ↗
  • InjuryApr 28, 2025Unexpected Shutdown; Unexpected Shutdown

    ADDITIONAL INFORMATION INDICATES SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. ADDITIONAL INFORMATION INDICATES, PROGRAMMING WAS DONE AROUND THE IMPEDANCES, AND EFFECTIVE THERAPY HAS BEEN RESTORED. THERE IS NO DEVICE MALFUNCTION; THEREFORE, THE LEADS ARE NO LONGER CONSIDERED REPORTABLE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 21923461 ↗
  • InjuryApr 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AT THE IPG SITE. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE.

    Patient problems: Implant Pain

    MAUDE report 21921291 ↗
  • InjuryApr 2, 2025Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT THE PATIENT'S IPG BECAME UNABLE TO COMMUNICATE WITH EXTERNAL DEVICES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2025 WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 21759070 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P010032. One submission often covers many models, so check the product and lot details in each record.

  • Z-0458-2026 · Oct 6, 2025 · Open, Classified
    Eterna SCS IPG (Implantable Pulse Generator), Model Number 32400
  • Z-2506-2023 · Jul 18, 2023 · Open, Classified
    Proclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
  • Z-2511-2023 · Jul 18, 2023 · Open, Classified
    Infinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2509-2023 · Jul 18, 2023 · Open, Classified
    Proclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
  • Z-2505-2023 · Jul 18, 2023 · Open, Classified
    Proclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
  • Z-2507-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2508-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2510-2023 · Jul 18, 2023 · Open, Classified
    Infinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2691-2017 · Jun 2, 2017 · Terminated
    Proclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-2692-2017 · Jun 2, 2017 · Terminated
    Proclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-1785-2013 · Nov 30, 2012 · Terminated
    Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
  • Z-0025-2013 · Sep 4, 2012 · Terminated
    The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
  • Z-1738-2013 · Jul 26, 2012 · Terminated
    Eon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
  • Z-3091-2011 · Jun 2, 2009 · Terminated
    Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im

Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices

All STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices →