Proclaim™ 3665Recalls & reported problems
St. Jude Medical, Inc. · Implantable Pulse Generator · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 20
- Deaths reported
- 0
- Injuries reported
- 18
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 05415067023247), received between Jul 4, 2018 and Jul 27, 2023 — classified as 18 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Premature Elective Replacement Indicator” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
14 other recalls cover devices cleared under the same FDA submission (P010032); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- May 16, 2024Open, Classified
The duration between the implantable pulse generator reaching the elective replacement indicator threshold and end of service may be shorter than indicated in product labeling.
Z-2362-2024 · Root cause: Vendor change control · FDA record ↗
Most reported problems
Device problems
- Premature Elective Replacement Indicator5
- Wireless Communication Problem4
- Application Program Freezes or Fails to Launch3
- Battery Problem3
- Improper or Incorrect Procedure or Method3
- No Apparent Adverse Event3
- Insufficient Device Problem Information2
- Premature End-of-Life Indicator1
Patient problems
- Inadequate Pain Relief12
- No Consequences Or Impact To Patient6
- No Clinical Signs, Symptoms or Conditions4
- Failure of Implant2
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.
InjuryJul 27, 2023Battery Problem; Battery Problem
IT WAS REPORTED THAT THE PATIENT'S IPG WAS AT END OF LIFE. SURGICAL INTERVENTION WAS UNDERTAKEN AND THE IPG WAS EXPLANTED AND REPLACED.
Patient problems: Inadequate Pain Relief
MAUDE report 17408175 ↗InjuryJan 12, 2023Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THE PATIENT'S IPG WOULD NOT COMMUNICATE WITH THE CHARGER, AND IS UNABLE TO CHARGE. SURGICAL INTERVENTION IS PENDING TO ADDRESS THIS ISSUE.
Patient problems: Inadequate Pain Relief
MAUDE report 16147391 ↗InjuryFeb 3, 2022No Apparent Adverse Event; Wireless Communication Problem; No Apparent Adverse Event; Wireless Communication Problem
IT WAS REPORTED THE PATIENTS IPG REACHED END OF LIFE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.
Patient problems: Inadequate Pain Relief; No Clinical Signs, Symptoms or Conditions
MAUDE report 13440203 ↗InjurySep 15, 2021Battery Problem; No Apparent Adverse Event; Battery Problem; No Apparent Adverse Event
IT WAS REPORTED THE IPG HAD REACHED END OF LIFE. IN TURN, THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.
Patient problems: Failure of Implant; Inadequate Pain Relief; No Clinical Signs, Symptoms or Conditions
MAUDE report 12476272 ↗InjurySep 15, 2021Battery Problem; No Apparent Adverse Event; Battery Problem; No Apparent Adverse Event
IT WAS REPORTED THE IPG HAD REACHED END OF LIFE. IN TURN, THE IPG WAS EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.
Patient problems: Failure of Implant; Inadequate Pain Relief; No Clinical Signs, Symptoms or Conditions
MAUDE report 12476267 ↗InjuryJun 9, 2021Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT THE WRONG IPG (NON MRI) WAS ACCIDENTALLY GIVEN TO THE PHYSICIAN FOR AN IMPLANT. AS A RESULT, SURGICAL INTERVENTION OCCURRED WHEREIN THE IPG WAS EXPLANTED AND REPLACED WITH THE CORRECT (MRI COMPATIBLE) IPG.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11971315 ↗InjurySep 29, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT'S IPG WAS INOPERABLE. AS A RESULT, THE IPG WAS EXPLANTED AND REPLACED. THERAPY WAS RESTORED FOLLOWING THE PROCEDURE.
Patient problems: Inadequate Pain Relief
MAUDE report 10601580 ↗InjuryAug 11, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR WAS SHOWING END OF LIFE. PATIENT LOST STIMULATION AS A RESULT. THE DEVICE WAS EXPLANTED, AND A NEW DEVICE WAS IMPLANTED. THERE WERE NO COMPLICATIONS DURING THE PROCEDURE. PATIENT WAS STABLE.
Patient problems: Inadequate Pain Relief
MAUDE report 10393861 ↗InjuryJul 6, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT POST PROCEDURE, IT WAS OBSERVED THAT THE INCORRECT DEVICE MODEL WAS IMPLANTED IN THE PATIENT. THE DEVICE WAS EXPLANTED, AND A NEW DEVICE WAS IMPLANTED. THERE WERE NO COMPLICATIONS DURING THE PROCEDURE. PATIENT WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10237343 ↗InjuryJul 1, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT POST PROCEDURE, UPON PROGRAMMING OF THE DEVICE IT WAS OBSERVED THAT THE INCORRECT DEVICE MODEL WAS IMPLANTED IN THE PATIENT. THE DEVICE WAS EXPLANTED, AND A NEW DEVICE WAS IMPLANTED. THERE WERE NO COMPLICATIONS DURING THE PROCEDURE. PATIENT WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10217527 ↗MalfunctionOct 7, 2019Premature Elective Replacement Indicator; Premature Elective Replacement Indicator
IT WAS REPORTED THAT DIAGNOSTICS INDICATED THE ELECTIVE REPLACEMENT INDICATOR (ERI) TRIGGERED EARLIER THAN INTENDED. THE WIRELESS SOFTWARE UPDATE WAS PERFORMED CLEARING THE ERI MESSAGE. THE DEVICE IS PROVIDING THERAPY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9162409 ↗InjuryAug 8, 2019Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
IT WAS REPORTED THE PATIENT UNDERWENT A SURGICAL PROCEDURE FOR A WOUND CLEANING IN HER FOOT. REPORTEDLY, AFTER THE PROCEDURE THE SCS SYSTEM PATIENT CONTROLLER (PC) NO LONGER COMMUNICATED WITH THE SCS IPG. THE SYSTEM HAD ONLY BEEN SWITCHED OFF, BUT NOT SET TO SURGERY MODE. A COMPANY REPRESENTATIVE ATTEMPTED TO CONNECT WITH CLINICAL PROGRAMMER BUT A MESSAGE: "UNABLE TO CONNECT TO THE GENERATOR. A PROBLEM HAS BEEN DETECTED..." WAS DISPLAYED. CONSEQUENTLY, SURGICAL INTERVENTION MAY BE NECESSARY TO REPLACE THE PATIENT'S SCS IPG.
Patient problems: Inadequate Pain Relief
MAUDE report 8876204 ↗InjuryJul 26, 2019Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
IT WAS REPORTED THAT THE PATIENT COULD NO LONGER CONNECT WITH THEIR IPG AFTER AN UNRELATED PROCEDURE. AS A RESULT, SURGICAL INTERVENTION WAS TAKEN TO REPLACE THE IPG.
Patient problems: Inadequate Pain Relief
MAUDE report 8832808 ↗InjuryMar 20, 2019Premature Elective Replacement Indicator; Premature Elective Replacement Indicator
IT WAS REPORTED THAT DIAGNOSTICS INDICATED THE ELECTIVE REPLACEMENT INDICATOR (ERI) TRIGGERED EARLIER THAN INTENDED. THE DEVICE HAD THE APPROPRIATE LEVEL OF BATTERY VOLTAGE TO PROVIDE THERAPY. THE DEVICE WAS EXPLANTED AND REPLACED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8436411 ↗InjuryFeb 28, 2019Premature Elective Replacement Indicator; Premature Elective Replacement Indicator
IT WAS REPORTED THE PATIENT'S PHYSICIAN REPLACED THE SCS IPG. REPORTEDLY, THE SCS PROGRAMMER DID NOT SHOW THE CORRECT BATTERY CAPACITY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8380592 ↗InjuryJan 24, 2019Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED COMMUNICATION ISSUES WITH THEIR IPG. AS A RESULT, THE IPG WAS REPLACED. ISSUE HAS BEEN RESOLVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8276652 ↗InjuryOct 3, 2018Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
FOLLOW UP INFORMATION IDENTIFIED POSTOPERATIVELY, EFFECTIVE THERAPY WAS RESTORED.
Patient problems: Inadequate Pain Relief
MAUDE report 7930405 ↗InjurySep 28, 2018Premature End-of-Life Indicator; Premature Elective Replacement Indicator; Premature End-of-Life Indicator; Premature Elective Replacement Indicator
IT WAS REPORTED (BELGIUM) THE DEVICE DISPLAYED THE END OF SERVICE (EOS) MESSAGE. THE PATIENT IS NOT RECEIVING THERAPY AND WILL LIKELY REQUIRE SURGICAL INTERVENTION. AS A RESULT, THE PHYSICIAN EXPLANTED AND REPLACED THE PATIENT¿S IPG WITH A NEW MODEL, RESOLVING THE ISSUE.
Patient problems: Inadequate Pain Relief; No Consequences Or Impact To Patient
MAUDE report 7918822 ↗MalfunctionSep 25, 2018Premature Elective Replacement Indicator; Premature Elective Replacement Indicator
IT WAS REPORTED (B)(6) THE ¿REPLACE GENERATOR SOON¿ MESSAGE WAS DISPLAYED ON THE PATIENT CONTROLLER INDICATING THE GENERATOR WAS APPROACHING ITS END OF SERVICE. THE IPG LOG WAS ASSESSED AND THE IPG WAS NOT AT A TRUE ERI, THE ELECTIVE REPLACEMENT INDICATOR MESSAGE DISPLAYED WAS PREMATURE. THE DEVICE HAS THE APPROPRIATE LEVEL OF BATTERY VOLTAGE TO PROVIDE THERAPY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7908121 ↗InjuryJul 4, 2018Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED ((B)(6)) THE PATIENT¿S IPG WOULD NOT COMMUNICATE WITH THE PATIENT PROGRAMMER OR CLINICIAN PROGRAMMER. MULTIPLE REPLACEMENT PROGRAMMERS WERE TRIED TO NO AVAIL. THE PHYSICIAN DECIDED TO REPLACE THE IPG WITH A NEW ONE IN (B)(6) 2018 (EXACT DATE UNKNOWN).
Patient problems: Inadequate Pain Relief
MAUDE report 7662341 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P010032. One submission often covers many models, so check the product and lot details in each record.
- Z-0458-2026 · Oct 6, 2025 · Open, ClassifiedEterna SCS IPG (Implantable Pulse Generator), Model Number 32400
- Z-2510-2023 · Jul 18, 2023 · Open, ClassifiedInfinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
- Z-2511-2023 · Jul 18, 2023 · Open, ClassifiedInfinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
- Z-2506-2023 · Jul 18, 2023 · Open, ClassifiedProclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
- Z-2509-2023 · Jul 18, 2023 · Open, ClassifiedProclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
- Z-2505-2023 · Jul 18, 2023 · Open, ClassifiedProclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
- Z-2507-2023 · Jul 18, 2023 · Open, ClassifiedProclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
- Z-2508-2023 · Jul 18, 2023 · Open, ClassifiedProclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
- Z-2691-2017 · Jun 2, 2017 · TerminatedProclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
- Z-2692-2017 · Jun 2, 2017 · TerminatedProclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
- Z-1785-2013 · Nov 30, 2012 · TerminatedEon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
- Z-0025-2013 · Sep 4, 2012 · TerminatedThe Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
- Z-1738-2013 · Jul 26, 2012 · TerminatedEon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
- Z-3091-2011 · Jun 2, 2009 · TerminatedSoftware version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im
Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page