Skip to content

GallantRecalls & reported problems across all models

St. Jude Medical, Inc. · 8 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

5 open recalls name at least one Gallant model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
12,885
Deaths reported
115
Injuries reported
2,265
Recalls
5

We found 8 models of Gallant by St. Jude Medical, Inc. in the FDA's device registry (GUDID). Together they are named in 12,885 adverse event reports in the FDA's MAUDE database, received between Aug 21, 2020 and Sep 30, 2026 — classified as 115 deaths, 2,265 injuries, 10,501 malfunctions.

The most frequently reported specific device problem is “Over-Sensing” (6,969 reports).

5 FDA recall records name at least one of these models by its device identifier, 5 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 246 reportsDec 2023: 200 reportsJan 2024: 237 reportsFeb 2024: 236 reportsMar 2024: 228 reportsApr 2024: 285 reportsMay 2024: 258 reportsJun 2024: 240 reportsJul 2024: 195 reportsAug 2024: 229 reportsSep 2024: 227 reportsOct 2024: 281 reportsNov 2024: 276 reportsDec 2024: 167 reportsJan 2025: 232 reportsFeb 2025: 234 reportsMar 2025: 242 reportsApr 2025: 184 reportsMay 2025: 278 reportsJun 2025: 229 reportsJul 2025: 227 reportsAug 2025: 229 reportsSep 2025: 169 reportsOct 2025: 166 reportsNov 2025: 225 reportsDec 2025: 247 reportsJan 2026: 240 reportsFeb 2026: 246 reportsMar 2026: 272 reportsApr 2026: 257 reportsMay 2026: 202 reportsJun 2026: 159 reportsJul 2026: 54 reportsAug 2026: 197 reportsSep 2026: 279 reportsOct 2026: 0 reports202420252026285
Last 36 months. The most recent month may be incomplete.

Gallant recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Aug 16, 2023Open, Classified

    Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication. In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity.

    Z-2615-2023 · Root cause: Component design/selection · Recall details · FDA record ↗

  2. Aug 16, 2023Open, Classified

    Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication. In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity.

    Z-2614-2023 · Root cause: Component design/selection · names 2 of these models · Recall details · FDA record ↗

  3. Mar 10, 2022Open, Classified

    Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.

    Z-0990-2022 · Root cause: Component design/selection · Recall details · FDA record ↗

  4. Mar 10, 2022Open, Classified

    Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.

    Z-0989-2022 · Root cause: Component design/selection · Recall details · FDA record ↗

  5. Mar 10, 2022Open, Classified

    Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.

    Z-0988-2022 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported Gallant problems

Device problems

  • Over-Sensing6,969
  • Failure to Interrogate1,774
  • Adverse Event Without Identified Device or Use Problem1,111
  • Inappropriate/Inadequate Shock/Stimulation778
  • Inappropriate or Unexpected Reset503
  • Incorrect Interpretation of Signal497
  • Under-Sensing436
  • Pacing Problem405

Patient problems

  • No Clinical Signs, Symptoms or Conditions10,994
  • Unspecified Infection921
  • Insufficient Information456
  • Arrhythmia354
  • Discomfort217
  • Pocket Erosion175
  • Wound Dehiscence100
  • Hematoma54

Gallant models

8 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Gallant adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other St. Jude Medical, Inc. products

All St. Jude Medical, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page