Aveir™ LSN25301Recalls & reported problems

St. Jude Medical, Inc. · Introducer Hydrophilic-Coated Introducer Sheath · Class 2

All models of Aveir (10)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05415067034823), received between Dec 22, 2025 and Sep 10, 2026 — classified as 1 death, 1 injury.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Arrhythmia1
  • Cardiac Perforation1
  • Low Blood Pressure/ Hypotension1
  • Vascular Dissection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 10, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that during an Aveir DR implant procedure on a congenital patient, the physician elected to start the procedure with the right atrium leadless pacemaker (RALP). The RALP was released at the target location. When attempting to manipulate the right ventricle leadless pacemaker (RVLP) with the delivery catheter into the RV, the catheter mechanically dislodged the fixed RALP. The RALP embolized in the left femoral artery. The physician inserted an introducer and retrieval catheter into the patient's right femoral artery and snared the RALP. The RALP was pulled toward the introducer but met resistance in the right femoral artery. Under fluoroscopy, contrast arteriograms showed a laceration in the right femoral artery between the introducer and the RALP. Vascular surgery was initiated to retrieve the RALP and repair the vessel. The patient was stable throughout the surgery and is recovering.

    Patient problems: Vascular Dissection

    MAUDE report 50722109 ↗
  • DeathDec 22, 2025Improper or Incorrect Procedure or Method; Adverse Event Without Identified Device or Use Problem; Improper or Incorrect Procedure or Method; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING THE IMPLANT OF A RIGHT VENTRICLE LEADLESS PACEMAKER (RVLP) THAT THE PHYSICIAN WAS UNABLE TO CROSS THE TRICUSPID VALVE DUE TO TORTUOUS ANATOMY. THE PHYSICIAN ELECTED TO ATTEMPT IMPLANT THROUGH THE RIGHT INTERNAL JUGULAR VEIN. THE DEVICE WAS ADVANCED IN THE RIGHT VENTRICLE AND POSITIONED IN THE AREA OF THE RIGHT VENTRICLE SEPTUM, MAPPING WAS PERFORMED AND THE RVLP WAS NOT CAPTURING AT 5V. THE RVLP WAS REPOSITIONED TO A DIFFERENT AREA, AND AROUND THIS TIME THEY COULD NOT MEASURE THE PATIENT'S BLOOD PRESSURE. THE RVLP, CATHETER, AND INTRODUCER WERE REMOVED AND CPR WAS COMMENCED BUT THE PATIENT COULD NOT BE REVIVED. THE PHYSICIAN SUSPECTS CARDIAC PERFORATION.

    Patient problems: Arrhythmia; Cardiac Perforation; Low Blood Pressure/ Hypotension

    MAUDE report 23863953 ↗