NA 405108Recalls & reported problems
St. Jude Medical, Inc. · Peel-Away Introducer · Class 2
- Adverse event reports
- 6
- Deaths reported
- 2
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05415067040879), received between Jun 24, 2022 and Apr 1, 2025 — classified as 2 deaths, 4 injuries.
The most frequently reported specific device problem is “Difficult to Insert” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Difficult to Insert2
- Activation, Positioning or Separation Problem1
- Gas/Air Leak1
Patient problems
- Insufficient Information2
- Air Embolism1
- Cardiac Perforation1
- No Clinical Signs, Symptoms or Conditions1
- Pneumothorax1
- Vascular Dissection1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
DeathApr 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED FOR AN IMPLANT PROCEDURE. THE DAY AFTER THE PROCEDURE, THE PATIENT PASSED AWAY. THERE IS NO KNOWN ALLEGATION FROM A HEALTH CARE PROFESSIONAL THAT SUGGESTS THAT THE DEATH WAS DEVICE RELATED. THE CAUSE OF DEATH WAS UNKNOWN. NO ADDITIONAL INFORMATION WAS REPORTED.
Patient problems: Insufficient Information
MAUDE report 21748600 ↗DeathApr 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED FOR AN IMPLANT PROCEDURE. THE DAY AFTER THE PROCEDURE, THE PATIENT PASSED AWAY. THERE IS NO KNOWN ALLEGATION FROM A HEALTH CARE PROFESSIONAL THAT SUGGESTS THAT THE DEATH WAS DEVICE RELATED. THE CAUSE OF DEATH WAS UNKNOWN. NO ADDITIONAL INFORMATION WAS REPORTED.
Patient problems: Insufficient Information
MAUDE report 21748599 ↗InjuryFeb 29, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING AN IMPLANT PROCEDURE, A PNEUMOTHORAX OCCURRED WHICH REQUIRED RESUSCITATION. DURING IMPLANTATION, THE PATIENT¿S VITAL SIGNS AND OXYGEN SATURATION BECAME POOR. THE ANESTHESIOLOGIST ASKED THAT THE OPERATION BE INTERRUPTED TO STABILIZE THE PATIENT. WHEN THE CS WAS VISUALIZED, USING THE BALLOON CATHETER, CONTRAST MEDIUM ENTERED THE ENTIRE HEART. AN ULTRASOUND WAS USED TO SEE WHERE THE PROBLEM WAS. THE PATIENTS BLOOD PRESSURE DROPPED AND A PNEUMOTHORAX WAS NOTED. AN ATTEMPT WAS MADE TO CORRECT THE PNEUMOTHORAX WITH HELP OF A DRAINAGE. THE PATIENT REQUIRED RESUSCITATION, THE PHYSICIANS BEGAN CHEST COMPRESSIONS, AND THE PATIENT WAS TRANSFERRED TO INTENSIVE CARE UNIT UNDER THE ACCOMPANIMENT OF THE INTERNAL RESCUE TEAM AND THE PROCEDURE WAS ABORTED. IT WAS NOTED THAT THE PNEUMOTHORAX LIKELY OCCURRED DURING PUNCTURE OF THE SUBCLAVIAN VEIN.
Patient problems: Pneumothorax
MAUDE report 18807372 ↗InjuryAug 27, 2023Difficult to Insert; Difficult to Insert
DURING A PACEMAKER IMPLANTATION PROCEDURE, IT WAS NOTED THAT THE GUIDE WIRE DID NOT PROGRESS INSIDE THE INTRODUCER LUMEN. THE INTRODUCER WAS EXCHANGED AND A NEW PUNCTURE IN THE SUBCLAVIAN VEIN WAS NECESSARY. THE PROCEDURE WAS THEN SUCCESSFULLY COMPLETED WITH NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17629119 ↗InjuryDec 4, 2022Difficult to Insert; Activation, Positioning or Separation Problem; Difficult to Insert; Activation, Positioning or Separation Problem
DURING THE PROCEDURE, THE INTRODUCER NEEDLE WAS NOT MOVING WHICH LED TO A PERFORATION IN THE SUBCLAVIAN VEIN.
Patient problems: Cardiac Perforation; Vascular Dissection
MAUDE report 15912790 ↗InjuryJun 24, 2022Gas/Air Leak; Gas/Air Leak
DURING THE IMPLANTATION OF A SINGLE-CHAMBER PACEMAKER WITH AN ELECTRODE IN THE ATRIAL LOCATION, AN AIR EMBOLISM WAS NOTED. WHEN THE TENDRIL ELECTRODE WAS INSIDE THE INTRODUCER, THE ENTRY OF AIR THROUGH THE SHEATH WAS OBSERVED. FOLLOWED BY AN IMMEDIATE DECOMPENSATION OF THE PATIENT WITH OXYGEN SATURATION BELOW 80% AND A BLOOD PRESSURE OF 90/50. VASOACTIVE DRUGS WERE ADMINISTERED WHICH STABILIZED THE PATIENT.
Patient problems: Air Embolism
MAUDE report 14804305 ↗
Other Introducer, catheter devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.