Proclaim™ 3670Recalls & reported problems

St. Jude Medical, Inc. · Plus 5 Implantable Pulse Generator · Class 3 · Implant

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This St. Jude Medical Proclaim Plus spinal cord stimulator has 1,243 adverse event reports, including 2 deaths and 1,166 injuries. Many reports describe inadequate pain relief, pain at the implant site, and wireless communication problems with the device. There are 2 open recalls: one about the battery warning period being shorter than the labeling states, and one about patients being unable to exit MRI mode.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,253
Deaths reported
2
Injuries reported
1,177
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1,253 adverse event reports that list this exact device (DI 05415067046383), received between Dec 22, 2022 and Sep 30, 2026 — classified as 2 deaths, 1,177 injuries, 74 malfunctions.

The most frequently reported specific device problem is “Wireless Communication Problem” (326 reports).

2 FDA recall records name this device's identifier, 2 still open.

13 other recalls cover devices cleared under the same FDA submission (P010032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 14 reportsDec 2023: 13 reportsJan 2024: 20 reportsFeb 2024: 23 reportsMar 2024: 21 reportsApr 2024: 27 reportsMay 2024: 24 reportsJun 2024: 14 reportsJul 2024: 25 reportsAug 2024: 26 reportsSep 2024: 27 reportsOct 2024: 31 reportsNov 2024: 37 reportsDec 2024: 22 reportsJan 2025: 26 reportsFeb 2025: 33 reportsMar 2025: 45 reportsApr 2025: 28 reportsMay 2025: 42 reportsJun 2025: 60 reportsJul 2025: 53 reportsAug 2025: 43 reportsSep 2025: 34 reportsOct 2025: 44 reportsNov 2025: 32 reportsDec 2025: 58 reportsJan 2026: 54 reportsFeb 2026: 50 reportsMar 2026: 62 reportsApr 2026: 42 reportsMay 2026: 27 reportsJun 2026: 17 reportsJul 2026: 15 reportsAug 2026: 21 reportsSep 2026: 44 reportsOct 2026: 0 reports20242025202662
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 16, 2024Open, Classified

    The duration between the implantable pulse generator reaching the elective replacement indicator threshold and end of service may be shorter than indicated in product labeling.

    Z-2364-2024 · Root cause: Vendor change control · FDA record ↗

  2. Jul 18, 2023Open, Classified

    Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).

    Z-2507-2023 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem546
  • Wireless Communication Problem326
  • Battery Problem148
  • Insufficient Device Problem Information78
  • Migration72
  • Premature Discharge of Battery70
  • No Apparent Adverse Event32
  • Inappropriate/Inadequate Shock/Stimulation10

Patient problems

  • Inadequate Pain Relief512
  • Implant Pain371
  • Failure of Implant322
  • Post Operative Wound Infection118
  • No Clinical Signs, Symptoms or Conditions98
  • Unspecified Infection55
  • Insufficient Information33
  • Wound Dehiscence27

Recent adverse event reports

The 25 most recent of 1,253 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that following weight loss, the patient experienced discomfort at the IPG site. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the IPG was replaced, and the pocket site was modified to address the issue.

    Patient problems: Implant Pain

    MAUDE report 50946088 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced pain at the IPG site. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the IPG was repositioned to address the issue.

    Patient problems: Implant Pain

    MAUDE report 50936725 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that patient experienced an infection at the IPG site. As a result, surgical intervention was undertaken wherein the IPG was repositioned to address the issue. The patient was administered oral antibiotics to address the issue.

    Patient problems: Post Operative Wound Infection

    MAUDE report 50936474 ↗
  • InjurySep 29, 2026Premature Discharge of Battery

    It was reported that the patient¿s IPG had reached end of life. It is unknown if this occurred prematurely. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein IPG was explanted and replaced to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50933621 ↗
  • InjurySep 29, 2026Wireless Communication Problem

    Patient problems: Inadequate Pain Relief

    MAUDE report 50932640 ↗
  • InjurySep 29, 2026Wireless Communication Problem

    It was reported the patient¿s IPG was unable to communicate with external devices. A manufacturer representative confirmed the issue and the IPG deemed inoperable. Surgical intervention may be taken at a later date to address the issue.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932033 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50924576 ↗
  • InjurySep 28, 2026Wireless Communication Problem

    Patient problems: Inadequate Pain Relief

    MAUDE report 50923999 ↗
  • InjurySep 28, 2026Wireless Communication Problem

    Patient problems: Inadequate Pain Relief

    MAUDE report 50923979 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient was experiencing pain at the IPG site following weight loss. In turn, surgical intervention was undertaken on (b)(6) 2026 wherein the IPG was moved up to address the issue.

    Patient problems: Implant Pain

    MAUDE report 50918919 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced reduced wound healing and erosion at the IPG site. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the system was explanted to address the issue.

    Patient problems: Erosion

    MAUDE report 50917922 ↗
  • InjurySep 28, 2026Wireless Communication Problem

    It was reported that following an unrelated surgical procedure wherein the device was not placed into surgery mode, the IPG became unable to communicate with external devices. As a result, surgical intervention was undertaken wherein the IPG was explanted and replaced to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50913701 ↗
  • MalfunctionSep 27, 2026Temperature Problem

    Patient problems: Burning Sensation

    MAUDE report 50910472 ↗
  • InjurySep 25, 2026Insufficient Device Problem Information

    Patient problems: Inadequate Pain Relief

    MAUDE report 50903658 ↗
  • InjurySep 25, 2026Wireless Communication Problem

    It was reported that following an unrelated surgical procedure wherein the device was not placed into surgery mode, the IPG became unable to communicate with external devices. Troubleshooting has been unable to resolve the issue. As a result, surgical intervention may be undertaken to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50899863 ↗
  • InjurySep 25, 2026Premature Discharge of Battery

    Patient problems: Inadequate Pain Relief

    MAUDE report 50899853 ↗
  • InjurySep 25, 2026Wireless Communication Problem

    It was reported the patient's IPG was inoperable. Surgical intervention may occur later to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50898835 ↗
  • InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem

    It was reported the patient experienced inadequate stimulation with their SCS system, pain at the IPG site and shocks when turning up the stimulation. As a result, surgical intervention was undertaken wherein the entire system was explanted.NOTE: It is unknown which lead is related to this issue.

    Patient problems: Implant Pain

    MAUDE report 50895948 ↗
  • InjurySep 25, 2026Wireless Communication Problem

    It was reported that following an unrelated surgical procedure wherein the device was not placed into surgery mode, the IPG became unable to communicate with external devices. Troubleshooting was unable to resolve the issue. As a result, surgical intervention was undertaken on (b)(6) 2026 where in the IPG was explanted and replaced to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50893068 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Implant Pain

    MAUDE report 50892252 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Post Operative Wound Infection

    MAUDE report 50892169 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    Patient underwent an implant procedure where in post-op, patient noted pain in their leg. Surgical intervention took place to explant system. Pain resolved post removal.

    Patient problems: Pain

    MAUDE report 50886416 ↗
  • InjurySep 24, 2026Insufficient Device Problem Information

    It was reported that the patient¿s IPG had reached end of life. It is unknown if this occurred prematurely. As a result, surgical intervention may be undertaken to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50886302 ↗
  • InjurySep 23, 2026Wireless Communication Problem

    It was reported that following an unrelated surgical procedure wherein the device was not placed into surgery mode, the IPG became unable to communicate with external devices. Troubleshooting has been unable to resolve the issue. As a result, surgical intervention may be undertaken to address the issue.

    Patient problems: Inadequate Pain Relief

    MAUDE report 50871750 ↗
  • InjurySep 22, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that patient experienced an infection at an unknown site. As a result, surgical intervention was undertaken in (b)(6) 2026 wherein the entire system was explanted to address the issue.

    Patient problems: Post Operative Wound Infection

    MAUDE report 50859982 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P010032. One submission often covers many models, so check the product and lot details in each record.

  • Z-0458-2026 · Oct 6, 2025 · Open, Classified
    Eterna SCS IPG (Implantable Pulse Generator), Model Number 32400
  • Z-2506-2023 · Jul 18, 2023 · Open, Classified
    Proclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
  • Z-2508-2023 · Jul 18, 2023 · Open, Classified
    Proclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
  • Z-2511-2023 · Jul 18, 2023 · Open, Classified
    Infinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2505-2023 · Jul 18, 2023 · Open, Classified
    Proclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
  • Z-2509-2023 · Jul 18, 2023 · Open, Classified
    Proclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
  • Z-2510-2023 · Jul 18, 2023 · Open, Classified
    Infinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
  • Z-2691-2017 · Jun 2, 2017 · Terminated
    Proclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-2692-2017 · Jun 2, 2017 · Terminated
    Proclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
  • Z-1785-2013 · Nov 30, 2012 · Terminated
    Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
  • Z-0025-2013 · Sep 4, 2012 · Terminated
    The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
  • Z-1738-2013 · Jul 26, 2012 · Terminated
    Eon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
  • Z-3091-2011 · Jun 2, 2009 · Terminated
    Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im

Other Stimulator, spinal-cord, totally implanted for pain relief devices

All Stimulator, spinal-cord, totally implanted for pain relief devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.