Proclaim™ 3670Recalls & reported problems
St. Jude Medical, Inc. · Plus 5 Implantable Pulse Generator · Class 3 · Implant
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This St. Jude Medical Proclaim Plus spinal cord stimulator has 1,243 adverse event reports, including 2 deaths and 1,166 injuries. Many reports describe inadequate pain relief, pain at the implant site, and wireless communication problems with the device. There are 2 open recalls: one about the battery warning period being shorter than the labeling states, and one about patients being unable to exit MRI mode.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,253
- Deaths reported
- 2
- Injuries reported
- 1,177
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1,253 adverse event reports that list this exact device (DI 05415067046383), received between Dec 22, 2022 and Sep 30, 2026 — classified as 2 deaths, 1,177 injuries, 74 malfunctions.
The most frequently reported specific device problem is “Wireless Communication Problem” (326 reports).
2 FDA recall records name this device's identifier, 2 still open.
13 other recalls cover devices cleared under the same FDA submission (P010032); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 16, 2024Open, Classified
The duration between the implantable pulse generator reaching the elective replacement indicator threshold and end of service may be shorter than indicated in product labeling.
Z-2364-2024 · Root cause: Vendor change control · FDA record ↗
- Jul 18, 2023Open, Classified
Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).
Z-2507-2023 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem546
- Wireless Communication Problem326
- Battery Problem148
- Insufficient Device Problem Information78
- Migration72
- Premature Discharge of Battery70
- No Apparent Adverse Event32
- Inappropriate/Inadequate Shock/Stimulation10
Patient problems
- Inadequate Pain Relief512
- Implant Pain371
- Failure of Implant322
- Post Operative Wound Infection118
- No Clinical Signs, Symptoms or Conditions98
- Unspecified Infection55
- Insufficient Information33
- Wound Dehiscence27
Recent adverse event reports
The 25 most recent of 1,253 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem
It was reported that following weight loss, the patient experienced discomfort at the IPG site. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the IPG was replaced, and the pocket site was modified to address the issue.
Patient problems: Implant Pain
MAUDE report 50946088 ↗InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient experienced pain at the IPG site. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the IPG was repositioned to address the issue.
Patient problems: Implant Pain
MAUDE report 50936725 ↗InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem
It was reported that patient experienced an infection at the IPG site. As a result, surgical intervention was undertaken wherein the IPG was repositioned to address the issue. The patient was administered oral antibiotics to address the issue.
Patient problems: Post Operative Wound Infection
MAUDE report 50936474 ↗InjurySep 29, 2026Premature Discharge of Battery
It was reported that the patient¿s IPG had reached end of life. It is unknown if this occurred prematurely. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein IPG was explanted and replaced to address the issue.
Patient problems: Inadequate Pain Relief
MAUDE report 50933621 ↗InjurySep 29, 2026Wireless Communication Problem
Patient problems: Inadequate Pain Relief
MAUDE report 50932640 ↗InjurySep 29, 2026Wireless Communication Problem
It was reported the patient¿s IPG was unable to communicate with external devices. A manufacturer representative confirmed the issue and the IPG deemed inoperable. Surgical intervention may be taken at a later date to address the issue.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50932033 ↗InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Unspecified Infection
MAUDE report 50924576 ↗InjurySep 28, 2026Wireless Communication Problem
Patient problems: Inadequate Pain Relief
MAUDE report 50923999 ↗InjurySep 28, 2026Wireless Communication Problem
Patient problems: Inadequate Pain Relief
MAUDE report 50923979 ↗InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient was experiencing pain at the IPG site following weight loss. In turn, surgical intervention was undertaken on (b)(6) 2026 wherein the IPG was moved up to address the issue.
Patient problems: Implant Pain
MAUDE report 50918919 ↗InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient experienced reduced wound healing and erosion at the IPG site. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the system was explanted to address the issue.
Patient problems: Erosion
MAUDE report 50917922 ↗InjurySep 28, 2026Wireless Communication Problem
It was reported that following an unrelated surgical procedure wherein the device was not placed into surgery mode, the IPG became unable to communicate with external devices. As a result, surgical intervention was undertaken wherein the IPG was explanted and replaced to address the issue.
Patient problems: Inadequate Pain Relief
MAUDE report 50913701 ↗MalfunctionSep 27, 2026Temperature Problem
Patient problems: Burning Sensation
MAUDE report 50910472 ↗InjurySep 25, 2026Insufficient Device Problem Information
Patient problems: Inadequate Pain Relief
MAUDE report 50903658 ↗InjurySep 25, 2026Wireless Communication Problem
It was reported that following an unrelated surgical procedure wherein the device was not placed into surgery mode, the IPG became unable to communicate with external devices. Troubleshooting has been unable to resolve the issue. As a result, surgical intervention may be undertaken to address the issue.
Patient problems: Inadequate Pain Relief
MAUDE report 50899863 ↗InjurySep 25, 2026Premature Discharge of Battery
Patient problems: Inadequate Pain Relief
MAUDE report 50899853 ↗InjurySep 25, 2026Wireless Communication Problem
It was reported the patient's IPG was inoperable. Surgical intervention may occur later to address the issue.
Patient problems: Inadequate Pain Relief
MAUDE report 50898835 ↗InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem
It was reported the patient experienced inadequate stimulation with their SCS system, pain at the IPG site and shocks when turning up the stimulation. As a result, surgical intervention was undertaken wherein the entire system was explanted.NOTE: It is unknown which lead is related to this issue.
Patient problems: Implant Pain
MAUDE report 50895948 ↗InjurySep 25, 2026Wireless Communication Problem
It was reported that following an unrelated surgical procedure wherein the device was not placed into surgery mode, the IPG became unable to communicate with external devices. Troubleshooting was unable to resolve the issue. As a result, surgical intervention was undertaken on (b)(6) 2026 where in the IPG was explanted and replaced to address the issue.
Patient problems: Inadequate Pain Relief
MAUDE report 50893068 ↗InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Implant Pain
MAUDE report 50892252 ↗InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Post Operative Wound Infection
MAUDE report 50892169 ↗InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem
Patient underwent an implant procedure where in post-op, patient noted pain in their leg. Surgical intervention took place to explant system. Pain resolved post removal.
Patient problems: Pain
MAUDE report 50886416 ↗InjurySep 24, 2026Insufficient Device Problem Information
It was reported that the patient¿s IPG had reached end of life. It is unknown if this occurred prematurely. As a result, surgical intervention may be undertaken to address the issue.
Patient problems: Inadequate Pain Relief
MAUDE report 50886302 ↗InjurySep 23, 2026Wireless Communication Problem
It was reported that following an unrelated surgical procedure wherein the device was not placed into surgery mode, the IPG became unable to communicate with external devices. Troubleshooting has been unable to resolve the issue. As a result, surgical intervention may be undertaken to address the issue.
Patient problems: Inadequate Pain Relief
MAUDE report 50871750 ↗InjurySep 22, 2026Adverse Event Without Identified Device or Use Problem
It was reported that patient experienced an infection at an unknown site. As a result, surgical intervention was undertaken in (b)(6) 2026 wherein the entire system was explanted to address the issue.
Patient problems: Post Operative Wound Infection
MAUDE report 50859982 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P010032. One submission often covers many models, so check the product and lot details in each record.
- Z-0458-2026 · Oct 6, 2025 · Open, ClassifiedEterna SCS IPG (Implantable Pulse Generator), Model Number 32400
- Z-2506-2023 · Jul 18, 2023 · Open, ClassifiedProclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device
- Z-2508-2023 · Jul 18, 2023 · Open, ClassifiedProclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext
- Z-2511-2023 · Jul 18, 2023 · Open, ClassifiedInfinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
- Z-2505-2023 · Jul 18, 2023 · Open, ClassifiedProclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr
- Z-2509-2023 · Jul 18, 2023 · Open, ClassifiedProclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens
- Z-2510-2023 · Jul 18, 2023 · Open, ClassifiedInfinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio
- Z-2691-2017 · Jun 2, 2017 · TerminatedProclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
- Z-2692-2017 · Jun 2, 2017 · TerminatedProclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The Proclaim implantable pulse generator (IPG) is an electronic device designed to be c
- Z-1785-2013 · Nov 30, 2012 · TerminatedEon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 reca
- Z-0025-2013 · Sep 4, 2012 · TerminatedThe Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode
- Z-1738-2013 · Jul 26, 2012 · TerminatedEon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or li
- Z-3091-2011 · Jun 2, 2009 · TerminatedSoftware version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting im
Other Stimulator, spinal-cord, totally implanted for pain relief devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.