Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05415067051844), received between Jun 10, 2025 and Feb 20, 2026 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak4
  • Fluid/Blood Leak2
  • Improper or Incorrect Procedure or Method1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Air Embolism1
  • Arrhythmia1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 20, 2026Gas/Air Leak

    DURING THE PULSED FIELD ABLATION PROCEDURE, AIR ASPIRATED FROM THE VALVE WHEN THE VOLT CATHETER WAS INSERTED. THE SHEATH WAS REPLACED BUT THE SAME ISSUE OCCURRED. THE SHEATH WAS REPLACED AGAIN AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT. THE SHEATHS WERE FLUSHED PROPERLY WITH EACH FLUSH PERFORMED BEFORE INTRODUCING THE VOLT CATHETER AND ONLY BLOOD WAS ASPIRATED, WITH NO BUBBLES. AIR WAS NOTED AFTER THE VOLT CATHETER WAS INSERTED. AS SOON AS THE VOLT WAS INSERTED, LARGE AIR BUBBLES WERE ASPIRATED. THE ISSUE DISAPPEARED WHEN THE VOLT WAS REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24402698 ↗
  • MalfunctionNov 19, 2025Fluid/Blood Leak; Improper or Incorrect Procedure or Method; Gas/Air Leak; Fluid/Blood Leak; Improper or Incorrect Procedure or Method; Gas/Air Leak

    DURING THE ATRIAL FIBRILLATION PROCEDURE USING PULSED FIELD ABLATION, THERE WAS A BLOOD AND AIR LEAK FROM THE SHEATH. THE SHEATH WAS PREPARED AND INSERTED INTO THE PATIENT. AFTER A SUCCESSFUL TRANSSEPTAL, THERE WAS NO AIR IN THE INITIAL ASPIRATION. WHILE MAPPING WITH THE DIAGNOSTIC CATHETER, BLEED BACK AND AIR WERE NOTED DURING ASPIRATION. THE SHEATH WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23594014 ↗
  • MalfunctionOct 30, 2025Fluid/Blood Leak; Fluid/Blood Leak

    DURING A PVI PROCEDURE WITH NON-ABBOTT (NON-ESI FARAPULSE BY BOSTON SCIENTIFIC), BLEED BACK FROM THE VALVE OF THE SHEATH WAS OBSERVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23424414 ↗
  • InjuryAug 21, 2025Gas/Air Leak; Gas/Air Leak

    DURING AF ABLATION PROCEDURE, AN AIR LEAK OCCURRED WHICH LED TO AN AIR EMBOLISM AND TRANSIENT ST ELEVATION. THERE WERE NO ISSUES WITH THE DEVICE DURING PREP; HOWEVER, AIR WAS NOTED TO BE LEAKING FROM THE HEMOSTATIC VALVE WHEN PULLING BACK AFTER TRANSSEPTAL. THE PHYSICIAN CONTINUED TO USE THE DEVICE. LATER IN THE PROCEDURE WHILE EXCHANGING THE PFA CATHETER TO THE MAPPING CATHETER, THE PATIENT EXPERIENCED TRANSIENT ST ELEVATION WHICH THE PHYSICIAN BELIEVED TO BE LINKED TO THE DEFECTIVE SHEATH. THE SHEATH WAS ASPIRATED TWICE WITH A 20CC SYRINGE. PATIENT WAS DISCHARGED.

    Patient problems: Air Embolism; Arrhythmia

    MAUDE report 22860634 ↗
  • MalfunctionJun 10, 2025Gas/Air Leak; Gas/Air Leak

    DURING AN ATRIAL FIBRILLATION ABLATION PROCEDURE, THE VALVE FROM THE 13F AGILIS WAS LEAKING FOLLOWING TRANSEPTAL. THE SHEATH WAS EXCHANGED AND THE PROCEDURE WAS COMPLETED WITHOUT ANY ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22175842 ↗