Skip to content

Titan ES8922Recalls & reported problems

Coloplast A/S · Scrotal Zero Degree Angle Cylinder Set with Pump · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Coloplast Titan penile implant, a fluid-filled pump and cylinder set, has 339 reports, 284 of them injuries and no deaths. The reports often describe breaks or fractures in parts, fluid leaks, and pumps that stick or fail to inflate, and some describe infections that led to removal and replacement. There is 1 open recall, started in March 2023, about pumps with a thinner wall than standard, which could fail earlier.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
334
Deaths reported
0
Injuries reported
280
Recalls naming it
1

Summary

The FDA's MAUDE database contains 334 adverse event reports that list this exact device (DI 05708932487382), received between Oct 19, 2017 and Jun 23, 2026 — classified as 280 injuries, 54 malfunctions.

The most frequently reported specific device problem is “Fracture” (78 reports).

1 FDA recall record name this device's identifier, 1 still open.

17 other recalls cover devices cleared under the same FDA submission (P000006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 5 reportsJan 2024: 6 reportsFeb 2024: 5 reportsMar 2024: 3 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 3 reportsJul 2024: 2 reportsAug 2024: 5 reportsSep 2024: 4 reportsOct 2024: 6 reportsNov 2024: 9 reportsDec 2024: 6 reportsJan 2025: 3 reportsFeb 2025: 3 reportsMar 2025: 7 reportsApr 2025: 7 reportsMay 2025: 4 reportsJun 2025: 2 reportsJul 2025: 4 reportsAug 2025: 4 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 6 reportsDec 2025: 4 reportsJan 2026: 3 reportsFeb 2026: 8 reportsMar 2026: 2 reportsApr 2026: 9 reportsMay 2026: 11 reportsJun 2026: 4 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 1, 2023Open, Classified

    A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

    Z-1375-2023 · Root cause: Process change control · FDA record ↗

Most reported problems

Device problems

  • Fracture78
  • Adverse Event Without Identified Device or Use Problem60
  • Fluid/Blood Leak36
  • Insufficient Device Problem Information27
  • Inflation Problem22
  • Break21
  • Inadequacy of Device Shape and/or Size19
  • Pumping Problem16

Patient problems

  • No Clinical Signs, Symptoms or Conditions193
  • Unspecified Infection47
  • No Information25
  • Erosion17
  • No Consequences Or Impact To Patient17
  • No Known Impact Or Consequence To Patient11
  • Failure of Implant7
  • Bacterial Infection6

Recent adverse event reports

The 25 most recent of 334 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 23, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE TITAN IMPLANT WAS REMOVED AND REPLACED WITH A MALLEABLE (GENESIS) DUE TO AN INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 25597652 ↗
  • InjuryJun 17, 2026Fluid/Blood Leak; Inflation Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE PATIENT UNDERWENT A REVISION PROCEDURE ABOUT 11 MONTHS AFTER HIS ORIGINAL IMPLANT. HE STATED THE IMPLANT WAS NO LONGER INFLATING AND BELIEVES IT LOST ALL SALINE FLUID. THERE WAS NO KNOWN INFECTION PRESENT OR FURTHER COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25534478 ↗
  • InjuryJun 11, 2026Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS EXPLANTED AND REPLACED DUE TO SIZING, WANTED LARGER CYLINDER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25477918 ↗
  • InjuryJun 9, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED WITH A MALLEABLE (GENESIS) DUE TO INFECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25448588 ↗
  • InjuryMay 29, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT WAS REMOVED AND REPLACED DUE TO POSSIBLE TUBING FRACTURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25331850 ↗
  • InjuryMay 29, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A TUBING BREAK RIGHT ABOVE THE PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25331319 ↗
  • InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED. DURING THE REVISION A PREVIOUS PERFORATION ON LEFT SIDE WAS IDENTIFIED.

    Patient problems: Perforation

    MAUDE report 25317082 ↗
  • InjuryMay 28, 2026Deflation Problem; Inflation Problem

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT WAS REMOVED AND REPLACED DUE TO HARD TO INFLATE AND DEFLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25316728 ↗
  • InjuryMay 26, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A VISIBLE CRACK IN THE TUBING FROM CYLINDER TO PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25282204 ↗
  • InjuryMay 12, 2026Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, THE LEFT CYLINDER WAS REMOVED AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25152715 ↗
  • MalfunctionMay 12, 2026Difficult to Insert; Air/Gas in Device

    ACCORDING TO THE AVAILABLE INFORMATION, DURING THE CASE THEY PREPPED A 22CM TITAN SCROTAL. WHEN THEY PREPPED THE IMPLANT, EVERYTHING WAS FINE. THEY WERE HAVING A DIFFICULT TIME PLACING THE PROXIMAL PORTION OF THE IMPLANT AS WELL AS THE PUMP DURING THE CASE. THEY WERE FINALLY ABLE TO MOVE FORWARD, AS THE SURGEON PREPARED TO PLACE THE PUMP HE REALIZED THERE WAS AN AIR BUBBLE IN THE PUMP. THEY CYCLED THE IMPLANT AGAIN AND THE AIR BUBBLE WAS CLEARED. AS THEY PREPARED TO PLACE THE PUMP AGAIN, HE NOTICED THE PUMP FELT DECOMPRESSED. THEY INITIALLY THOUGHT IT WAS DUE TO A BAD RUBBER SHOD SO THEY SWITCHED TO A NEW RUBBER SHOD AFTER CLEARING THE BUBBLE AGAIN. WHEN THE PHYSICIAN PLACED THE PUMP IN THE SCROTUM, HE NOTICED THE PUMP WAS DECOMPRESSED AGAIN. THEY REMOVED THE ENTIRE IMPLANT FROM THE CORPORA AND PERFORMED A TEST TO ENSURE THERE WAS NO LEAK IN THE SYSTEM. THERE WASN'T ANYTHING VISIBLE BUT AT THAT POINT THE PHYSICIAN DID NOT FEEL COMFORTABLE SINCE THEY COULD NOT UNDERSTAND WHY THAT OCCURRED. THE PHYSICIAN MENTIONED THAT HE WAS VERY FORCEFUL WITH THE PUMP WHEN PLACING IT IN THE SCROTUM AND WONDERED IF THAT COULD HAVE CONTRIBUTED TO THE ISSUE. THE PHYSICIAN DECIDED IT WOULD BE BEST TO R

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25148867 ↗
  • MalfunctionMay 11, 2026Fluid/Blood Leak; Material Puncture/Hole

    ACCORDING TO THE AVAILABLE INFORMATION, WHEN THE PHYSICIAN DEPLOYED THE CYLINDER INTO THE CORPORA, THE CYLINDER OUTSIDE THE BODY FILLED UP WITH BLOODY LIQUID. THIS IMPLANT WAS REMOVED FROM THE BODY. IT SHOWED A HOLE DISTALLY ON THE IMPLANT CAUSING LIQUID TO ESCAPE. ANOTHER IMPLANT WAS OPENED AND USED FOR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25143638 ↗
  • InjuryMay 11, 2026Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT REMOVED AND REPLACED DUE TO INCORRECT SIZING: TOO SHORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25135637 ↗
  • InjuryMay 5, 2026Self-Activation or Keying

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS REMOVED AND REPLACED DUE TO AN AUTO INFLATION MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25081491 ↗
  • InjuryMay 4, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT REMOVED AND REPLACED DUE TO A BREAK IN THE CYLINDER TUBING CLOSE TO THE PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25068083 ↗
  • InjuryApr 23, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A BREAK IN TUBING BY THE PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24967219 ↗
  • InjuryApr 22, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, CYLINDERS REMOVED AND REPLACED DUE TO A TUBING BREAK [FRACTURE] ON THE LEFT CYLINDER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24955765 ↗
  • InjuryApr 21, 2026Crack

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS EXPLANTED AND REPLACED DUE TO A CRACK AT THE TOP OF PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24940065 ↗
  • InjuryApr 20, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED WITH A MALLEABLE (GENESIS) DUE TO AN INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 24932532 ↗
  • InjuryApr 20, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE PUMP WAS EXPLANTED AND REPLACED DUE TO BROKEN TUBING AT THE PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24927562 ↗
  • InjuryApr 17, 2026Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO AN UNKNOWN MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24912874 ↗
  • InjuryApr 16, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS EXPLANTED AND REPLACED DUE TO BROKEN TUBING AT THE PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24903781 ↗
  • InjuryApr 16, 2026Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT REMOVED AND REPLACED DUE TO UNKNOWN REASONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24899762 ↗
  • InjuryApr 14, 2026Migration

    ACCORDING TO THE AVAILABLE INFORMATION THE IMPLANT WAS REMOVED AND REPLACED DUE TO LEFT CYLINDER MIGRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24873539 ↗
  • InjuryMar 9, 2026Fracture; Material Puncture/Hole

    ACCORDING TO THE AVAILABLE INFORMATION THE DEVICE WAS REMOVED AND REPLACED DUE TO HOLE AND LEAK IN TUBING TO CYLINDERS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24544509 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000006. One submission often covers many models, so check the product and lot details in each record.

  • Z-1364-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 18cm, Catalog Number EN29182400; inflatable penile Prosthesis
  • Z-1367-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 18cm, Catalog Number ES28182400; inflatable penile Prosthesis
  • Z-1371-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 22cm, Catalog Number ES29222400; inflatable penile Prosthesis
  • Z-1363-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 16cm, Catalog Number EN29162400; inflatable penile Prosthesis
  • Z-1360-2023 · Mar 1, 2023 · Open, Classified
    Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
  • Z-1373-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 18 Cm, Catalog Number ES89182400; inflatable penile Prosthesis
  • Z-1374-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 20 Cm, Catalog Number ES89202400; inflatable penile Prosthesis
  • Z-1365-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis
  • Z-1369-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 16cm, Catalog Number ES29162400; inflatable penile Prosthesis
  • Z-1370-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 20cm, Catalog Number ES29202400; inflatable penile Prosthesis
  • Z-1362-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 18cm, Catalog Number EN28182400; inflatable penile Prosthesis
  • Z-1359-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Pump, Catalog Numbers 5177502400; inflatable penile Prosthesis
  • Z-1361-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 16cm, Catalog Number EN28162400; inflatable penile Prosthesis
  • Z-1368-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 22cm, Catalog Number ES28222400; inflatable penile Prosthesis
  • Z-1372-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Infrapubic 20 Cm, Catalog Number ES88202400; inflatable penile Prosthesis

…and 2 more.

Other Device, Impotence, Mechanical/Hydraulic devices

All Device, Impotence, Mechanical/Hydraulic devices →

Sources and updates