1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Coloplast Titan Touch inflatable penile implant appears in 865 adverse event reports, with 766 injuries and no deaths. Many describe tubing cracking or breaking, fluid leaks, and unexpected inflation, with some mentioning infections and implants that had to be removed and replaced. One open recall, started in March 2023, concerns pumps with thinner walls that may fail sooner than standard pumps.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
862
Deaths reported
0
Injuries reported
763
Recalls naming it
1

Summary

The FDA's MAUDE database contains 862 adverse event reports that list this exact device (DI 05708932539197), received between Jul 21, 2017 and Jul 8, 2026 — classified as 763 injuries, 99 malfunctions.

The most frequently reported specific device problem is “Fracture” (177 reports).

1 FDA recall record name this device's identifier, 1 still open.

17 other recalls cover devices cleared under the same FDA submission (P000006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 8 reportsDec 2023: 5 reportsJan 2024: 14 reportsFeb 2024: 8 reportsMar 2024: 18 reportsApr 2024: 11 reportsMay 2024: 11 reportsJun 2024: 8 reportsJul 2024: 9 reportsAug 2024: 13 reportsSep 2024: 12 reportsOct 2024: 10 reportsNov 2024: 16 reportsDec 2024: 6 reportsJan 2025: 9 reportsFeb 2025: 10 reportsMar 2025: 10 reportsApr 2025: 7 reportsMay 2025: 9 reportsJun 2025: 5 reportsJul 2025: 15 reportsAug 2025: 8 reportsSep 2025: 6 reportsOct 2025: 7 reportsNov 2025: 6 reportsDec 2025: 11 reportsJan 2026: 11 reportsFeb 2026: 9 reportsMar 2026: 11 reportsApr 2026: 17 reportsMay 2026: 5 reportsJun 2026: 7 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202618
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 1, 2023Open, Classified

    A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

    Z-1370-2023 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture177
  • Adverse Event Without Identified Device or Use Problem123
  • Fluid/Blood Leak98
  • Insufficient Device Problem Information91
  • Inflation Problem68
  • Pumping Problem56
  • Inadequacy of Device Shape and/or Size48
  • Break45

Patient problems

  • No Clinical Signs, Symptoms or Conditions489
  • Unspecified Infection96
  • No Information94
  • Erosion38
  • Failure of Implant23
  • No Consequences Or Impact To Patient23
  • No Known Impact Or Consequence To Patient21
  • Pain21

Recent adverse event reports

The 25 most recent of 862 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 8, 2026Self-Activation or Keying

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT REMOVED AND REPLACED DUE TO AUTO INFLATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25749238 ↗
  • InjuryJul 8, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE INFLATABLE PENILE PROSTHESIS WAS REMOVED AND REPLACED WITH A MALLEABLE (GENESIS) DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 25748822 ↗
  • InjuryJun 23, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO AN INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 25600057 ↗
  • InjuryJun 22, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED WITH A MALLEABLE (GENESIS) A COUPLE OF DAYS LATER DUE TO AN INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 25587348 ↗
  • InjuryJun 22, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A CRACK IN THE TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25587112 ↗
  • InjuryJun 22, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A BREAK IN TUBING AT PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25584558 ↗
  • InjuryJun 19, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A POSSIBLE TUBE FRACTURE FROM PUMP. THE FRACTURE WAS NOT VERIFIED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25563366 ↗
  • InjuryJun 11, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS EXPLANTED AND REPLACED DUE TO A TUBING FRACTURE IN TWO PLACES - ONE ON GRAY TUBING NEAR PUMP, ONE ON CLEAR TUBING FROM CYLINDER TO PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25477983 ↗
  • InjuryJun 4, 2026Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT WAS REMOVED AND REPLACED DUE TO AN UNSPECIFIED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25391684 ↗
  • InjuryMay 11, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT REMOVED AND REPLACED DUE TO A TUBING BREAK AT THE CYLINDER PUMP JUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25135534 ↗
  • InjuryMay 6, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT REMOVED AND REPLACED DUE TO A TUBING FRACTURE FROM PUMP TO CYLINDER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25094782 ↗
  • InjuryMay 5, 2026Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT PUMP REMOVED AND REPLACED FOR AN UNSPECIFIED REASON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25081915 ↗
  • InjuryMay 5, 2026Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO AN UNSPECIFIED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25081339 ↗
  • InjuryMay 4, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE INFLATABLE PENILE PROSTHESIS IMPLANT WAS REMOVED AND REPLACED WITH A MALLEABLE (GENESIS) DUE TO EROSION.

    Patient problems: Erosion

    MAUDE report 25068662 ↗
  • InjuryApr 30, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT REMOVED AND REPLACED DUE TO AN INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 25050613 ↗
  • InjuryApr 28, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A TUBING BREAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25018915 ↗
  • InjuryApr 24, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED WITH A GENESIS DUE TO AN INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 24984424 ↗
  • InjuryApr 22, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT REMOVED AND REPLACED DUE TO A TUBING BREAK BY PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24960210 ↗
  • InjuryApr 21, 2026Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, THE TITAN IMPLANT WAS REMOVED AND REPLACED WITH A GENESIS FOR UNKNOWN REASON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24942582 ↗
  • InjuryApr 20, 2026Device Appears to Trigger Rejection

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO EXTRUSION.

    Patient problems: Erosion

    MAUDE report 24925753 ↗
  • InjuryApr 20, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION THE IMPLANT WAS REMOVED AND REPLACED DUE TO TUBING FRACTURE NEAR PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24925131 ↗
  • InjuryApr 17, 2026Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO THE TUBING LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24912012 ↗
  • InjuryApr 16, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, TITAN IMPLANT REMOVED AND REPLACED WITH A GENESIS DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 24903221 ↗
  • InjuryApr 15, 2026Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT REMOVED AND REPLACED WITH GENESIS DUE TO UNKNOWN REASONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24889403 ↗
  • InjuryApr 15, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION IMPLANT REMOVED AND REPLACED DUE TO A TUBING BREAK IN THE MIDDLE BETWEEN CYLINDER AND PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24885457 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000006. One submission often covers many models, so check the product and lot details in each record.

  • Z-1372-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Infrapubic 20 Cm, Catalog Number ES88202400; inflatable penile Prosthesis
  • Z-1374-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 20 Cm, Catalog Number ES89202400; inflatable penile Prosthesis
  • Z-1375-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 22 Cm, Catalog Number ES89222400; inflatable penile Prosthesis
  • Z-1361-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 16cm, Catalog Number EN28162400; inflatable penile Prosthesis
  • Z-1362-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 18cm, Catalog Number EN28182400; inflatable penile Prosthesis
  • Z-1363-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 16cm, Catalog Number EN29162400; inflatable penile Prosthesis
  • Z-1359-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Pump, Catalog Numbers 5177502400; inflatable penile Prosthesis
  • Z-1371-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 22cm, Catalog Number ES29222400; inflatable penile Prosthesis
  • Z-1366-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Scrotal Zero Ang 18cm, Catalog Number EN88182400; inflatable penile Prosthesis
  • Z-1369-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 16cm, Catalog Number ES29162400; inflatable penile Prosthesis
  • Z-1368-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 22cm, Catalog Number ES28222400; inflatable penile Prosthesis
  • Z-1365-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis
  • Z-1367-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 18cm, Catalog Number ES28182400; inflatable penile Prosthesis
  • Z-1360-2023 · Mar 1, 2023 · Open, Classified
    Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
  • Z-1373-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 18 Cm, Catalog Number ES89182400; inflatable penile Prosthesis

…and 2 more.

Other Device, Impotence, Mechanical/Hydraulic devices

All Device, Impotence, Mechanical/Hydraulic devices →