Skip to content

Titan ES2922Recalls & reported problems

Coloplast A/S · Titan® Touch Scrotal Zero Degree Angle Cylinder Set with Pump · Class 3 · Implant

All models of Titan (62)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Coloplast Titan Touch penile implant has 761 reports, with 1 death and 689 injuries. Many reports describe the implant being removed and replaced after infection, erosion, or breaks in the tubing, and some mention leaks. A recall that started in March 2023 is open, and it concerns pumps with thinner walls than standard, which could fail earlier than expected.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
758
Deaths reported
1
Injuries reported
686
Recalls naming it
1

Summary

The FDA's MAUDE database contains 758 adverse event reports that list this exact device (DI 05708932539203), received between Mar 24, 2016 and Jul 8, 2026 — classified as 1 death, 686 injuries, 71 malfunctions.

The most frequently reported specific device problem is “Fracture” (174 reports).

1 FDA recall record name this device's identifier, 1 still open.

17 other recalls cover devices cleared under the same FDA submission (P000006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 11 reportsDec 2023: 7 reportsJan 2024: 9 reportsFeb 2024: 8 reportsMar 2024: 16 reportsApr 2024: 15 reportsMay 2024: 7 reportsJun 2024: 7 reportsJul 2024: 18 reportsAug 2024: 10 reportsSep 2024: 12 reportsOct 2024: 9 reportsNov 2024: 15 reportsDec 2024: 9 reportsJan 2025: 11 reportsFeb 2025: 6 reportsMar 2025: 8 reportsApr 2025: 10 reportsMay 2025: 9 reportsJun 2025: 12 reportsJul 2025: 9 reportsAug 2025: 7 reportsSep 2025: 7 reportsOct 2025: 8 reportsNov 2025: 6 reportsDec 2025: 10 reportsJan 2026: 4 reportsFeb 2026: 7 reportsMar 2026: 6 reportsApr 2026: 23 reportsMay 2026: 7 reportsJun 2026: 3 reportsJul 2026: 4 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202623
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 1, 2023Open, Classified

    A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

    Z-1371-2023 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture174
  • Adverse Event Without Identified Device or Use Problem94
  • Insufficient Device Problem Information89
  • Fluid/Blood Leak72
  • Inflation Problem67
  • Pumping Problem43
  • Break39
  • Inadequacy of Device Shape and/or Size30

Patient problems

  • No Clinical Signs, Symptoms or Conditions433
  • No Information83
  • Unspecified Infection76
  • Erosion51
  • Pain27
  • Failure of Implant21
  • No Known Impact Or Consequence To Patient14
  • Bacterial Infection13

Recent adverse event reports

The 25 most recent of 758 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 8, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO AN INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 25749153 ↗
  • InjuryJul 8, 2026Material Puncture/Hole

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A CYLINDER PUNCTURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25748580 ↗
  • InjuryJul 8, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO EROSION.

    Patient problems: Erosion

    MAUDE report 25747570 ↗
  • InjuryJul 6, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO TUBING FRACTURE FROM LEFT CYLINDER TO PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25719937 ↗
  • InjuryJun 18, 2026Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED WITH DUE TO A SIZING ISSUE AND THE PATIENT WANTING A NEW DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25551547 ↗
  • InjuryJun 9, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A TUBING FRACTURE FROM PUMP TO TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25447971 ↗
  • InjuryJun 3, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED WITH A GENESIS DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 25379194 ↗
  • InjuryMay 21, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE PATIENT HAD A REAR TIP SLING FROM A PREVIOUS INFLATABLE PENILE PROSTHESIS. THE REAR TIP SLING FAILED, AND THE PATIENT'S CYLINDER ERODED PROXIMALLY. THE PHYSICIAN REMOVED THE RIGHT CYLINDER, INSERTED A TRULOCK PLUG, AND KEPT LEFT CYLINDER.

    Patient problems: Erosion

    MAUDE report 25244137 ↗
  • InjuryMay 21, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, DURING IMPLANT SURGERY THE SURGEON CUT THE RIGHT CYLINDER AND PLUGGED, IMPLANTING ONLY THE LEFT CYLINDER DUE TO A PARTIAL PERFORATION IN THE RIGHT CORPORA.

    Patient problems: Perforation

    MAUDE report 25242121 ↗
  • InjuryMay 19, 2026Fracture

    ACCORDING TO THE AVAILABLE, IMPLANT WAS REMOVED AND REPLACED DUE TO TUBING BREAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25216797 ↗
  • InjuryMay 18, 2026Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO INCORRECT SIZING (TOO LONG).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25210544 ↗
  • InjuryMay 11, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A TUBING BREAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25143398 ↗
  • InjuryMay 1, 2026Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT REMOVED AND REPLACED DUE TO AN UNSPECIFIED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25057369 ↗
  • InjuryMay 1, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT WAS REMOVED AND REPLACED WITH A GENESIS DUE TO PAIN AND APPARENT INFECTION. UPON EXPLANT, NO INFECTION WAS NOTED.

    Patient problems: Pain

    MAUDE report 25055910 ↗
  • InjuryApr 28, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A CRACK [FRACTURE] IN THE TUBING BY THE PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25020517 ↗
  • InjuryApr 24, 2026Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A TUBING LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24985309 ↗
  • InjuryApr 23, 2026Physical Resistance/Sticking

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A STIFF PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24966822 ↗
  • InjuryApr 22, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION, IMPLANT REMOVED AND REPLACED DUE TO THE RIGHT AND LEFT CYLINDER TUBING FRACTURES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24955426 ↗
  • InjuryApr 21, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 24945598 ↗
  • InjuryApr 20, 2026Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO A MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24927742 ↗
  • MalfunctionApr 20, 2026Fluid/Blood Leak

    ACCORDING TO THE AVAILABLE INFORMATION, THERE IS A LEAK IN THE DEVICE, AND ALL THE FLUID HAS LEAKED OUT. A REVISION SURGERY IS PLANNED BUT HAS NOT YET BEEN PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24926639 ↗
  • InjuryApr 20, 2026Fracture

    ACCORDING TO THE AVAILABLE INFORMATION THE IMPLANT WAS REMOVED AND REPLACED DUE TO MALFUNCTION. TEAR IN PUMP TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24925076 ↗
  • InjuryApr 17, 2026Pumping Problem

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANTED PUMP WAS REMOVED AND REPLACED DUE TO A PUMP MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24916490 ↗
  • InjuryApr 17, 2026Inadequacy of Device Shape and/or Size

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO POTENTIAL DISTAL EROSION. REIMPLANTED A NEW SMALLER 20 CM CYLINDER AND CONNECTED TO THE EXISTING RESERVOIR.

    Patient problems: Erosion

    MAUDE report 24915471 ↗
  • InjuryApr 17, 2026Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, THE IMPLANT WAS REMOVED AND REPLACED DUE TO AN UNSPECIFIED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24912636 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000006. One submission often covers many models, so check the product and lot details in each record.

  • Z-1361-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 16cm, Catalog Number EN28162400; inflatable penile Prosthesis
  • Z-1370-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 20cm, Catalog Number ES29202400; inflatable penile Prosthesis
  • Z-1374-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 20 Cm, Catalog Number ES89202400; inflatable penile Prosthesis
  • Z-1366-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Scrotal Zero Ang 18cm, Catalog Number EN88182400; inflatable penile Prosthesis
  • Z-1363-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 16cm, Catalog Number EN29162400; inflatable penile Prosthesis
  • Z-1367-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Infr Zero Ang 18cm, Catalog Number ES28182400; inflatable penile Prosthesis
  • Z-1364-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Scrot Zero 18cm, Catalog Number EN29182400; inflatable penile Prosthesis
  • Z-1373-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 18 Cm, Catalog Number ES89182400; inflatable penile Prosthesis
  • Z-1372-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Infrapubic 20 Cm, Catalog Number ES88202400; inflatable penile Prosthesis
  • Z-1365-2023 · Mar 1, 2023 · Open, Classified
    Titan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis
  • Z-1359-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Pump, Catalog Numbers 5177502400; inflatable penile Prosthesis
  • Z-1369-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch Scro Zero Ang 16cm, Catalog Number ES29162400; inflatable penile Prosthesis
  • Z-1362-2023 · Mar 1, 2023 · Open, Classified
    Titan Touch NB Infr Zero 18cm, Catalog Number EN28182400; inflatable penile Prosthesis
  • Z-1360-2023 · Mar 1, 2023 · Open, Classified
    Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
  • Z-1375-2023 · Mar 1, 2023 · Open, Classified
    Titan 0-Deg Scrotal 22 Cm, Catalog Number ES89222400; inflatable penile Prosthesis

…and 2 more.

Other Device, Impotence, Mechanical/Hydraulic devices

All Device, Impotence, Mechanical/Hydraulic devices →

Sources and updates