Skip to content

WIDEX WPT102Recalls & reported problems

Widex A/S · ACC. MRIC WPT102 W CABLE US/JP · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05714879048043), received between Mar 25, 2024 and Mar 25, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Integrity Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 25, 2024Material Integrity Problem; Material Integrity Problem

    IN THIS EVENT, HEALTH CARE PROFESSIONAL HAD INFORMED THAT A PATIENT'S CHARGER WAS DAMAGED AND USB ENTRANCE WAS BURNT, WHILE CHARGING. INVESTIGATION RESULTS: BURNT MARKS FOUND ON THE CHARGER AND CONNECTOR; THE MICRO-USB WAS LEFT ON THE RECEPTABLE. X-RAY RESULT SHOWS NO ABNORMALITY INSIDE THE CHARGER THAT PROVES BURNING/ MELTING STARTED ON THE CONNECTOR ITSELF. SINCE PART OF THE CONNECTOR IS THOROUGHLY BURNT, CANNOT VALIDATE IF IT WAS INDUCED BY CABLE RELATED DEFECT OR BY USER. THE RESULTS OF THE VISUAL INSPECTION AND TECHNICAL ASSESSMENT INDICATE THAT THE ROOT-CAUSE IS A SHORT CIRCUIT DEVELOPED IN THE USB CONNECTOR OR USB RECEPTACLE DUE TO CORROSION, BENT PINS, OR FOREIGN DEBRIS. CORRECTIVE AND PREVENTATIVE ACTIONS HAVE NOW BEEN IMPLEMENTED AND DOCUMENTED IN A CAPA. PREVENTATIVE ACTION WAS TAKEN TO REDUCE THE PORTABLE CHARGER'S MAXIMUM BATTERY CAPACITY FROM 100% TO 80% TO PREVENT EXTENSIVE BATTERY REACTIONS DURING BATTERY SHORT-CIRCUIT. PREVENTIVE MEASURE WAS VERIFIED TO BE EFFECTIVE IN FURTHER REDUCING THE POSSIBILITY AND SUFFICIENT ENERGY TO CAUSE SEVERE MELTING CASES. NO FURTHER ROOT CAUSES HAVE BEEN IDENTIFIED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18968966 ↗

Other Hearing aid, air conduction with wireless technology devices

All Hearing aid, air conduction with wireless technology devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page