Signia PURE 312 5AXRecalls & reported problems
Ws Audiology Usa, Inc. · Signia Pure 312 5AX hearing aid/ DKC · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05714880154627), received between Jun 28, 2024 and Jun 28, 2024 — classified as 1 injury.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Patient Device Interaction Problem1
Patient problems
- Discomfort1
- Ear Pain1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJun 28, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
IN THIS EVENT, PATIENT WAS REMOVING HEARING AID FROM EAR WHEN UNKOWNING TO THE USER, THE DOME HAD DETACHED FROM THE RECEIVER AND REMAINED IN EAR CANAL. PATIENT CONTINUED TO REINSERT HEARING AID FOR THE FOLLOWING DAYS, PUSHING THE DETACHED DOME DEEPER INTO EAR CANAL. PATIENT WENT TO ENT DOCTOR WHO FOUND THE DETACHED DOME LODGED IN EAR CANAL AND WAS ABLE TO REMOVE FROM EAR. USER REPORTED PAIN AND DISCOMFORT IN THE EAR. PATIENT WAS PRESCRIBED AN ANTIBIOTIC. RESULTS OF TECHNICAL INVESTIGATION: HEARING AIDS WITH ERU WAX GUARD WERE RETURNED FOR INVESTIGATION. DOME THAT WAS LODGED IN PATIENT'S EAR WAS NOT RETURNED. DEVICES THAT WERE RETURNED WERE TESTED AND ARE FUNCTIONING WITHIN SPECIFICATION. USER GUIDE INSTRUCTS PATIENT TO SEEK ASSISTANCE FROM MEDICAL PROFESSIONAL SHOULD A PIECE REMAIN IN THE EAR.
Patient problems: Discomfort; Ear Pain
MAUDE report 19632401 ↗
Other Hearing Aid, Air Conduction With Wireless Technology devices
- GN ReSound 20472100Gn Hearing A/S · Hearing aid, air conduction with wireless technology55 reports
- GN ReSound 21625000Gn Hearing A/S · Hearing aid, air conduction with wireless technology20 reports
- GN Hearing 21655200Gn Hearing A/S · Hearing aid, air conduction with wireless technology8 reports
- Phonak Brand 050-0889-P5Sonova AG · Hearing Aid, Air Conduction With Wireless Technology8 reports
- Beltone 20623200Gn Hearing A/S · Hearing aid, air conduction with wireless technology7 reports
- Beltone 21625700Gn Hearing A/S · Hearing aid, air conduction with wireless technology7 reports
- Phonak Audéo I90-Sphere (graphite gray)Sonova AG · Hearing Aid, Air-Conduction With Wireless Technology, Prescription5 reports
- GN Hearing 20863700Gn Hearing A/S · Hearing aid, air conduction with wireless technology5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page