DropSafe, safety pen needles 31G x 6mmRecalls & reported problems
Htl Strefa S A · 100 pcs sterile single use DropSafe pen needles per box, intended for use with pen injector devices for the injection of drugs. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05907996091584), received between Nov 4, 2019 and May 20, 2026 — classified as 5 injuries, 1 malfunction.
The most frequently reported specific device problem is “Material Twisted/Bent” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Twisted/Bent4
- Fail-Safe Did Not Operate3
- Material Deformation1
- Material Fragmentation1
Patient problems
- Needle Stick/Puncture5
- Unspecified Tissue Injury1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryMay 20, 2026Material Twisted/Bent; Fail-Safe Did Not Operate
ON 04/22/2026 HTL-STREFA INC. RECEIVED AN EMAIL NOTIFICATION. CUSTOMER STATED: THE NEEDLE CAME OUT OF THE SIDE AND STUCK A NURSE. WE TESTED MULTIPLE ONES AND IT HAPPENS WITH THEM ALL. SUPPOSED TO BE A SAFETY NEEDLE. PHOTO HAS BEEN ATTACHED TO THE NOTIFICATION. HTL-STREFA MADE ATTEMPT TO OBTAIN ADDITIONAL INFORMATION TO THE EVENT CIRCUMSTANCES. WITHOUT RESULTS. ON 2026-05-15 SAMPLE UNDER COMPLAINT HAS BEEN RETURNED FOR FURTHER ANALYSIS.
Patient problems: Needle Stick/Puncture
MAUDE report 25228051 ↗InjuryApr 14, 2026Material Twisted/Bent; Fail-Safe Did Not Operate
ON 04/01/2026 HTL-STREFA INC. RECEIVED AN EMAIL COMPLAINT NOTIFICATION. CUSTOMER REPORTED A NEEDLESTICK AT THE FACILITY. AFTER YOU INJECT THE INSULIN, THE NEEDLE IS RETRACTING BACK BUT STICKS OUT THE SIDE AND CAUSED A NEEDLESTICK. PHOTO HAS BEEN ATTACHED TO THE NOTIFICATION. ON THE PROVIDED PHOTO WE CAN SEE BENT NEEDLE FROM THE USER END. THE NEEDLE IS EXPOSED FROM SAFETY PEN NEEDLE. HTL-STREFA MADE ATTEMPT TO OBTAIN ADDITIONAL INFORMATION TO THE EVENT CIRCUMSTANCES.
Patient problems: Needle Stick/Puncture
MAUDE report 24870589 ↗InjuryApr 14, 2026Material Twisted/Bent; Fail-Safe Did Not Operate
ON 03/23/2026 HTL-STREFA INC. RECEIVED AN EMAIL COMPLAINT NOTIFICATION. CUSTOMER STATED: THE NEEDLE CAME OUT OF THE SIDE AND STUCK A NURSE. WE TESTED MULTIPLE ONES AND IT HAPPENS WITH THEM ALL. SUPPOSED TO BE A SAFETY NEEDLE.CUSTOMER WAS CALLED SEVERAL TIMES WITH NO ANSWER AND NO OPTION FOR LEAVING A CALL BACK MESSAGE. EMAIL WAS SENT FOR ADDITIONAL INFORMATION.ON 2026-04-22 SAMPLE UNDER COMPLAINT HAS BEEN RETURNED FOR FURTHER ANALYSIS.
Patient problems: Needle Stick/Puncture
MAUDE report 24870298 ↗MalfunctionMar 21, 2023Material Fragmentation; Material Fragmentation
JW SENT THE INITIAL EMAIL TO HTL INC. SHE WRITES, "RECENTLY, WE HAVE HAD AN ISSUE WITH AN ACCIDENTAL NEEDLE STICK DUE TO A FAULTY LOT OF NEEDLES WE WERE SENT. PLEASE SEE THE ATTACHED PHOTOS OF THE ISSUE AND ADVISE HOW THIS ISSUE CAN BE PREVENTED IN THE FUTURE." JW FORWARDED INFORMATION FROM THE CLINICAL TEAM LEADER, JK, AT THE FACILITY WHERE THE INCIDENT OCCURS THE FACILITY IS OCONTO HOSPITAL CLINIC. THE TEAM LEADER WRITES, "HELLO TEAMS, SITUATION: A NEEDLESTICK OCCURRED RELATED TO A FAULTY INSULIN PEN NEEDLE- DROPSAFE BRAND LOT C68H4 (SEE PICTURES) BACKGROUND: THE NEEDLE INSERTED INTO THE PEN AT AN ANGLE AND CAUSED IT TO GET LODGED IN THE PEN EVEN AFTER THE CAP WAS REMOVED. THE NURSE ADMINISTERING THE NEXT DOSE OF INSULIN WENT TO SCRUB THE TOP OF THE INSULIN PEN AND GOT POKED BY THE FAULTY, RETAINED NEEDLE. ASSESSMENT: BHOH'S SUPPLY CHAIN SWITCHED TO THESE INSULIN PEN NEEDLES A COUPLE OF MONTHS AGO. REVIEW WHAT PRODUCTS YOU HAVE IN YOUR AREA. RECOMMENDATION: ADVISE NURSES TO WATCH FOR RETAINED NEEDLES AS THEY REMOVE THE INSULIN PEN NEEDLES AND BEFORE THEY SCRUB THE CAP TO PUT A NEW ONE ON."
Patient problems: Unspecified Tissue Injury
MAUDE report 16581175 ↗InjuryNov 4, 2019Material Twisted/Bent; Material Twisted/Bent
'I WAS JUST MADE AWARE OF A NEEDLE STICK THAT OCCURRED IN OUR (B)(6) FACILITY USING ONE OF THE DROP SAFE PEN NEEDLES THAT CAME OUT OF THE SIDE OF THE DEVICE. ADDITIONAL DETAILS - HERE IS THE INFO REGARDING THE NEEDLE STICK:. THE NURSE - SHE WAS ATTEMPTING TO REMOVE THE USED NEEDLE AT THE END OF THE INSULIN PEN, THE NEEDLE WAS BENT THROUGH ONE OF THE OPENING ON THE DROP SAFE PEN NEEDLES AND IT CAME OUT OF THE SIDE OF THE DEVICE. SHE QUICKLY REPORTED AND SHOWED THE SUPERVISOR WHAT HAD HAPPENED. WE SENT HER TO (B)(6) LAB FOR BLOOD DRAW AND THE FIRST ROUND OF RESULTS WERE NEGATIVE. SHE WILL FOLLOW THE SERIAL LAB TESTING PROTOCOL UNTIL COMPLETED'.
Patient problems: Needle Stick/Puncture
MAUDE report 9271437 ↗InjuryNov 4, 2019Material Deformation; Material Deformation
'I WAS JUST MADE AWARE OF A NEEDLE STICK THAT OCCURRED IN OUR (B)(6) FACILITY USING ONE OF THE DROP SAFE PEN NEEDLES THAT CAME OUT OF THE SIDE OF THE DEVICE. ADDITIONAL DETAILS - HERE IS THE INFO REGARDING THE NEEDLE STICK:. THE NURSE - SHE WAS ATTEMPTING TO REMOVE THE USED NEEDLE AT THE END OF THE INSULIN PEN, THE NEEDLE WAS BENT THROUGH ONE OF THE OPENING ON THE DROP SAFE PEN NEEDLES AND IT CAME OUT OF THE SIDE OF THE DEVICE. SHE QUICKLY REPORTED AND SHOWED THE SUPERVISOR WHAT HAD HAPPENED. WE SENT HER TO (B)(6) LAB FOR BLOOD DRAW AND THE FIRST ROUND OF RESULTS WERE NEGATIVE. SHE WILL FOLLOW THE SERIAL LAB TESTING PROTOCOL UNTIL COMPLETED'.
Patient problems: Needle Stick/Puncture
MAUDE report 9271407 ↗
Other Needle, Hypodermic, Single Lumen devices
- BD Nano 2nd Gen 320550Becton Dickinson Pen Limited · Needle, hypodermic, single lumen2,044 reports
- BD SafetyGlide 305916Becton, Dickinson And Company · Needle, hypodermic, single lumen1,767 reports
- BD Nano Ultra-Fine 320122Becton Dickinson Pen Limited · Needle, hypodermic, single lumen601 reports
- N/A 305106Becton, Dickinson And Company · Needle, hypodermic, single lumen1 recall467 reports
- N/A 305110Becton, Dickinson And Company · Needle, hypodermic, single lumen418 reports
- BD Safe-Clip 328235Becton, Dickinson and Company Inc · Needle, hypodermic, single lumen268 reports
- BD Ultra-Fine 320109Becton Dickinson Pen Limited · Needle, hypodermic, single lumen229 reports
- BD AutoShield Duo 329515Becton, Dickinson And Company · Needle, hypodermic, single lumen208 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page