Skip to content

Surgify Halo | 3.0 mm | Extendable 30.000.SEE.U2Recalls & reported problems

Surgify Medical Oy · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 06429811532229), received between Jun 23, 2026 and Jun 23, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 30, 2026Class IIOpen, Classified

    Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS).

    Z-2555-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 23, 2026Detachment of Device or Device Component

    DURING A SPINE PROCEDURE THE BURR DISASSEMBLED. THE USE TIME OF THE DEVICE EXCEEDED APPROXIMATELY 6 TIMES THE USE TIME SPECIFIED IN THE IFU. ALSO 50% HIGHER RPM WAS USED COMPARED TO THE RPM STATED IN THE IFU. NO REPORTED HARM WAS CAUSED TO THE PATIENT. POST-OPERATIVE PICTURES (TAKEN FOR OTHER REASONS) CONFIRMED THAT NO BURR FRAGMENTS REMAINED IN THE BODY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25589715 ↗

Other Drills, Burrs, Trephines & Accessories (Simple, Powered) devices

All Drills, Burrs, Trephines & Accessories (Simple, Powered) devices →

Sources and updates