Single-Use Marked Spring Tip Guidewire DG-18-08-21Recalls & reported problems
Micro-Tech (Nanjing) Co., Ltd. · The diameter of the spring tip is 1.8mm, the diameter of the wire guide rod is 0.8mm, and the effective working length is 2100mm. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 06932503525114), received between Jun 20, 2022 and Oct 8, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Fracture1
- Material Deformation1
Patient problems
- Perforation of Esophagus2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 8, 2024Material Deformation; Material Deformation
IT WAS REPORTED THAT MONDAY, ON (B)(6) GI TECH (B)(6) AND DR. (B)(6) HAD AN ESOPHAGEAL PERFORATION WHILE USING MICROTECH'S SPRING TIP GUIDEWIRE. THE ACCOUNT WAS TRAILING MICROTECH'S GUIDEWIRE. (B)(6) SAID WHILE TEACHING THE DR WAS ADVANCING THE SAVORY 1:1 BETWEEN MARKING POINT 3 AND 4 ON THE GUIDEWIRE. THE PHYSICIAN FELT NO RESISTANCE DURING ADVANCEMENT OF SAVORY. THERE WAS A DENT OR KINK SEEN BETWEEN THE 3 AND 4 LANDING MARK ON OUR GUIDEWIRE. A PERFORATION OCCURRED AND THE PATIENT WAS ADMITTED AND IS RECOVERING AS EXPECTED. THE DR PLACED A 25MM BY 155 ESOPHAGEAL STENT. DURING THE TEAM DEBRIEF THE DR & TECH CONCLUDED THE GUIDEWIRE CAUSED TO THE PERFORATION.
Patient problems: Perforation of Esophagus
MAUDE report 20394683 ↗MalfunctionJun 20, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED TO MICRO-TECH(NANJING) CO.,LTD. THAT DR. BATTAGLINO WAS USING A SAVORY WITH GUIDEWIRE TO DILATE AND STATED THAT HE FELT RESISTANCE WITH ADVANCING THE SAVORY, SO REMOVED THE SAVORY WITH THE WIRE AND NOTICED THAT THE SPRING END OF THE WIRE WAS NO LONGER ATTACHED TO THE REST OF THE WIRE THAT HE REMOVED, SO THEN HAD TO GO BACK IN TO RETRIEVE THE BROKEN END. ONCE REENTERING, THEY FOUND THAT THE BROKEN END OF THE TIP THAT WAS RETAINED HAD PERFORATED THROUGH THE ESOPHAGUS AND POSSIBLY INTO OTHER TISSUE. THAT PIECE WAS THEN REMOVED AND THE CRNA OBTAINED AN ADVANCED AIRWAY AS HE THEN HAD A BLEED AND DID HAVE A CHANGE IN O2 SATURATION AND BLOOD PRESSURE. THE PATIENT WAS TRANSPORTED VIA EMS TO WAKE MED, WHERE HE UNDERWENT SURGICAL REPAIR OF THE PERFORATION. HERE ARE THE PHOTOS FROM THE WIRE THAT WAS USED. IT SEEMS TO HAVE BROKEN AT ONE OF THE DISTANCE MARKER SITES. REC FROM CUSTOMER ON 5/27/2022 - THANK YOU FOR LOOKING INTO THIS. AS OF NOW, PER DR. BATTAGLINO, THE PATIENT IS AWAKE AND HAS BEEN EXTUBATED, AND IS DOING MUCH BETTER AT THE MOMENT. HE DID HAVE SURGICAL REPAIR OF AN ESOPHAGEAL PERFORATION AND ALSO SUFFERED A PNEUMOTHORAX, REQUIRING CHEST TUBE PLACEMENT AS THE WIRE H
Patient problems: Perforation of Esophagus
MAUDE report 14741739 ↗
Other Endoscopic Guidewire, Gastroenterology-Urology devices
- HiWire G30474Cook Incorporated · Endoscopic guidewire, gastroenterology-urology82 reports
- Roadrunner G18154Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall18 reports
- Roadrunner G17542Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall9 reports
- Roadrunner G18153Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall6 reports
- Roadrunner G18155Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall3 reports
- Roadrunner G17539Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall2 reports
- Roadrunner G17540Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall0 reports
- Jagwire M00556580Boston Scientific Corporation · Endoscopic guidewire, gastroenterology-urology1 death49 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page