Skip to content

IceRod® 1.5 Cryoablation Needle FPRPR3203Recalls & reported problems

Boston Scientific Corporation · 1.5mm Cryoablation Needle · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07290104830198), received between Nov 7, 2024 and Jul 21, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Thermal Decomposition of Device” (3 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 18, 2020Class IICompleted

    Complaint trend regarding needle shaft gas leaks.

    Z-0669-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Thermal Decomposition of Device3
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Fever1
  • Hemoptysis1
  • Hemothorax1
  • Pain1
  • Pleural Effusion1
  • Pneumothorax1
  • Unspecified Gastrointestinal Problem1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 21, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ADVERSE EVENTS OCCURRED, AND CATHETERIZATION WAS REQUIRED. THIS RETROSPECTIVE ANALYSIS AIMED TO COMPARE THE SAFETY AND EFFICACY OF MICROWAVE ABLATION (MWA) VERSUS CRYOABLATION (CA) IN PATIENTS WITH PRIMARY HIGH-RISK PULMONARY NODULES. ALL LUNG MWA AND CA PROCEDURES WERE PERFORMED BY TWO EXPERIENCED (>5 YEARS) INTERVENTIONAL RADIOLOGISTS UNDER COMPUTED TOMOGRAPHY (CT) GUIDANCE. GENERAL ANESTHESIA WAS ADMINISTERED THROUGH A COMBINATION OF INTRAVENOUS AND INHALATIONAL AGENTS. CONTINUOUS MONITORING OF CARDIAC ACTIVITY AND VITAL PARAMETERS WAS MAINTAINED THROUGHOUT EACH PROCEDURE. UPON COMPLETION, A FOLLOW-UP CT SCAN WAS PERFORMED TO DETECT ANY IMMEDIATE COMPLICATIONS, AND PATIENTS WERE SUBSEQUENTLY TRANSFERRED TO AN INPATIENT UNIT FOR OBSERVATION. CRYOABLATION WAS PERFORMED USING VISUAL ICE? SYSTEM WITH ICEROD1.5 CRYOPROBES UNDER CT GUIDANCE TO ENSURE THE TARGETED NODULE WAS ENCOMPASSED BY AN ADEQUATE MARGIN. THE ABLATION PROTOCOL CONSISTED OF THREE FREEZE-THAW CYCLES. THE PROTOCOL INCLUDED A FIRST FREEZE OF 3 MINUTES, A THAW PERIOD OF 3 MINUTES, A SECOND FREEZE LASTING 8 MINUTES, FOLLOWED BY A 5-MINUTE THAW, AND A FINAL 8-MINUTE FREEZE BEFORE ACTIVE THAWING. THE A

    Patient problems: Fever; Hemoptysis; Hemothorax; Pain; Pleural Effusion; Pneumothorax; Unspecified Gastrointestinal Problem

    MAUDE report 25894295 ↗
  • MalfunctionNov 7, 2024Thermal Decomposition of Device; Thermal Decomposition of Device

    IT WAS REPORTED THAT THERE WAS FROST ALONG THE LENGTH OF THE NEEDLE SHAFT. ICEROD NEEDLES WERE SELECTED FOR A CRYOABLATION PROCEDURE IN THE LIVER. DURING THE PROCEDURE, ABOUT 1-3 MINUTES INTO THE FREEZE, FROST FORMED ON THE SHAFTS OF ALL THE NEEDLES. A GLOVE FILLED WITH HOT WATER WAS USED TO PROTECT THE SKIN FROM FREEZING. THE PROCEDURE WAS CONTINUED WITH THE ORIGINAL NEEDLES. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20632398 ↗
  • MalfunctionNov 7, 2024Thermal Decomposition of Device; Thermal Decomposition of Device

    IT WAS REPORTED THAT THERE WAS FROST ALONG THE LENGTH OF THE NEEDLE SHAFT. ICEROD NEEDLES WERE SELECTED FOR A CRYOABLATION PROCEDURE IN THE LIVER. DURING THE PROCEDURE, ABOUT 1-3 MINUTES INTO THE FREEZE, FROST FORMED ON THE SHAFTS OF ALL THE NEEDLES. A GLOVE FILLED WITH HOT WATER WAS USED TO PROTECT THE SKIN FROM FREEZING. THE PROCEDURE WAS CONTINUED WITH THE ORIGINAL NEEDLES. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20631500 ↗
  • MalfunctionNov 7, 2024Thermal Decomposition of Device; Thermal Decomposition of Device

    IT WAS REPORTED THAT THERE WAS FROST ALONG THE LENGTH OF THE NEEDLE SHAFT. ICEROD NEEDLES WERE SELECTED FOR A CRYOABLATION PROCEDURE IN THE LIVER. DURING THE PROCEDURE, ABOUT 1-3 MINUTES INTO THE FREEZE, FROST FORMED ON THE SHAFTS OF ALL THE NEEDLES. A GLOVE FILLED WITH HOT WATER WAS USED TO PROTECT THE SKIN FROM FREEZING. THE PROCEDURE WAS CONTINUED WITH THE ORIGINAL NEEDLES. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20631491 ↗

Other Unit, Cryosurgical, Accessories devices

All Unit, Cryosurgical, Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page