Skip to content

IceRod® 1.5 MRI 90° Cryoablation Needle FPRPR3195Recalls & reported problems

Boston Scientific Corporation · 1.5mm 90 degree Cryoablation Needle · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07290104830297), received between Mar 16, 2019 and Aug 11, 2022 — classified as 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Failure to Fire” (4 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Nov 18, 2020Class IICompleted

    Complaint trend regarding needle shaft gas leaks.

    Z-0678-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Fire4
  • Gas/Air Leak2
  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • No Known Impact Or Consequence To Patient4
  • Burn, Thermal2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 11, 2022Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT A NEEDLE AIR LEAK OCCURRED. AN ICEROD 90 DEGREE MRI NEEDLE WAS SELECTED FOR THE INDEX CRYOABLATION PROCEDURE. DURING PREPARATION, HOWEVER, THERE WAS AN AIR LEAK OBSERVED. THE PROCEDURE WAS COMPLETED USING A DIFFERENT DEVICE OF THE SAME MODEL. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15213922 ↗
  • MalfunctionAug 10, 2022Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT A NEEDLE AIR LEAK OCCURRED. AN ICEROD 90 DEGREE MRI NEEDLE WAS SELECTED FOR THE INDEX CRYOABLATION PROCEDURE. DURING PREPARATION, HOWEVER, THERE WAS AN AIR LEAK OBSERVED. THE PROCEDURE WAS COMPLETED USING A DIFFERENT DEVICE OF THE SAME MODEL. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15203975 ↗
  • MalfunctionFeb 27, 2020Failure to Fire; Failure to Fire

    4 NEEDLES WERE USED IN A MRI GUIDED RENAL CRYOABLATION PROCEDURE. DURING THE CASE, ALL 4 ICEROD MRI NEEDLES ONLY REACHED AN ICEBALL SIZE OF APPROX. 1 CM, WHICH WAS INSUFFICIENT FOR THE PROCEDURE. EXTRA ACTIVE THAWING CYCLES DID NOT RESOLVE THIS. MULTIPLE FREEZING CYCLES WERE CONDUCTED IN AN ATTEMPT TO REACH AN ACCEPTABLE ABLATION ZONE. THE TREATMENT WAS DEEMED UNSUCCESSFUL, AND A NEW CRYOABLATION PROCEDURE HAS BEEN SCHEDULED FOR THIS PATIENT. ALL FOUR NEEDLES WERE RETURNED TO THE MANUFACTURER FOR INVESTIGATION, AND THE CRYOABLATION SYSTEM WAS INVESTIGATED IN A FIELD SERVICE CALL. THIS MDR SUMMARISES THE INVESTIGATION RESULTS OF THE FOURTH NEEDLE USED IN THIS COMPLAINT. PLEASE SEE THE FOLLOWING MDR'S FOR THE OTHER DEVICES USED DURING THIS CRYOABLATION PROCEDURE: 3004462490-2020-00001 - CRYOABLATION SYSTEM, 9616793-2020-00035 - NEEDLE 1, 9616793-2020-00036 - NEEDLE 2, 9616793-2020-00037 - NEEDLE 3

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9767066 ↗
  • MalfunctionFeb 27, 2020Failure to Fire; Failure to Fire

    4 NEEDLES WERE USED IN A MRI GUIDED RENAL CRYOABLATION PROCEDURE. DURING THE CASE, ALL 4 ICEROD MRI NEEDLES ONLY REACHED AN ICEBALL SIZE OF APPROX. 1 CM, WHICH WAS INSUFFICIENT FOR THE PROCEDURE. EXTRA ACTIVE THAWING CYCLES DID NOT RESOLVE THIS. MULTIPLE FREEZING CYCLES WERE CONDUCTED IN AN ATTEMPT TO REACH AN ACCEPTABLE ABLATION ZONE. THE TREATMENT WAS DEEMED UNSUCCESSFUL, AND A NEW CRYOABLATION PROCEDURE HAS BEEN SCHEDULED FOR THIS PATIENT. ALL FOUR NEEDLES WERE RETURNED TO THE MANUFACTURER FOR INVESTIGATION, AND THE CRYOABLATION SYSTEM WAS INVESTIGATED IN A FIELD SERVICE CALL. THIS MDR SUMMARISES THE INVESTIGATION RESULTS OF THE FIRST NEEDLE USED IN THIS COMPLAINT. PLEASE SEE THE FOLLOWING MDR'S FOR THE OTHER DEVICES USED DURING THIS CRYOABLATION PROCEDURE: 3004462490-2020-00001 - CRYOABLATION SYSTEM, 9616793-2020-00036 - NEEDLE 2, 9616793-2020-00037 - NEEDLE 3, 9616793-2020-00038 - NEEDLE 4.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9767060 ↗
  • MalfunctionFeb 27, 2020Failure to Fire; Failure to Fire

    4 NEEDLES WERE USED IN A MRI GUIDED RENAL CRYOABLATION PROCEDURE. DURING THE CASE, ALL 4 ICEROD MRI NEEDLES ONLY REACHED AN ICEBALL SIZE OF APPROX. 1 CM, WHICH WAS INSUFFICIENT FOR THE PROCEDURE. EXTRA ACTIVE THAWING CYCLES DID NOT RESOLVE THIS. MULTIPLE FREEZING CYCLES WERE CONDUCTED IN AN ATTEMPT TO REACH AN ACCEPTABLE ABLATION ZONE. THE TREATMENT WAS DEEMED UNSUCCESSFUL, AND A NEW CRYOABLATION PROCEDURE HAS BEEN SCHEDULED FOR THIS PATIENT. ALL FOUR NEEDLES WERE RETURNED TO THE MANUFACTURER FOR INVESTIGATION, AND THE CRYOABLATION SYSTEM WAS INVESTIGATED IN A FIELD SERVICE CALL. THIS MDR SUMMARISES THE INVESTIGATION RESULTS OF THE SECOND NEEDLE USED IN THIS COMPLAINT. PLEASE SEE THE FOLLOWING MDR'S FOR THE OTHER DEVICES USED DURING THIS CRYOABLATION PROCEDURE: 3004462490-2020-00001 - CRYOABLATION SYSTEM, 9616793-2020-00035 - NEEDLE 1, 9616793-2020-00037 - NEEDLE 3, 9616793-2020-00038 - NEEDLE 4.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9767059 ↗
  • MalfunctionFeb 27, 2020Failure to Fire; Failure to Fire

    4 NEEDLES WERE USED IN A MRI GUIDED RENAL CRYOABLATION PROCEDURE. DURING THE CASE, ALL 4 ICEROD MRI NEEDLES ONLY REACHED AN ICEBALL SIZE OF APPROX. 1 CM, WHICH WAS INSUFFICIENT FOR THE PROCEDURE. EXTRA ACTIVE THAWING CYCLES DID NOT RESOLVE THIS. MULTIPLE FREEZING CYCLES WERE CONDUCTED IN AN ATTEMPT TO REACH AN ACCEPTABLE ABLATION ZONE. THE TREATMENT WAS DEEMED UNSUCCESSFUL, AND A NEW CRYOABLATION PROCEDURE HAS BEEN SCHEDULED FOR THIS PATIENT. ALL FOUR NEEDLES WERE RETURNED TO THE MANUFACTURER FOR INVESTIGATION, AND THE CRYOABLATION SYSTEM WAS INVESTIGATED IN A FIELD SERVICE CALL. THIS MDR SUMMARISES THE INVESTIGATION RESULTS OF THE THIRD NEEDLE USED IN THIS COMPLAINT. PLEASE SEE THE FOLLOWING MDR'S FOR THE OTHER DEVICES USED DURING THIS CRYOABLATION PROCEDURE: 3004462490-2020-00001 - CRYOABLATION SYSTEM, 9616793-2020-00035 - NEEDLE 1, 9616793-2020-00036 - NEEDLE 2, 9616793-2020-00038 - NEEDLE 4

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9767057 ↗
  • InjuryNov 1, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS IS A FOLLOW UP #1 TO REPORT INVESTIGATION FINDINGS. AS REPORTED IN THE INITIAL: LIMITED INFORMATION WAS REPORTED THAT FOLLOWING A LIVER CRYOABLATION PROCEDURE WITH AN ICEROD MRI NEEDLE, IT WAS NOTICED THAT THE PATIENT HAD A HEAT SKIN BURN. INFORMATION REGARDING TREATMENT OR INTERVENTION WAS NOT PROVIDED. THE NEEDLE WILL BE RETURNED FOR INVESTIGATION.

    Patient problems: Burn, Thermal

    MAUDE report 9266838 ↗
  • InjuryMar 16, 2019Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING A MRI PROCEDURE THERE WAS OBSERVATION OF A SKIN BURN DUE TO AN APPARENT PROCEDURAL ISSUE. DURING FOLLOW-UP, THE DOCTOR CONFIRMED THAT HE DOES NOT USE ANY FORM OF EXTERNAL/TOPICAL SKIN PROTECTION AS A STANDARD OF PRACTICE. HE STATED THAT THEY HAD A GLOVE FILLED WITH WARM SALINE AS WELL AS A CONTINUOUS FLOW OF WARM SALINE THAT WAS RUN OVER THE NEEDLE INSERTION SITES. THE TUMOR WAS SUPERFICIAL (APPX 2CM BELOW THE SKIN). 7 ICEROD NEEDLES WERE PLACED AT FAIRLY STEEP ANGLES IN ALTERNATING OPPOSING DIRECTIONS TO ALLOW FOR TUMOR COVERAGE AT A MORE SUPERFICIAL ANGLE. BASED ON THE IMAGING, THE FROST BITE WAS FROM THE SHAFT CONTACTING THE SKIN. IT IS UNKNOWN IF ANY TREATMENT WAS PERFORMED.

    Patient problems: Burn, Thermal

    MAUDE report 8426629 ↗

Other Unit, Cryosurgical, Accessories devices

All Unit, Cryosurgical, Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page