IceSphere® 1.5 S 90° Cryoablation Needle FPRPR3561Recalls & reported problems
Boston Scientific Corporation · 1.5mm 90 degree Cryoablation Needle · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07290104830556), received between Mar 18, 2019 and Jun 26, 2023 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Wrong Label” (2 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Nov 18, 2020Class IICompleted
Complaint trend regarding needle shaft gas leaks.
Z-0672-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Wrong Label2
- Adverse Event Without Identified Device or Use Problem1
- Gas/Air Leak1
- Thermal Decomposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Consequences Or Impact To Patient1
- Pain1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJun 26, 2023Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT PNEUMATOSIS OCCURRED AFTER USE OF THE CRYOABLATION NEEDLE. AN ICESPHERE S 90 DEGREE NEEDLE WAS SELECTED FOR A CRYOABLATION PROCEDURE TO TREAT A KIDNEY LESION. THE PROBE WAS TESTED. IT WAS REPORTED THAT NO GAS LEAKS WERE SEEN, AND THE ICE BALL WAS VISUALIZED; HOWEVER, THERE WAS ONE SMALL BUBBLE SEEN, BUT NO STREAM OF BUBBLES. THE PHYSICIAN INSPECTED THE NEEDLE AND DETERMINED THAT IT LOOKED FINE, SO THE NEEDLE WAS USED. THE PROBE WAS INSERTED INTO THE PATIENT THROUGH THE POSTERIOR BACK MUSCLE, AND IT COULD NOT BE VISUALIZED ON THE SCREEN, SO IT WAS LOCKED INTO THE CHANNEL AGAIN AND PROMPTED TO RE-TEST IT. THIS TIME, THE PROBE WAS TESTED INSIDE THE PATIENT. THE FIRST FREEZE WAS COMPLETED WITH NO ISSUE. AT THE END OF THE SECOND FREEZE, HOWEVER, AFTER 10 MINUTES, THE CT IMAGE SHOWED AIR IN THE SUBCUTANEOUS MUSCLE. THE NEEDLE WAS REMOVED AND RE-TESTED IN SALINE. SMALL BUBBLES WERE SEEN FROM WHERE THE NEEDLE MEETS THE HANDLE. AFTER PRESSING THE FREEZE FUNCTION WHILE IN THE SALINE, MULTIPLE GAS BUBBLES WERE PRESENT. THE NEEDLE HAD BEEN COMPLETED SUCCESSFULLY WITH THE ORIGINAL DEVICE. THE PATIENT WAS IN STABLE CONDITION AFTER THE PROCEDURE, AND NO INTERVENTION WAS NEEDE
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 17201669 ↗MalfunctionJan 18, 2023Wrong Label; Wrong Label
IT WAS REPORTED THAT THE DEVICE PACKAGING WAS MISLABELED. AN ICESPHERE S 90 DEGREE NEEDLE WAS OPENED FROM THE PACKAGING. THE PACKAGE INCORRECTLY LABELED THE ICESPHERE S 90 DEGREE NEEDLE AS A 175CM NEEDLE INSTEAD OF A 100CM NEEDLE. THE NEEDLE WAS NOT USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16196883 ↗MalfunctionJan 18, 2023Wrong Label; Wrong Label
IT WAS REPORTED THAT THE DEVICE PACKAGING WAS MISLABELED. AN ICESPHERE S 90 DEGREE NEEDLE WAS OPENED FROM THE PACKAGING. THE PACKAGE INCORRECTLY LABELED THE ICESPHERE S 90 DEGREE NEEDLE AS A 175CM NEEDLE INSTEAD OF A 100CM NEEDLE. THE NEEDLE WAS NOT USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16196881 ↗MalfunctionOct 27, 2020Thermal Decomposition of Device; Thermal Decomposition of Device
IT WAS REPORTED THAT THE INSULATION OF A CRYOABLATION NEEDLE FAILED DURING A PROCEDURE. THE ICESPHERE NEEDLE PASSED INTEGRITY TESTING WITHOUT ISSUE; HOWEVER, THREE MINUTES INTO THE FIRST FREEZE CYCLE FROST DEVELOPED ON THE NEEDLE SHAFT. THE PHYSICIAN USED STERILE WARMING BAGS AND SALINE TO PROTECT THE PATIENT'S SKIN UNTIL THE NEEDLE WAS PASSIVELY THAWED. THE DEVICE WAS REMOVED AND REPLACED WITH A NEW ICESPHERE NEEDLE. NO PATIENT COMPLICATIONS WERE ENCOUNTERED AND THE TREATMENT WAS COMPLETED SUCCESSFULLY WITHOUT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10744654 ↗InjuryMar 18, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT SHORTLY, AFTER A NERVE PROCEDURE WAS COMPLETED THAT PATIENT EXPERIENCED PAIN. ONE DOSE OF MORPHINE WAS GIVEN AND THE PATIENT FELT FINE. THE HOSPITAL TECHNICIAN THOUGHT THE PAIN WAS DUE TO HYDRO DISSECTION. THERE WAS NO ALLEGED DEVICE MALFUNCTION. NO OTHER DETAILS PROVIDED SUGGEST A MORE SIGNIFICANT PROCEDURE RELATED ISSUE.
Patient problems: Pain
MAUDE report 8429972 ↗
Other Unit, Cryosurgical, Accessories devices
- IceFORCE® 2.1 CX 90° Cryoablation Needle FPRPR3602Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls4 deaths274 reports
- IceRod® 1.5 CX 90° Cryoablation Needle FPRPR3533Boston Scientific Corporation · Unit, Cryosurgical, Accessories1 recall168 reports
- IcePearl® 2.1 CX 90° Cryoablation Needle FPRPR3601Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls78 reports
- IceFORCE™ 2.1 CX L 90° Cryoablation Needle FPRPR3618Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls6 reports
- IceFORCE® 2.1 CX Cryoablation Needle FPRPR3604Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls3 reports
- IcePearl™ 2.1 CX L 90° Cryoablation Needle FPRPR3617Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls1 report
- IceRod® He-Thaw Prostate Cryoablation Kit / Presice® System FPRPR4570Galil Medical Ltd · Unit, Cryosurgical, Accessories2 recalls0 reports
- IceRod® i-Thaw® Prostate Cryoablation Kit / Presice® System FPRPR4070Galil Medical Ltd · Unit, Cryosurgical, Accessories2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page