Skip to content

Piccolo Composite Plate System MTP Plate - NarrowRecalls & reported problems

Carbofix Orthopedics Ltd · A plate indicated for treatment of deformations, fractures, nonunions and replantations of the 1st metatarsophalangeal joint and 1st metatarsal bone. · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07290110201692), received between Feb 17, 2021 and Feb 17, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Crack” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Crack1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 17, 2021Crack

    DURING IMPLANTATION OF AN MTP PLATE, TWO SCREWS OF A LARGER DIAMETER THAN INDICATED WERE INSERTED INTO THE PLATE HOLES. THE PLATE CRACKED (NEAR THE HOLE WHERE LOCKING SCREW WAS INSERTED). THE PLATE WAS REMOVED AND REPLACED WITH CROSSING CANNULATED SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11336543 ↗

Other Plate, Fixation, Bone devices

All Plate, Fixation, Bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page