XVS Straight Clamp AMCH00005Recalls & reported problems
Augmedics Ltd · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07290113780002), received between Aug 25, 2020 and Aug 25, 2020 — classified as 3 injuries.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method3
Patient problems
- Spinal Column Injury2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryAug 25, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
THE MOVEMENT OF THE CLAMP MAY CAUSE FOR INACCURATE REGISTRATION AND OR DISPLAY OF TRACKING INFORMATION, WHICH MAY LEAD TO SERIOUS INJURY. THE DISLOCATION OF THE CLAMP BY MUSCLE AND FASCIA WAS EVIDENT, HENCE, THE SURGEON PLACED THE CLAMP AGAIN AND REPEATED THE SCAN. IT IS SUSPECTED THAT THE CLAMP WAS NOT CLOSED AS REQUIRED. AUGMEDICS' PERSONNEL REVIEWED THE INSTRUCTIONS AND INDICATIONS OF COMPLETE CLAMP CLOSURE AGAIN WITH THE SURGEON. NO PATIENT HARM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10446962 ↗InjuryAug 25, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
THE EVENT IS RELATED TO THE PROCEDURE PERFORMED ON (B)(6) 2020. THE SURGEON REPORTED THAT 3 OF THE SCREWS DEVIATED IN POSITION. THE FIRST SCREW INSERTED WAS ACCURATE. DURING THE INVESTIGATION IT WAS FOUND OUT THE SURGEON WAS WORKING THOUGH SMALL INCISION IN OPEN PROCEDURE WHICH CAUSED FOR CONSTANT REARRANGEMENTS OF RETRACTORS AND INSTRUMENTATION HANDLING IN CLOSE VICINITY OF THE CLAMP. THE AUGMEDICS REPRESENTATIVE SUSPECTS THAT THE CLAMP CLOSING SCREW WAS NOT FULLY TIGHTNENED. IT WAS CONCLUDED THAT THE CLAMP WAS NOT TIGHTEN ENOUGH BY THE SURGEON AND THEREFORE WAS MOVED BY THE HANDLING OF THE RETRACTOR OR INSTRUMENTATION THE 3 SCREWS WERE REMOVED AND REPLACED USING FLUORO. NO ADDITIONAL INTERVENTION WAS REQUIRED. NO ADDITIONAL ADVERSE EVENTS WERE REPORTED. THE SURGEON WAS RE-TRAINED ON INDICATION FOR FULLY CLOSED CLAMP.
Patient problems: Spinal Column Injury
MAUDE report 10446961 ↗InjuryAug 25, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
THE EVENT IS RELATED TO THE PROCEDURE PERFORMED ON (B)(6) 2020. THE SURGEON REPORTED THAT 3 OF THE SCREWS DEVIATED IN POSITION. THE FIRST SCREW INSERTED WAS ACCURATE. DURING THE INVESTIGATION IT WAS FOUND OUT THE SURGEON WAS WORKING THOUGH SMALL INCISION IN OPEN PROCEDURE WHICH CAUSED FOR CONSTANT REARRANGEMENTS OF RETRACTORS AND INSTRUMENTATION HANDLING IN CLOSE VICINITY OF THE CLAMP. THE AUGMEDICS REPRESENTATIVE SUSPECTS THAT THE CLAMP CLOSING SCREW WAS NOT FULLY TIGHTENED. IT WAS CONCLUDED THAT THE CLAMP WAS NOT TIGHTEN ENOUGH BY THE SURGEON AND THEREFORE WAS MOVED BY THE HANDLING OF THE RETRACTOR OR INSTRUMENTATION. THE 3 SCREWS WERE REMOVED AND REPLACED USING FLUORO. NO ADDITIONAL INTERVENTION WAS REQUIRED. NO ADDITIONAL ADVERSE EVENTS WERE REPORTED. THE SURGEON WAS RE-TRAINED ON INDICATION FOR FULLY CLOSED CLAMP.
Patient problems: Spinal Column Injury
MAUDE report 10446960 ↗
Other Orthopedic Augmented Reality devices
- ARVIS ® IN-28000Kico Knee Innovation Company Pty Limited · Orthopedic Augmented Reality2 recalls0 reports
- Perc Pin Adaptor AMCH07200Augmedics Ltd · Orthopedic Augmented Reality1 recall0 reports
- All-in-one computer + Pre-installed Software ACC00008Augmedics Ltd · Orthopedic Augmented Reality26 reports
- XVS Headset IPD-60 Clear Display AHMD601WLTAugmedics Ltd · Orthopedic Augmented Reality4 reports
- NextAR Spine Platform 03.57.10.0026Medacta International SA · Orthopedic Augmented Reality2 reports
- NextAR Spine Platform 03.57.10.0173Medacta International SA · Orthopedic Augmented Reality2 reports
- MyKnee NextAR Tibial Distal Cutting Block Efficiency 15.1242Medacta International SA · Orthopedic Augmented Reality1 report
- NextAR TKA Platform 02.23.10.0011Medacta International SA · Orthopedic Augmented Reality1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page