The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630345728968), received between Oct 26, 2023 and Jan 19, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 19, 2024Fracture; Fracture

    DURING NEXTAR SURGERY, THE TIP OF SCREWDRIVER BROKE DURING PEDICLE SCREW 8.0X85MM INSERTION. TAPPING 8.0 WAS PERFORMED INITIALLY. IT TOOK TIME TO RETRIEVE THE BROKEN PIECE FROM THE PEDICLE SCREW HEAD, BUT FINALLY PEDICLE SCREW WAS ABLE TO RETRIEVE. ADDITIONAL TAPPING PERFORMED AND THE SURGERY FINISHED WITH PEDICLE SCREW 8.0X75MM. BACKUP SCREWDRIVER (SAME LOT) SUCCESSFULLY USED. DELAY TIME 40 MIN TO RETRIEVE THE BROKEN PART OF THE SCREWDRIVER AND REVISE THE PEDICLE SCREW. TOTAL SURGERY TIME ABOUT 4 HOURS AND 30 MIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18543116 ↗
  • MalfunctionOct 26, 2023Fracture; Fracture

    DURING A NEXTAR CASE, WHILE INSERTING S2AI SCREW (10X100MM), THE TIP OF NEXTAR SPINE MUST SCREWDRIVER - SHORT BROKE. NO SCREW WAS INSERTED IN S2AI AND THE SURGERY WAS COMPLETED SUCCESSFULLY. TAPPING WAS PERFORMED USING REFERENCES 03.51.10.0438 -> 03.51.10.0439 -> 03.51.10.0440. ALL FRAGMENTS WERE REMOVED. DELAY TIME 30 MIN. TOTAL SURGERY TIME 7H. THE DELAY WAS DUE TO THE TIME NECESSARY TO RETRIEVE THE FRAGMENTS OF THE TIP OF SCREWDRIVER.

    Patient problems: No Clinical Signs, Symptoms or Conditions