NextAR Spine Platform 03.57.10.0026Recalls & reported problems

Medacta International SA · NextAR Spine Double Short Clamp 25° · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630971292499), received between Feb 6, 2024 and Apr 11, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1
  • Mechanical Jam1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Bone Fracture(s)1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 11, 2024Fracture; Fracture

    THE FIXATION SCREW OF THE NEXTAR SPINE DOUBLE SHORT CLAMP 25° WAS BROKEN AND IT WAS NOT POSSIBLE TO REMOVE THE CLAMP AT THE END OF THE SURGERY. IT HAS BEEN NECESSARY TO BREAK THE SPINOUS PROCESS IN ORDER TO REMOVE THE INSTRUMENT. THE DELAY TIME WAS 30/40 MINUTE AND THE TOTAL TIME OF THE SURGERY WAS 4H.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 19085957 ↗
  • InjuryFeb 6, 2024Mechanical Jam; Mechanical Jam

    DURING NEXTAR SPINE SURGERY, AFTER 4 K-WIRES (USED FOR PEDICLE SCREW INSERTION GUIDE) HAVE BEEN POSITIONED, THE NEXTAR SPINE DOUBLE SHORT CLAMP 25° WAS NOT STABLE ANYMORE ON THE BONE AND SOME MOVEMENTS OF IT HAVE BEEN DETECTED. WHEN THE SURGEON TRIED TO REPOSITION THE CLAMP, IT WAS NOT POSSIBLE TO MOUNT IT NEWLY ON THE PATIENT'S BONE. THE SCREW OF THE CLAMP WAS BLOCKED. THEREFORE, THE SURGEON POSITIONED THE LONG CLAMP 25° (ALTERNATIVE INSTRUMENT) ON THE SAME PROCESSUS SPINOUS TO INSERT 2 ADDITIONAL K-WIRES AND IN THAT MOMENT, A FRACTURE OF THE PROCESSUS SPINOUS HAS BEEN DETECTED IN TH5/TH6. THE PATIENT'S BONE QUALITY WAS POOR AND SEVERELY OSTEOPOROTIC. THE CONVENTIONAL INSTRUMENTATION WAS USED TO REPOSITION THE 2 K-WIRES AND FINISH THE SURGERY. THE SURGEON DECIDED TO NOT FIX THE FRACTURE AND TO CLOSE THE PATIENT. DELAY TIME IS APPROXIMATELY 15 MINUTES.

    Patient problems: Bone Fracture(s)

    MAUDE report 18648847 ↗