Servo-s Base Unit 6640440Recalls & reported problems

Maquet Critical Care AB · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Maquet Servo-s ventilator has 799 adverse event reports, including 3 deaths and 18 injuries. Many describe output problems, failed self-tests and gas leaks, and most mention no effect on the patient. No recalls name this device. Reports fell from 263 in the previous 12 months to 168 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
795
Deaths reported
3
Injuries reported
20
Recalls naming it
0

Summary

The FDA's MAUDE database contains 795 adverse event reports that list this exact device (DI 07325710001127), received between Nov 20, 2020 and Sep 30, 2026 — classified as 3 deaths, 20 injuries, 772 malfunctions.

The most frequently reported specific device problem is “Output Problem” (392 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 19 reportsDec 2023: 24 reportsJan 2024: 10 reportsFeb 2024: 9 reportsMar 2024: 6 reportsApr 2024: 20 reportsMay 2024: 33 reportsJun 2024: 24 reportsJul 2024: 29 reportsAug 2024: 24 reportsSep 2024: 32 reportsOct 2024: 31 reportsNov 2024: 30 reportsDec 2024: 27 reportsJan 2025: 32 reportsFeb 2025: 22 reportsMar 2025: 19 reportsApr 2025: 26 reportsMay 2025: 22 reportsJun 2025: 21 reportsJul 2025: 26 reportsAug 2025: 17 reportsSep 2025: 10 reportsOct 2025: 11 reportsNov 2025: 14 reportsDec 2025: 45 reportsJan 2026: 17 reportsFeb 2026: 22 reportsMar 2026: 12 reportsApr 2026: 16 reportsMay 2026: 15 reportsJun 2026: 15 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 8 reportsOct 2026: 0 reports20242025202645
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem392
  • Failure of Device to Self-Test305
  • Mechanical Problem56
  • Infusion or Flow Problem52
  • Contamination44
  • Gas/Air Leak35
  • Improper Flow or Infusion17
  • Leak/Splash13

Patient problems

  • No Clinical Signs, Symptoms or Conditions771
  • Low Oxygen Saturation21
  • Bradycardia2
  • High Blood Pressure/ Hypertension2
  • Insufficient Information2
  • Irregular Pulse2
  • Loss of consciousness1
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 795 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Infusion or Flow Problem

    It was reported that the ventilator generated an alarm indicating high O2 concentration. There was no patient harm. Manufacturer's Ref. #: (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946619 ↗
  • InjurySep 25, 2026Infusion or Flow Problem

    It was reported that during patient ventilation, ventilator generated an O2 Sensor failure alarm while connecting to an O2 concentrator. Patient's saturation decreased to unknown level. The ventilator was replaced.Final patient outcome was no harm.Manufacturer's Ref.#: (b)(4).

    Patient problems: Low Oxygen Saturation

    MAUDE report 50898137 ↗
  • MalfunctionSep 25, 2026Failure of Device to Self-Test

    It was reported that the ventilator failed Internal leakage test during Pre-use Check. There was no patient involvement. Manufacturer's Ref. #:(b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50895928 ↗
  • InjurySep 25, 2026Infusion or Flow Problem

    It was reported that during patient ventilation, ventilator generated an O2 Sensor failure alarm while connecting to an O2 concentrator. Patient's saturation decreased to unknown level. The ventilator was replaced.Final patient outcome was no harm.Manufacturer's Ref.#: (b)(4).

    Patient problems: Low Oxygen Saturation

    MAUDE report 50894406 ↗
  • MalfunctionSep 24, 2026Mechanical Problem; Failure of Device to Self-Test

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884408 ↗
  • MalfunctionSep 24, 2026Failure of Device to Self-Test

    It was reported that the ventilator failed Internal leakage test with a message 'Pressure Transducer difference > 5 cmH2O' during Pre-use Check. There was no patient involvement.Manufacturer's Ref.#: (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884403 ↗
  • MalfunctionSep 24, 2026Failure of Device to Self-Test

    It was reported that the ventilator failed Pressure transducer test during Pre-use Check. There was no patient involvement. Manufacturer's Ref. #: (b)(4)

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50880938 ↗
  • MalfunctionSep 24, 2026Failure of Device to Self-Test

    It was reported that the ventilator failed Internal leakage test with a message 'System volume too large' during Pre-use Check. There was no patient involvement.Manufacturer's Ref. #: (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50880049 ↗
  • MalfunctionJun 29, 2026Leak/Splash

    IT WAS REPORTED THAT THE VENTILATOR HAD A LEAKAGE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25649997 ↗
  • MalfunctionJun 22, 2026Contamination; Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED FLOW TRANSDUCER TEST DURING PRE-USE CHECK. MOREOVER, THE AIR GAS MODULE WAS CONTAMINATED WITH WATER. THERE WAS NO PATIENT HARM.MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25576675 ↗
  • MalfunctionJun 22, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED O2 CELL/SENSOR TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25576373 ↗
  • MalfunctionJun 19, 2026Mechanical Problem; Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED INTERNAL LEAKAGE TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25558134 ↗
  • MalfunctionJun 18, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED PRESSURE TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REFERENCE #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25545205 ↗
  • MalfunctionJun 18, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED PRESSURE TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25543175 ↗
  • MalfunctionJun 18, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED THE INTERNAL LEAKAGE TEST DURING THE PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25542470 ↗
  • MalfunctionJun 16, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED PRESSURE TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25516187 ↗
  • MalfunctionJun 12, 2026Infusion or Flow Problem

    IT WAS REPORTED THAT THE VENTILATOR GENERATED AN ALARM OF LOW O2 CONCENTRATION. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF#: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25486098 ↗
  • MalfunctionJun 9, 2026Improper Flow or Infusion

    IT WAS REPORTED THAT THE VENTILATOR GENERATED ALARMS FOR O2 CONCENTRATION. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25439397 ↗
  • MalfunctionJun 8, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED INTERNAL LEAKAGE TEST, PRESSURE TRANSDUCER TEST, SAFETY VALVE TESTS, O2 CELL/SENSOR TEST, FLOW TRANSDUCER TEST AND PATIENT CIRCUIT TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25428993 ↗
  • MalfunctionJun 3, 2026Gas/Air Leak

    IT WAS REPORTED THAT THERE WAS CONTINUOUS GAS LEAKAGE. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF#: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25369863 ↗
  • MalfunctionJun 2, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED INTERNAL LEAKAGE TEST AND PRESSURE TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25358545 ↗
  • MalfunctionJun 2, 2026Output Problem

    IT WAS REPORTED THAT THE VENTILATOR GENERATED A TECHNICAL ERROR CODE INDICATING DISABLED VALVES. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25357210 ↗
  • MalfunctionJun 1, 2026Mechanical Problem; Gas/Air Leak

    IT WAS REPORTED THAT THE VENTILATOR HAD CONTINUOUS GAS LEAKAGE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25353475 ↗
  • MalfunctionMay 26, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED INTERNAL LEAKAGE TEST AND FLOW TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF.:# (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25276053 ↗
  • MalfunctionMay 24, 2026Gas/Air Leak

    IT WAS REPORTED THAT THE VENTILATOR HAD A CONTINUOUS GAS LEAK. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25266738 ↗