NA 6431233BRecalls & reported problems

Gambro Dasco SPA · S-660-C Accessory Y connector · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07332414004699), received between Nov 23, 2021 and May 20, 2025 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak5
  • Air/Gas in Device4
  • Break1
  • Connection Problem1
  • Crack1
  • Fitting Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 20, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    IT WAS REPORTED THAT AN ACCESSORY Y CONNECTOR WAS OBSERVED TO BE CRACKED AT THE ARTERIAL BRANCH CONNECTION AND LEAKED BLOOD DURING THERAPY. THE VOLUME OF BLOOD LOSS WAS UNKNOWN. THERE WAS NO REPORT OF MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22052000 ↗
  • MalfunctionMar 28, 2025Air/Gas in Device; Air/Gas in Device

    IT WAS REPORTED THAT DURING DIALYSIS TREATMENT USING A Y CONNECTOR LINE ACCESSORY, ¿SIGNALING OF AN AIR INGRESS WAS IDENTIFIED. IT WAS REPORTED THAT THE LINE WAS CLAMPED, AND THE Y CONNECTOR WAS REPLACED PRIOR TO ANY AIR BUBBLES ENTERING THE PATIENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21724592 ↗
  • MalfunctionMar 18, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN EXTERNAL BLOOD LEAK WAS OBSERVED IN THE AREA OF THE ARTERIAL LIMB OF AY CONNECTOR S660 LINE ACCESSORY DURING THERAPY. THE VOLUME OF BLOOD LOSS WAS APPROXIMATELY 100ML. THEY CONNECTOR WAS CHANGED TO CONTINUE TREATMENT. THE PATIENT WAS REPORTED TO HAVE EXPERIENCED "AIR INSUFFLATION", HOWEVER THERE WAS NO REPORT OF MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21635133 ↗
  • MalfunctionJan 31, 2025Fluid/Blood Leak; Connection Problem; Fluid/Blood Leak; Connection Problem

    IT WAS REPORTED THAT A Y CONNECTOR S 660 LINE ACCESSORY WAS DISCONNECTED FROM THE PATIENT CATHETER AND LEAKED BLOOD DURING HEMODIALYSIS THERAPY. THE VOLUME OF BLOOD LOSS WAS UNKNOWN. THERE WAS NO REPORT OF MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21287502 ↗
  • MalfunctionJan 12, 2025Fluid/Blood Leak; Air/Gas in Device; Fluid/Blood Leak; Air/Gas in Device

    IT WAS REPORTED THAT AN EXTERNAL BLOOD LEAK WAS OBSERVED FROM THE VENOUS LINE OF A S 660 ACCESSORY Y CONNECTOR. AIR ENTERED THE LINE AT THE ARTERIAL LINE CONNECTION POINT AND RESULTED IN THE LEAK. THE VOLUME OF BLOOD LOSS WAS UNKNOWN. THERE WAS NO REPORT OF A MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21134136 ↗
  • MalfunctionJan 10, 2025Fluid/Blood Leak; Air/Gas in Device; Fluid/Blood Leak; Air/Gas in Device

    IT WAS REPORTED THAT AN EXTERNAL BLOOD LEAK WAS OBSERVED FROM A Y CONNECTOR S 660 LINE ACCESSORY DURING HEMODIALYSIS THERAPY. A MICROFOAM ALARM WAS TRIGGERED AFTER (2) HOURS AND (5) MINUTES OF TREATMENT. THE VOLUME OF BLOOD LOSS WAS UNKNOWN. THE CONNECTOR WAS REPLACED TO CONTINUE THERAPY. THERE WAS NO REPORT OF MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21127742 ↗
  • InjuryDec 20, 2023Break; Air/Gas in Device; Break; Air/Gas in Device

    IT WAS REPORTED THAT THE PATIENT UNDERWENT CONTINUOUS RENAL REPLACEMENT THERAPY (CRRT) USING A FILTER SET AND A Y CONNECTOR. IT WAS FURTHER REPORTED THAT WHILE ATTEMPTING TO RETURN BLOOD AND START A NORMAL SALINE RECIRCULATION, THE DEVICE ¿MALFUNCTIONED¿ AND AS A RESULT, THE EXTRACORPOREAL (EC) BLOOD WAS NOT RETURNED. ACCORDING TO THE REPORTER, A SECOND Y CONNECTOR WAS OBTAINED, HOWEVER, IT WAS NOTICED TO BE BROKEN AT THE CONNECTION HUB. A THIRD Y CONNECTOR WAS REPORTED TO BE NEEDED, HOWEVER, TOO MUCH AIR HAD ENTERED THE CIRCUIT DUE TO THE BROKEN Y-CONNECTOR, AND TREATMENT WAS TERMINATED WITHOUT THE EC BLOOD RETURNING TO THE PATIENT. THE PATIENT RECEIVED A BLOOD TRANSFUSION LATER IN THE SHIFT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18369771 ↗
  • MalfunctionNov 23, 2021Fitting Problem; Fitting Problem

    IT WAS REPORTED THAT THREE (3) ACCESSORIES Y CONNECTOR S 660 C DID NOT CONNECT PROPERLY ON THE DIALYSIS LINE. NO LEAKS WERE REPORTED TO HAVE OCCURRED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12863366 ↗

Other SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE devices

All SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE devices →