Skip to content

PRISMAFLEX SET M60Recalls & reported problems

Gambro Industries · The Prismaflex M60 set is indicated for use only with the PrismaFlex control unit or with the PrisMax control unit (in countries where PrisMax is cleared or registered) in providing continuous fluid management and renal replacement therapies. The system is intended for patients who have acute renal failure, fluid overload, or both. · Class 2

All models of PRISMAFLEX SET (8)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07332414064549), received between Mar 21, 2018 and Aug 6, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (6 reports).

1 FDA recall record name this device's identifier, 1 still open.

6 other recalls cover devices cleared under the same FDA submission (K041005); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 14, 2022Class IIOpen, Classified

    Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.

    Z-0081-2023 · Root cause: Error in labeling · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak6
  • Leak/Splash2
  • Break1
  • Crack1
  • Hole In Material1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient2
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2026Obstruction of Flow

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 50160442 ↗
  • MalfunctionNov 25, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT PRIOR TO USE, A LEAK WAS OBSERVED AT THE "GREEN LINE" OF A PRISMAFLEX M60 SET. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23645625 ↗
  • MalfunctionApr 26, 2024Break; Fluid/Blood Leak; Break; Fluid/Blood Leak

    IT WAS REPORTED THAT THE "PRESSURE BYPASS VALVE" OF A PRISMAFLEX M60 SET WAS BROKEN AND LEAKING DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19193429 ↗
  • MalfunctionMay 9, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED AN EXTERNAL FLUID LEAK WAS OBSERVED ORIGINATING FROM THE AIR CHAMBER OF A PRISMAFLEX M60 SET DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16894767 ↗
  • MalfunctionSep 13, 2022Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT FEW MINUTES AFTER STARTING THERAPY, ONE UNIT OF PRISMAFLEX M60 SET HAD A BLOOD LEAK AT THE RETURN LINE UP NEAR THE RETURN / VENOUS CLAMP AND JUST BELOW THAT AREA. THE AMOUNT OF BLOOD LOSS WAS ESTIMATED TO BE AROUND 200ML. THERE WAS PATIENT INVOLVEMENT HOWEVER, NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15412830 ↗
  • MalfunctionMar 29, 2022Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    IT WAS REPORTED THAT DURING PRIMING WITH ONE UNIT OF PRISMAFLEX M60 SET, A CRACK WAS OBSERVED IN THE DE-AERATION CHAMBER AND FLUID LEAKED OUT RESULTING IN AN UNSPECIFIED ALARM BEING TRIGGERED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13939265 ↗
  • MalfunctionSep 23, 2021Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK WAS OBSERVED FROM THE TOP AREA OF THE DIALYZER ON A PRISMAFLEX M60 SET DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12519180 ↗
  • MalfunctionJul 20, 2018Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT TEN MINUTES INTO TREATMENT WITH A PRISMAFLEX M60 SET , A BLOOD LEAK AT THE JUNCTION AFTER BLOOD PUMP TUBE WAS OBSERVED. BLOOD WAS RETURNED TO THE PATIENT AND THEY WERE DISCONNECTED AND TREATMENT ENDED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7707742 ↗
  • MalfunctionMar 21, 2018Fluid/Blood Leak; Hole In Material; Fluid/Blood Leak; Hole In Material

    IT WAS REPORTED THAT A PRISMAFLEX M60 SET CLOTTED DURING USE, AND BLOOD LEAKED FROM THE PBP TUBING. THE FILTER CLOTTED AT THE PBP LINE, WHERE IT WRAPS AROUND THE BLOOD PUMP. AFTER THE SET CLOTTED, THE REPORTER NOTICED A LEAK FROM WHAT APPEARED TO BE A TINY PIN HOLE IN THE TUBING. TREATMENT WAS STOPPED AND THE PATIENT WAS DISCONNECTED. THE BLOOD IN THE CIRCUIT WAS NOT RETURNED TO THE PATIENT. THE PATIENT HAD MINIMAL BLOOD LOSS. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7360091 ↗

FDA clearance and classification

Cleared through Abbreviated 510(k) K041005 on Oct 7, 2004 (decision: Substantially Equivalent). Submitted by Gambro Renal Products as “Gambro Prismaflex And Gambro Prismaflex M60 & M100 Sets”. The FDA review took 171 days. FDA record ↗

The FDA regulates product code KDI (dialyzer, high permeability with or without sealed dialysate system) under 21 CFR 876.5860 as a Class II device. 21 CFR 876.5860 ↗

The FDA lists this type of device as life-sustaining or life-supporting.

Recalls under the same FDA submission

These recalls involve devices cleared under K041005. One submission often covers many models, so check the product and lot details in each record.

  • Z-1227-2026 · Jan 6, 2026 · Open, Classified
    PRIMSAFLEX M100 SET, Product Code 106697; Dialyzer, High Permeability With Or Without Sealed Dialysate System
  • Z-2378-2021 · Jul 22, 2021 · Terminated
    Baxter PrismaFlex, REF numbers: a) 106696 (M 60 SET) b) 106697 (M 100 SET) Expansion: c) 107144 (TPE2000 SET) d) 115307 (M 150 SET) e) 115309 (ST 100 SET) f) 955467 (M 150 SET)
  • Z-2202-2014 · Jul 18, 2014 · Terminated
    Prismaflex M100. Product number 106697. Each set is packaged in a plastic pouch, 4 pouches per carton box
  • Z-2201-2014 · Jul 18, 2014 · Terminated
    Prismaflex M60 set. Product number 106696. Each set is packaged in a plastic pouch, 4 pouches per carton box
  • Z-0695-2007 · Jan 4, 2007 · Terminated
    Gambro Prismaflex Continuous Renal Replacement System, Catalog Number 602314700, Gambro Lundia AB, Sweden.
  • Z-0032-2007 · May 16, 2006 · Terminated
    Gambro Prismaflex Hemodialysis Machine, Catalogue Number 6023014700, Gambro Dasco S.p.A, Italy.

Common questions about the PRISMAFLEX SET M60

Has the PRISMAFLEX SET M60 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0081-2023 (Class II), began on Sep 14, 2022: Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the PRISMAFLEX SET M60?

The FDA's MAUDE database lists 9 adverse event reports for this device, received between Mar 21, 2018 and Aug 6, 2026. In those reports, the most common specific device problems are “Fluid/Blood Leak” (6), “Leak/Splash” (2) and “Break” (1); the most common patient problem is “Hemorrhage/Blood Loss/Bleeding” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PRISMAFLEX SET M60?

None of the reports are classified as deaths, 0 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the PRISMAFLEX SET M60 FDA approved?

It was cleared through Abbreviated 510(k) K041005 on Oct 7, 2004. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the PRISMAFLEX SET M60, and is it still sold?

The FDA's device database lists Gambro Industries as the company responsible for it (brand PRISMAFLEX SET, model M60). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Dialyzer, high permeability with or without sealed dialysate system devices

All Dialyzer, high permeability with or without sealed dialysate system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page