Nobel Biocare® 33452Recalls & reported problems
Nobel Biocare AB · Healing Abutment NobRpl RP Ø5.3x5mm · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 07332747002881), received between May 23, 2019 and May 23, 2019 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Inaccurate Information” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Inaccurate Information10
Patient problems
- No Known Impact Or Consequence To Patient10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON (B)(6) 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBRPL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: THE AFFECTED LOT 13076845 WAS RESTRICTED ON MAY 01, 2019. A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLEMENT
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636046 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON (B)(6) 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBEL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: THE AFFECTED LOT 13076845 WAS RESTRICTED ON (B)(6) 2019. A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLEMENT C
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636044 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON (B)(6) 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBEL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: THE AFFECTED LOT 13076845 WAS RESTRICTED ON (B)(6) 2019. A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLEMENT C
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636043 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON APRIL 23, 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBRPL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: · THE AFFECTED LOT 13076845 WAS RESTRICTED ON MAY 01, 2019. · A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IM
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636041 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON APRIL 23, 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBRPL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN. THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: THE AFFECTED LOT 13076845 WAS RESTRICTED ON MAY 01, 2019. A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLEM
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636040 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON (B)(6) 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBRPL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: THE AFFECTED LOT 13076845 WAS RESTRICTED ON (B)(6) 2019. A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLEMENT
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636038 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON APRIL 23, 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBRPL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT. A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: THE AFFECTED LOT 13076845 WAS RESTRICTED ON MAY 01, 2019. A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLEM
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636036 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON APRIL 23, 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBRPL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: THE AFFECTED LOT 13076845 WAS RESTRICTED ON MAY 01, 2019. A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLEM
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636034 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON (B)(6) 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBRPL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: · THE AFFECTED LOT 13076845 WAS RESTRICTED ON MAY 01, 2019. · A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLE
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636031 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON (B)(6) 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBRPL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: · THE AFFECTED LOT 13076845 WAS RESTRICTED ON MAY 01, 2019. · A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLE
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636029 ↗
FDA clearance and classification
The FDA regulates product code NHA (abutment, implant, dental, endosseous) under 21 CFR 872.3630 as a Class II device. 21 CFR 872.3630 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: To be used in conjunction with an endosseous dental implant fixture to aid in prosthetic rehabilitation.
Common questions about the Nobel Biocare® 33452
Has the Nobel Biocare® 33452 been recalled?
No FDA recall record in our data names this exact device identifier (DI 07332747002881), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Nobel Biocare® 33452?
The FDA's MAUDE database lists 10 adverse event reports for this device, received on May 23, 2019. In those reports, the most common specific device problem is “Inaccurate Information” (10). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Nobel Biocare® 33452?
None of the reports are classified as deaths, 0 as injuries and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Nobel Biocare® 33452 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the Nobel Biocare® 33452, and is it still sold?
The FDA's device database lists Nobel Biocare AB as the company responsible for it (brand Nobel Biocare®, model 33452). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Abutment, implant, dental, endosseous devices
- CERTAIN® GOLD-TITE® IUNIHGZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS150 reports
- Nobel Biocare® 36662Nobel Biocare AB · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS2 recalls5 reports
- CERTAIN® IUNIHTZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS67 reports
- ATLANTIS® 35502Dentsply Ih, Inc. · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS66 reports
- ATLANTIS® 35502Dentsply Ih, Inc. · Abutment, implant, dental, endosseous61 reports
- Esthetic Abutment 36669Nobel Biocare AB · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS1 recall2 reports
- CERTAIN® BELLATEK® IEDAT5ZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS1 recall1 report
- CERTAIN® BELLATEK® ILDAN3ZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page