Nobel Biocare® 33452Recalls & reported problems
Nobel Biocare AB · Healing Abutment NobRpl RP Ø5.3x5mm · Class 2 · Implant
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 07332747002881), received between May 23, 2019 and May 23, 2019 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Inaccurate Information” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Inaccurate Information10
Patient problems
- No Known Impact Or Consequence To Patient10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON (B)(6) 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBRPL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: THE AFFECTED LOT 13076845 WAS RESTRICTED ON MAY 01, 2019. A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLEMENT
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636046 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON (B)(6) 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBEL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: THE AFFECTED LOT 13076845 WAS RESTRICTED ON (B)(6) 2019. A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLEMENT C
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636044 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON (B)(6) 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBEL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: THE AFFECTED LOT 13076845 WAS RESTRICTED ON (B)(6) 2019. A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLEMENT C
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636043 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON APRIL 23, 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBRPL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: · THE AFFECTED LOT 13076845 WAS RESTRICTED ON MAY 01, 2019. · A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IM
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636041 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON APRIL 23, 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBRPL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN. THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: THE AFFECTED LOT 13076845 WAS RESTRICTED ON MAY 01, 2019. A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLEM
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636040 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON (B)(6) 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBRPL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: THE AFFECTED LOT 13076845 WAS RESTRICTED ON (B)(6) 2019. A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLEMENT
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636038 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON APRIL 23, 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBRPL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT. A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: THE AFFECTED LOT 13076845 WAS RESTRICTED ON MAY 01, 2019. A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLEM
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636036 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON APRIL 23, 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBRPL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: THE AFFECTED LOT 13076845 WAS RESTRICTED ON MAY 01, 2019. A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLEM
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636034 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON (B)(6) 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBRPL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: · THE AFFECTED LOT 13076845 WAS RESTRICTED ON MAY 01, 2019. · A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLE
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636031 ↗MalfunctionMay 23, 2019Inaccurate Information; Inaccurate Information
ON (B)(6) 2019 A CUSTOMER IN THE US RETURNED 10 PRODUCTS LABELED 36644 (HEALING ABUTMENT CC RP Ø3.6X5MM), LOT 13076845, STATING THE PACKAGES CONTAIN AN INCORRECT HEALING ABUTMENT. INSPECTION PERFORMED BY NOBEL BIOCARE CONFIRMED THAT ALL TEN PACKAGES CONTAIN ARTICLE 33452 (HEALING ABUTMENT NOBRPL RP Ø5.3X5MM), LOT 13076868. INVESTIGATION CONCLUDED THAT 16 PIECES OF ARTICLE 33452, LOT 13076868 HAD BEEN INCORRECTLY LABELED AS ARTICLE 36644, LOT 13076845. AN EXTRA LABEL SHEET (INCLUDING 16 INDIVIDUAL LABELS) HAD BEEN PRINTED FOR COMPLAINED LOT 13076845. INSTEAD OF THE SHEET BEING SCRAPPED, IT WAS IN ERROR USED ON ARTICLE 33452, LOT 13076868. A HEALTH HAZARD EVALUATION WAS PERFORMED CONCLUDING ON A HIGH RISK FOR THE PATIENT: A MISFIT BETWEEN THE ABUTMENT AND THE IMPLANT COULD POTENTIALLY LEAD TO INFLAMMATION OR DAMAGE TO THE IMPLANT DUE TO INCREASED FORCES ON THE IMPLANT THREADS. BOTH ISSUES COULD RESULT IN A NEED FOR REVISION SURGERY. CORRECTIVE OR REMOVAL ACTIONS THAT HAVE BEEN, AND ARE EXPECTED TO BE TAKEN: THE FOLLOWING CORRECTIVE ACTIONS WERE IMPLEMENTED: · THE AFFECTED LOT 13076845 WAS RESTRICTED ON MAY 01, 2019. · A CAPA WAS INITIATED TO FURTHER INVESTIGATE THE ISSUE AND TO IMPLE
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8636029 ↗
Other Abutment, implant, dental, endosseous devices
- CERTAIN® GOLD-TITE® IUNIHGZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS151 reports
- Nobel Biocare® 36662Nobel Biocare AB · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS2 recalls5 reports
- CERTAIN® IUNIHTZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS67 reports
- ATLANTIS® 35502Dentsply Ih, Inc. · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS66 reports
- ATLANTIS® 35502Dentsply Ih, Inc. · Abutment, implant, dental, endosseous54 reports
- Esthetic Abutment 36669Nobel Biocare AB · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS1 recall2 reports
- CERTAIN® BELLATEK® IEDAT5ZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS1 recall1 report
- Custom Abutments Final AbutmentsInstitut Straumann AG · Abutment, implant, dental, endosseous1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.