GETINGE 9100-SERIES 9128ERecalls & reported problems
Getinge Disinfection AB · WASHER DISINFECTOR 9128E · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 07340153710061), received between Aug 13, 2021 and Jan 5, 2024 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Device Handling Problem” (6 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 1, 2022Open, Classified
Inadequate documentation verifying whether device installation has been executed in accordance with the manufacturer's requirements. This can lead to installation failure modes and hazardous situations.
Z-1078-2022 · Root cause: Other · FDA record ↗
Most reported problems
Device problems
- Device Handling Problem6
- Adverse Event Without Identified Device or Use Problem2
- Contamination /Decontamination Problem1
- Detachment of Device or Device Component1
- Device Dislodged or Dislocated1
- Device Fell1
- Misassembly During Maintenance/Repair1
Patient problems
- Bruise/Contusion4
- Muscle/Tendon Damage3
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 5, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem
ON DECEMBER 8TH, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH THE 91E-SERIES WASHER DISINFECTOR, WITH THE MODEL NAME 9128E. AS IT WAS STATED DEBRIS WAS ON THE INSTUMENTS. THERE WAS NO ALARM ON THE DEVICE. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION, THAT ANY IMPROPERLY CLEANED LOADS USED FOR PATIENTS¿ TREATMENT COULD BE THE SOURCE OF INFECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18449613 ↗MalfunctionDec 14, 2023Device Handling Problem; Device Handling Problem
MANUFACTURER'S REFERENCE NUMBER: (B)(4).
Patient problems: Muscle/Tendon Damage
MAUDE report 18325077 ↗MalfunctionDec 14, 2023Device Handling Problem; Device Handling Problem
MANUFACTURER'S REFERENCE NUMBER: (B)(4).
Patient problems: Muscle/Tendon Damage
MAUDE report 18325073 ↗MalfunctionDec 14, 2023Device Handling Problem; Device Handling Problem
MalfunctionDec 14, 2023Device Handling Problem; Device Handling Problem
MalfunctionApr 7, 2023Device Fell; Misassembly During Maintenance/Repair; Device Fell; Misassembly During Maintenance/Repair
MalfunctionDec 1, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON 21ST NOVEMBER, 2022 GETINGE BECAME AWARE OF AN ISSUES RELATED TO MANUAL TROLLEY. THE MANUAL TROLLEY IS NOT REGISTERED AS A MEDICAL DEVICE, HOWEVER UPON THE SITUATION OCCURRENCE IT WAS USED WITH THE MEDICAL DEVICE WASHER DISINFECTOR WITH THE MODEL NAME 9128E. AS IT WAS STATED, THE LOADING TROLLEY ALMOST FELL ON THE FLOOR DUE TO A FAULTY STOP PIN THAT IS INSTALLED ON THE MANUAL TROLLEY TO PREVENTS THE LOADING TROLLEYS FROM FALLING OFF. THE MALFUNCTIONED PART WAS REPLACED AND THE MANUAL TROLLEY WAS RETURNED TO THE SERVICE IN FULL WORKING CONDITION. THERE WAS NO INJURY OR DAMAGE REPORTED, HOWEVER WE DECIDED TO REPORT THE ISSUE BASED ON A POTENTIAL AS LOADING TROLLEY FALLING TO THE FLOOR COULD BRING A HAZARDOUS SITUATION FOR THE OPERATOR AND LEAD TO SERIOUS INJURY IF THE SITUATION WAS TO REOCCUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15894321 ↗MalfunctionMar 19, 2022Adverse Event Without Identified Device or Use Problem; Device Handling Problem; Adverse Event Without Identified Device or Use Problem; Device Handling Problem
ON 2ND MARCH, 2022 GETINGE RECEIVED THE INFORMATION ABOUT THE EVENT WHICH WAS RELATED TO MANUAL TROLLEY UNIVERSAL WASH CART 2000 THAT IS USED WITH THE 91E-SERIES WASHER DISINFECTOR. AS IT WAS STATED ON (B)(6) 2022 AN OPERATOR REPORTED A BRUISE ON HIS FOOT DURING THE EXTRACTION OF THE INSTRUMENTS FROM THE HIGHEST SHELF OF THE TROLLEY. AFTER THE MEDICAL EXAMINATION THE OPERATOR WAS 7 DAYS OFF WORK TO RECOVER FROM THE INJURY. BASED ON THE AVAILABLE INFORMATION WE CANNOT CLASSIFY THE HEALTH IMPACT, HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AND BASED ON THE POTENTIAL AS WE CANNOT CONFIRM THAT SIMILAR EVENTS COULD NOT CONTRIBUTED TO THE SERIOUS INJURY IF REOCUR.
Patient problems: Bruise/Contusion
MAUDE report 13820059 ↗MalfunctionMar 19, 2022Adverse Event Without Identified Device or Use Problem; Device Handling Problem; Adverse Event Without Identified Device or Use Problem; Device Handling Problem
ON 2ND MARCH, 2022 GETINGE RECEIVED THE INFORMATION ABOUT THE EVENT WHICH WAS RELATED TO THE MANUAL TROLLEY UNIVERSAL WASH CART 2000 USED WITH THE 91E-SERIES WASHER DISINFECTOR. AS IT WAS STATED ON (B)(6) 2021 AN OPERATOR REPORTED A STRAIN INJURY WHILE TOWING THE TROLLEY IN THE WASHING TUNNEL. THE OPERATOR REPORTED THAT THE WHEELS WERE STUCK IN THE GROOVES. AFTER THE MEDICAL EXAMINATION THE OPERATOR WAS 7 DAYS OFF WORK TO RECOVER FROM THE INJURY. BASED ON THE AVAILABLE INFORMATION WE CANNOT CLASSIFY THE HEALTH IMPACT, HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AND BASED ON THE POTENTIAL AS WE CANNOT CONFIRM THAT SIMILAR EVENTS COULD NOT CONTRIBUTED TO THE SERIOUS INJURY IF REOCCUR.
Patient problems: Muscle/Tendon Damage
MAUDE report 13820049 ↗MalfunctionAug 13, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
MANUFACTURER REFERENCE NUMBER: (B)(6) .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12318871 ↗
Other Disinfector, Medical Devices devices
- GETINGE 88-SERIES 88-5Getinge Disinfection AB · Disinfector, Medical Devices3 recalls6 reports
- GETINGE 88-SERIES 88-5Getinge Disinfection AB · Disinfector, Medical Devices3 recalls0 reports
- GETINGE CM320-SERIESGetinge Disinfection AB · Disinfector, Medical Devices2 recalls3 reports
- GETINGE CM320-SERIESGetinge Disinfection AB · Disinfector, Medical Devices2 recalls1 report
- GETINGE 86-SERIES S-8668Getinge Disinfection AB · Disinfector, Medical Devices1 recall37 reports
- Getinge Aquadis 56 56MGetinge Disinfection AB · Disinfector, Medical Devices1 recall11 reports
- AMSCO 7052HPCorporation Steris Canada · Disinfector, medical devices1 recall10 reports
- AMSCO 7053HPCorporation Steris Canada · Disinfector, medical devices1 recall10 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.