Skip to content

GETINGE CM320-SERIES (REF CM320WWD)Recalls & reported problems

Getinge Disinfection AB · CM320-3 CHAMBER WWD. Washer disinfector with three chambers; Washer chamber followed by another wash chamber and a dryer chamber. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Dec 21, 2020Terminated

    The potential that the plastic inlet hose connections may deform or break as a result of material fatigue and/or over tightening of the clamps. A broken hose connection may lead to a hazardous situation of water leakage or water spray. The water leakage or water spray may result in a burn to the user and/or an inconvenience due to a procedural delay if the Washer Disinfector is taken out of service.

    Z-1124-2021 · Root cause: Device Design · FDA record ↗

Other Disinfector, Medical Devices devices

All Disinfector, Medical Devices devices →